Basic + Good = Elegant Simplicity
Keysight Technologies
U3400 Series 4½ and 5½ Digit
Digital Multimeters
Data Sheet
Simply right for your basic needs, without
compromising quality
The U3400 Series digital multimeters (DMMs) come with functions vital for your everyday
measurements: DC, AC and AC+ DC voltage and current, 2- and 4-wire resistance,
frequency, continuity and diode tests.
Math functions dBm, Min/Max, Relative, Compare, Hold and Percentage** are also
available to ease your measurement analysis.
Not only are the U3400 Series built robust for lasting reliability over the long haul, these
DMMs also ensure dependable measurements with up to 0.012% basic DCV accuracy.
Efcient testing with dual display and selectable
resolutions*
Dual display on the U3400 Series lets you view two parameters simultaneously as you
measure, and further enhances your troubleshooting tasks. For example, theability to see
both AC voltage and frequency simultaneously would help you to measure the frequency
response of amplier circuits more efciently and effectively. Refer to Page 3 for more
typical dual display combinations and applications.
Up to three measurement speeds are available on the U3402A: Slow, Medium and Fast.
This means greater exibility in catering to different testing needs: a faster measurement
speed at lower resolution or a higher resolution at slower measurement speed.
Physical security
Your instruments may be at risk of theft or misplacement whenever you leave them
unattended on the bench. With the U3400 Series’ rear Kensington lock slot, you can
secure your DMM and be assured that it is where you expect it to be for your continued
testing the next day.
Features
Up to 120,000 counts
resolution
Up to 0.012% basic DCV
accuracy
11 basic measurements
and up to six built-in math
functions
Dual display on bright VFD
Selectable resolutions for
variable measurement
speeds*
Kensington lock slot
security
* U3402A only
** U3401A only
02 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Take a closer look
Primary display Secondary display
Input terminals
and current fuse
AC line
voltage selector
Line voltage fuse
selection table
AC power
connector
RS232 interface connector
(For calibration use only)
AC line fuse Chassis ground lug
Measurement
function
keypads
Autoranging,
manual range
and comparator
operation
Math
operation
keypads
Power
on/off
switch
Shift
Figure 1 Front panel of the U3402A. The front panel of the U3401A is similar with
slight differences in the position of certain functions. Please refer to the U3401A’s
User’s and Service Guide U3401-90001 for details.
Figure 2 Rear panel of the U3401A/U3402A.
03 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Typical dual display combinations and applications
Primary display Secondary display Application Available in
DCV ACV Testing of DC-to-AC or AC-to-DC converter circuit U3402A, U3401A
ACV + DCV DCV Power supply DC level and AC ripple measurement U3402A, U3401A
DCV DCI Testing of power supply load regulation U3402A
DCV ACI Loop current and voltage drop level checking U3402A
ACI + DCI DCV Testing of line and load regulation U3402A
ACV DCI Testing of AC-to-DC or DC-to-AC converter U3402A
ACI + DCI ACV Power supply DC level and AC ripple measurement U3402A
ACV ACI Transformer testing U3402A
ACV Hz Amplier circuit’s AC frequency response measurement U3402A, U3401A
ACI Hz Adjustment of AC motor control U3402A, U3401A
DCI ACI Power supply AC ripple and DC current measurement U3402A, U3401A
ACI + DCI DCI Current dissipation measurement U3402A, U3401A
dBm Reference W
RF frequency measurement
U3401A
dBm DCV U3402A, U3401A
dBm ACV U3402A, U3401A
dBm Hz Frequency response checking U3402A
04 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
U3400 Series specications
DC voltage
DCV resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Typical input impedance1
500.00 mV 10 µV 510.00 0.02% + 4 10.0 MW
5.0000 V 100 µV 5.1000 0.02% + 4 11.1 MW
50.000 V 1 mV 51.000 0.02% + 4 10.1 MW
500.00 V 10 mV 510.00 0.02% + 4 10.0 MW
1000.0 V 100 mV 1200.020.02% + 4 10.0 MW
1. Input impedance is in parallel with capacitance <100 pF.
2. In VDC 1000 V range, 1200 V is readable with audio warning.
Rate Range Resolution Maximum reading Accuracy
(1 year; 23 ºC ± 5 ºC)
Typical input impedance1
Slow 120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 V
1 µV
10 µV
100 µV
1 mV
10 mV
119.999
1.19999
11.9999
119.999
1000.003
0.012% + 82
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
Medium 400.00 mV
4.0000 V
40.000 V
400.00 V
1000.0 V
10 µV
100 µV
1 mV
10 mV
100 mV
399.99
3.9999
39.999
399.99
1000.03
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
Fast 400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
100 µV
1 mV
10 mV
100 mV
1 V
399.9
3.999
39.99
399.9
10003
0.012% + 2
0.012% + 2
0.012% + 2
0.012% + 2
0.012% + 2
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
1. Input impedance is in parallel with capacitance <120 pF.
2. Relative (REL) operation is used.
3. In VDC 1000 V range, 1050 V is readable.
05 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
AC voltage (true RMS, AC coupling mode)
ACV resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)1
30 Hz to 50 Hz 50 Hz to 10 kHz 10 kHz to 30 kHz 30 kHz to 100 kHz
500.00 mV 10 µV 510.00 1% + 40 0.05% + 40 2% + 60 3% + 20
5.0000 V 100 µV 5.1000 1% + 20 0.35% + 15 1% + 20 3% + 50
50.000 V 1 mV 51.000 1% + 20 0.35% + 15 1% + 20 3% + 50
500.00 V 10 mV 510.00 Not specied 0.5% + 15 1% + 2023% + 502
750.0 V 100 mV 1000.0 Not specied 0.5% + 1531% + 202Not specied
1. Accuracy specied at input >5% of full scale.
2. Input voltage <200 V RMS.
3. For 5 kHz to 10 kHz, accuracy is 0.7% + 15.
4. In VAC 750 V range, 1000.0 V RMS is readable with audio warning.
Rate Range Resolution Maximum
reading
Accuracy (1 year; 23 ºC ± 5 ºC)1
20 Hz to 45 Hz 45 Hz to 10 kHz 10 kHz to 30 kHz 30 kHz to 100 kHz
Slow 120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
1 µV
10 µV
100 µV
1 mV
10 mV
119.999
1.19999
11.9999
119.999
750.004
1% + 100
1% + 100
1% + 100
1% + 100
1% + 1002
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
1.5% + 100
1% + 100
1% + 100
1% + 100
1% + 100
5% + 3002
3% + 2002
3% + 2002
3% + 2002
3% + 2003
Medium 400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
10 µV
100 µV
1 mV
10 mV
100 mV
399.99
3.9999
39.999
399.99
750.0
1% + 40
1% + 40
1% + 40
1% + 402
1% + 402
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
1.5% + 80
1% + 40
1% + 40
1% + 40
1% + 40
5% + 1202
3% + 802
3% + 802
3% + 802
3% + 803
Fast 400.0 mV
4.000 V
40.00 V
400.0 V
750 V
100 µV
1 mV
10 mV
100 mV
1 V
399.9
3.999
39.99
399.9
750
1% + 5
1% + 5
1% + 5
1% + 52
1% + 52
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
1.5% + 10
1% + 5
1% + 5
1% + 5
1% + 5
5% + 152
3% + 102
3% + 102
3% + 102
3% + 103
1. Accuracy specied at input >5% of full scale.
2. Input voltage <200 V RMS.
3. Input voltage <500 V RMS.
4. In VAC 750 V range, 787.5 V RMS is readable.
06 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
AC voltage (true RMS, AC+DC coupling mode)
AC+DC voltage resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)1
50 Hz to 10 kHz 10 kHz to 30 kHz 30 kHz to 100 kHz
500.00 mV 10 µV 510.00 0.5% + 50 2% + 70 3% + 130
5.0000 V 100 µV 5.1000 0.5% + 25 1% + 30 3% + 60
50.000 V 1 mV 51.000 0.5% + 25 1% + 30 3% + 60
500.00 V 10 mV 510.00 0.5% + 25 1% + 3023% + 602
750.0 V 100 mV 1000.030.5% + 2541% + 302Not specied
1. Accuracy specied at input >5% of full scale.
2. Input voltage <200 V RMS.
3. In VAC 750 V range, 1000.0 V RMS is readable with audio warning.
4. For 5 kHz to 10 kHz, accuracy is 0.7% + 25.
Rate Range Resolution Maximum
reading
Accuracy (1 year; 23 ºC ± 5 ºC)1
45 Hz to 10 kHz 10 kHz to 30 kHz 30 kHz to 100 kHz
Slow 120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
1 µV
10 µV
100 µV
1 mV
10 mV
119.999
1.19999
11.9999
119.999
750.0002
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
1.5% + 300
1% + 100
1% + 100
1% + 100
1% + 100
5% + 300
3% + 200
3% + 200
3% + 200
3% + 2003
Medium 400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
10 µV
100 µV
1 mV
10 mV
100 mV
399.99
3.9999
39.999
399.99
750.00
0.2% + 45
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 43
1.5% + 83
1% + 43
1% + 43
1% + 43
1% + 43
5% + 125
3% + 83
3% + 83
3% + 83
3% + 833
Fast 400.0 mV
4.000 V
40.00 V
400.0 V
750 V
100 µV
1 mV
10 mV
100 mV
1 V
399.9
3.999
39.99
399.9
750.0
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
1.5% + 12
1% + 7
1% + 7
1% + 7
1% + 7
5% + 18
3% + 12
3% + 12
3% + 12
3% + 123
1. Accuracy specied at input >5% of full scale.
2. In VAC 750 V range, 787.5 V RMS is readable.
3. Input voltage <500 V RMS.
07 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
DC current
DCI resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Burden voltage1 and shunt resistor
500.00 µA 10 nA 510.00 0.05% + 5 <0.06 V/100 W
5.0000 mA 100 nA 5.1000 0.05% + 4 <0.6 V/100 W
50.000 mA 1 µA 51.000 0.05% + 4 <0.08 V/1 W
500.00 mA 10 µA 510.00 0.05% + 4 <0.8 V/1 W
5.0000 A 100 µA 5.1000 0.25% + 5 <0.3 V/0.01 W
10.000 A 1 mA 20.00020.25% + 5 <0.6 V/0.01 W
1. Typical at full-scale reading and voltage across the input terminals.
2. In 10 A range, >10 to 20 ADC is readable for 20 seconds maximum with audio warning.
Rate Range Resolution Maximum reading Accuracy
(1 year; 23 ºC ± 5 ºC)
Burden voltage1
and shunt resistor
Slow 12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
0.1 µA
1 µA
10 µA
100 µA
11.9999
119.999
1199.99
11.9999
0.05% + 152
0.05% + 5
0.2% + 5
0.2% + 5
<0.15 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
Medium 40.000 mA
120.00 mA
1200.0 mA
12.000 A
1 µA
10 µA
100 µA
1 mA
39.999
119.99
1199.9
11.999
0.1% + 6
0.1% + 3
0.2% + 3
0.2% + 3
<0.5 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
Fast 40.00 mA
120.0 mA
1200 mA
12.00 A
10 µA
100 µA
1 mA
10 mA
39.99
119.9
1199
11.99
0.1% + 2
0.1% + 2
0.2% + 2
0.2% + 2
<0.5 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
1. Typical at full-scale reading and voltage across the input terminals.
2. Relative (REL) operation is used.
08 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
AC current (true RMS, AC coupling mode)
ACI resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)1Burden voltage2
and shunt
resistor
30 Hz to 50 Hz 50 Hz to 2 kHz 2 kHz to 5 kHz 5 kHz to 20 kHz
500.00 µA 10 nA 510.00 1.5% + 50 0.5% + 20 1.5% + 50 3% + 753<0.06 V/100 W
5.0000 mA 100 nA 5.1000 1.5% + 40 0.5% + 20 1.5% + 40 3% + 60 <0.6 V/100 W
50.000 mA 1 µA 51.000 1.5% + 40 0.5% + 20 1.5% + 40 3% + 60 <0.08 V/1 W
500.00 mA 10 µA 510.00 1.5% + 40 0.5% + 20 1.5% + 40 3% + 60 <0.8 V/1 W
5.0000 A 100 µA 5.1000 2% + 4040.5% + 20 Not specied Not specied <0.3 V/0.01 W
10.000 A 1 mA 20.00052% + 4040.5% + 30
(<1 kHz)
Not specied Not specied <0.6 V/0.01 W
1. Accuracy specied at input >5% of full scale and >1 A for 10 A range unless otherwise stated.
2. Typical at full-scale reading and voltage across the input terminals.
3. Input current >35 µA RMS.
4. Input current <3 A RMS.
5. In 10 A range, >10 to 20 ADC is readable for 20 seconds maximum with audio warning.
Rate Range Resolution Maximum
reading
Accuracy (1 year; 23 ºC ± 5 ºC)1Burden voltage2
and shunt
resistor
20 Hz to 45 Hz 45 Hz to 2 kHz 2 kHz to 10 kHz
Slow 12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
0.1 µA
1 µA
10 µA
100 µA
11.9999
119.999
1199.99
11.9999
1.5% + 100
1.5% + 100
1.5% + 100
2% + 100 (<1.2 A)
0.5% + 100
0.5% + 100
0.5% + 100
1% + 100
2% + 200
2% + 200
2% + 200
Not specied
<0.15 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
Medium 40.000 mA
120.00 mA
1200.0 mA
12.000 A
1 µA
10 µA
100 µA
1 mA
39.999
119.99
1199.9
11.999
1.5% + 40
1.5% + 12
1.5% + 12
1.5% + 12 (<1.2 A)
0.5% + 40
0.5% + 12
0.5% + 12
1% + 12
2% + 80
2% + 30
2% + 30
Not specied
<0.5 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
Fast 40.00 mA
120.0 mA
1200 mA
12.00 A
10 µA
100 µA
1 mA
10 mA
39.99
119.9
1199
11.99
1.5% + 5
1.5% + 2
1.5% + 2
2% + 2 (<1.2 A)
0.5% + 5
0.5% +2
0.5% +2
1% +2
2% + 10
2.2% + 5
2.2% + 5
Not specied
<0.5 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
1. Accuracy specied at input >5% of full scale.
2. Typical at full-scale reading and voltage across the input terminals.
09 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
AC current (true RMS, AC+DC coupling mode)
AC+DC current resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)1Burden voltage2
and shunt
resistor
50 Hz to 2 kHz 2 kHz to 5 kHz 5 kHz to 20 kHz
500.00 µA 10 nA 510.00 0.5% + 30 1.5% + 60 3% + 853<0.06 V/100 W
5.0000 mA 100 nA 5.1000 0.5% + 30 1.5% + 50 3% + 70 <0.6 V/100 W
50.000 mA 1 µA 51.000 0.5% + 30 1.5% + 50 3% + 70 <0.08 V/1 W
500.00 mA 10 µA 510.00 0.5% + 30 1.5% + 50 3% + 70 <0.8 V/1 W
5.0000 A 100 µA 5.1000 0.5% + 30 Not specied Not specied <0.3 V/0.01 W
10.000 A 1 mA 20.00040.5% + 40 (<1 kHz) Not specied Not specied <0.6 V/0.01 W
1. Accuracy specied at input >5% of full scale and >1 A for 10 A range unless otherwise stated.
2. Typical at full-scale reading and voltage across the input terminals.
3. Input current >35 µA RMS.
4. In 10 A range, >10 to 20 ADC is readable for 20 seconds maximum with audio warning.
Rate Range Resolution Maximum
reading
Accuracy (1 year; 23 ºC ± 5 ºC)1Burden voltage2
and shunt resistor
45 Hz to 2 kHz 2 kHz to 10 kHz
Slow 12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
0.1 µA
1 µA
10 µA
100 µA
11.9999
119.999
1199.99
11.9999
0.5% + 100
0.5% + 100
0.5% + 100
1% + 100
2% + 200
2% + 200
2% + 200
Not specied
<0.15 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
Medium 40.000 mA
120.00 mA
1200.0 mA
12.000 A
1 µA
10 µA
100 µA
1 mA
39.999
119.99
1199.9
11.999
0.5% + 42
0.5% + 15
0.5% + 15
1% + 15
2% + 80
2% + 30
2% + 30
Not specied
<0.5 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
Fast 40.00 mA
120.0 mA
1200 mA
12.00 A
10 µA
100 µA
1 mA
10 mA
39.99
119.9
1199
11.99
0.5% + 7
0.5% + 4
0.5% + 4
1% + 4
2% + 12
2% + 7
2% + 7
Not specied
<0.5 V/10 W
<1.5 V/10 W
<0.3 V/0.1 W
<0.6 V/0.01 W
1. Accuracy specied at input >5% of full scale.
2. Typical at full-scale reading and voltage across the input terminals.
10 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Resistance
Resistance resolution, full scale reading, and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
Range1Resolution Maximum reading Test current Accuracy (1 year; 23 ºC ± 5 ºC)
500.00 W10 mW 510.00 0.5 mA 0.1% + 52
5.0000 kW 100 mW 5.1000 0.45 mA 0.1% + 32
50.000 kW1 W 51.000 45 µA 0.1% + 3
500.00 kW 10 W510.00 4.5 µA 0.1% + 3
5.0000 MW100 W5.1000 450 nA 0.1% + 3
50.000 MW1 kW 51.000 45 nA 0.3% + 3
1. To reduce noise interference that may be induced by the test leads, we recommend using a shielded test cable for measuring resistances above 500 kW.
2. Relative (REL) operation is used.
Rate Range1 Resolution Maximum
reading
Test current Accuracy (1 year; 23 ºC ± 5 ºC)
2-wire 4-wire
Slow 120.000 W
1.20000 kW
12.0000 kW
120.000 kW
1.20000 MW
12.0000 MW
120.000 MW
1 mW
10 mW
100 mW
1 W
10 W
100 W
1 kW
119.999
1.19999
11.9999
119.999
1.19999
11.9999
119.999
0.5 mA
0.5 mA
100 µA
10 µA
1 µA
100 nA
10 nA
0.1% + 82
0.08% + 52
0.06% + 52
0.06% + 5
0.06% + 5
0.3% + 5
3% + 8
0.05% + 82
0.05% + 52
0.05% + 5
0.05% + 5
0.05% + 5
0.3% + 5
3% + 8
Medium 400.00 W
4.0000 kW
40.000 kW
400.00 kW
4.0000 MW
40.000 MW
300.00 MW
10 mW
100 mW
1 W
10 W
100 W
1 kW
10 kW
399.99
3.9999
39.999
399.99
3.9999
39.999
299.99
0.5 mA
100 µA
50 µA
5 µA
500 nA
50 nA
10 nA
0.1% + 52
0.08% + 32
0.06% + 3
0.06% + 3
0.15% + 3
1.5% + 3
5.0% + 5
0.05% + 52
0.05% + 3
0.05% + 3
0.05% + 3
0.15% + 3
1.5% + 3
5.0% + 5
Fast 400.0 W
4.000 kW
40.00 kW
400.0 kW
4.000 MW
40.00 MW
300.0 MW
100 mW
1 W
10 W
100 W
1 kW
10 kW
100 kW
399.9
3.999
39.99
399.9
3.999
39.99
299.9
0.5 mA
100 µA
50 µA
5 µA
500 nA
50 nA
10 nA
0.1% + 22
0.08% + 2
0.06% + 2
0.06% + 2
0.15% + 2
1.5% + 2
5.0% + 2
0.05% + 2
0.05% + 2
0.05% + 2
0.05% + 2
0.15% + 2
1.5% + 2
5.0% + 2
1. To reduce noise interference that may be induced by the test leads, we recommend using a shielded test cable for measuring resistances above 100 kW.
2. Relative (REL) operation is used.
11 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Frequency
Frequency resolution, full-scale reading and accuracy [± (% of reading + count)]
Diode test/continuity
U3401A 4.5-digit DMM
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
U3402A 5.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)
2.3000 V 100 µV 2.3000 V 0.05% + 5
Rate Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)
Slow 1.19999 V 10.0000 µV 0.012% + 5
Medium 2.4999 V 100.00 µV 0.012% + 5
Fast 2.499 V 1.000 mV 0.012% + 2
Range Input sensitivity for voltage measurement (sine-wave)
5 Hz to 100 kHz 100 kHz to 500 kHz
500 mV 35 mV RMS 200 mV RMS
5 V 0.25 V RMS 0.5 V RMS
50 V 2.5 V RMS 5 V RMS
500 V 25 V RMS Not specied
750 V 50 V RMS Not specied
Range Measurement range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)
500.00 Hz 5 Hz to 500 Hz 0.01 Hz 510.00 0.01% + 5
5.0000 kHz 500 Hz to 5 kHz 0.1 Hz 5.1000 0.01% + 3
50.000 kHz 5 kHz to 50 kHz 1 Hz 51.000 0.01% + 3
500.00 kHz 50 kHz to 500 kHz 10 Hz 999.99 0.01% + 3
Range Measurement range Resolution Maximum reading Accuracy
(1 year; 23 ºC ± 5 ºC)1
Input sensitivity
(sine wave)
1200.00 Hz 5 Hz to 1200 Hz 10 mHz 1199.99 0.005% + 3 40 mV RMS
12.0000 kHz 10 Hz to 12 kHz 100 mHz 11.9999 0.005% + 2 40 mV RMS
120.000 kHz 100 Hz to 120 kHz 1 Hz 119.999 0.005% + 2 40 mV RMS
1.0000 MHz 1 kHz to 1 MHz 10 Hz 1.1999 0.005% + 2 0.5 V RMS
1. Specied accuracy at input >5% of full scale.
12 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Decibel (dB) calculation
Range and accuracy (±dB)
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
Voltage range1, 2 Input voltage dBm3 range at 600 W ref Accuracy (1 year; 23 ºC ± 5 ºC)
30 Hz to 50 Hz 50 Hz to 10 kHz 10 kHz to 100 kHz
500.00 mV 20 mV to 500 mV –29.82 to –3.80 0.3 0.3 0.7
5.0000 V 5000 mV to 5 V 3.80 to 16.20 0.2 0.2 0.5
50.000 V 5 V to 50 V 16.20 to 36.20 0.2 0.2 0.5
500.00 V 50 V to 500 V 36.20 to 56.20 0.240.2 0.54
1000.0 VDC 500 V to 1000 V 56.20 to 62.22 Not specied 0.25Not specied
750.0 VAC 500 V to 750 V 56.20 to 59.72 Not specied 0.25Not specied
1. Auto ranging is used when dBm operation is selected.
2. In VDC 1000 V range, 1200 V is readable. In VAC 750 V range, 1000 V is readable.
3. Reading is displayed in dB when relative (REL) operation is used.
4. Input voltage <200 V RMS.
5. For input voltage within the frequency range of 50 Hz to 1 kHz.
Rate Voltage range 1, 2 Input voltage dBm3 range
at 600 W ref
Accuracy (1 year; 23 ºC ± 5 ºC)
20 Hz to 45 Hz 45 Hz to 10 kHz 10 kHz to 100 kHz
Slow 120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 VDC
750.00 V
6 mV to 120 mV
120 mV to 1.2 V
1.2 V to 12 V
12 V to 120 V
120 V to 1000 V
120 V to 750 V
–42.20 to16.20
16.20 to 3.80
3.80 to 23.80
23.80 to 43.80
43.80 to 62.22
43.80 to 59.72
1.0
0.8
0.8
0.8
Not specied
Not specied
0.2
0.1
0.1
0.1
1.04
1.04
1.0
0.8
0.8
0.8
Not specied
Not specied
Medium 400.00 mV
4.0000 V
40.000 V
400.00 V
1000.0 VDC
750.0 V
20 mV to 400 mV
400 mV to 4 V
4 V to 40 V
40 V to 400 V
400 V to 1000 V
400 V to 750 V
–31.76 to –5.74
–5.74 to 14.26
14.26 to 34.26
34.26 to 54.26
54.26 to 62.22
54.26 to 59.72
1.0
0.8
0.8
0.8
Not specied
Not specied
0.2
0.1
0.1
0.1
1.04
1.04
1.0
0.8
0.8
0.8
Not specied
Not specied
Fast 400.0 mV
4.000 V
40.00 V
400.0 V
1000 VDC
750 V
20 mV to 400 mV
400 mV to 4 V
4 V to 40 V
40 V to 400 V
400 V to 1000 V
400 V to 750 V
–31.76 to –5.74
–5.74 to 14.26
14.26 to 34.26
34.26 to 54.26
54.26 to 62.22
54.26 to 59.72
1.0
0.8
0.8
0.8
Not specied
Not specied
0.2
0.1
0.1
0.1
1.04
1.04
1.0
0.8
0.8
0.8
Not specied
Not specied
1. Auto ranging is used when dBm operation is selected.
2. In VAC 750 V range, 5% over-range is readable.
3. Reading is displayed in dB when relative (REL) operation is used.
4. For input voltage within the frequency range of 45 Hz to 1 kHz.
13 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Full-scale display counts
Supplemental measurement specications
Supplemental specications
Model Reading rate Display counts
U3401A N/A 51,000
U3402A Slow 120,000
Medium 40,000
Fast 4,000
Measurement Specication
DC voltage Measurement method Sigma Delta A-to-D converter
Input resistance 10 MW ± 2% range (typical)
Maximum input voltage 1000 VDC or PEAK AC on all ranges
Input protection 1000 V on all ranges
Response time Approximately 1.0 second when the displayed reading reaches 99.9% DC value of the tested input
signal at the same range
DC current Shunt resistance U3401A
0.01 W to 100 W for 500 µA to 10 A ranges
U3402A
0.1 W to 10 W for 12 mA to 1.2 A ranges
0.01 W for 12 A range
Maximum input and
overload protection
(U3402A only)
mA input terminal: 1200 mADC or AC RMS. Protected with 1.25 A/500 V, IEC-127 sheet, FB fuse
12 A input terminal: 10 ADC or AC RMS continuous; or 12 ADC or AC RMS for 30 seconds maximum.
Protected with 15 A/600 V, breaking capacity 10,000 A FB fuse
Response time Approximately 1.0 second when the displayed reading reaches 99.9% DC value of the tested input
signal at the same range
AC voltage Measurement method AC-coupled true RMS: measures the AC component with up to 400 VDC bias on any range
Crest factor Maximum 3:0 at full scale
Input impedance U3401A: 1 MW in parallel with <100 pF
U3402A: 1 MW ± 2% in parallel with <120 pF
Maximum input voltage U3401A: 1000 V RMS/1400 V PEAK
U3402A: 750 V RMS/1200 V PEAK
2x107 V-Hz product on any range, normal mode input
1x106 V-Hz product on any range, common mode input
Overload ranging Selects higher range if peak input overload is detected during auto range. Overload is reported in
manual ranging
Input protection U3401A: 1000 V RMS on all ranges
U3402A: 750 V RMS on all ranges
Response time Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested
input signal at the same range.
14 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Supplemental measurement specications (continued)
Measurement Specication
AC+DC
voltage
Measurement method AC+DC coupled true RMS: measures the AC component with up to 400 VDC bias on any range
Crest factor Maximum 3:0 at full scale
Input impedance U3401A: 1 MW in parallel with <100 pF
U3402A: 1 MW ± 2% in parallel with <120 pF
Maximum input voltage U3401A: 1000 V RMS/1400 V PEAK
U3402A: 750 V RMS/1100 V PEAK
2x107 V-Hz product on any range, normal mode input
1x106 V-Hz product on any range, common mode input
Overload ranging Selects higher range if peak input overload is detected during auto range. Overload is reported in
manual ranging
Input protection U3401A: 1000 V RMS on all ranges
U3402A: 750 V RMS on all ranges
Response time Approximately 1.5 seconds (2.5 seconds for U3402A) when the displayed reading reaches 99.9% AC
RMS value of the tested input signal at the same range
AC current Measurement method DC coupled to the fuse and current shunt, AC coupled true rms measurement (measures the AC
component only)
Crest factor Maximum 3:0 at full scale
Shunt resistance U3401A
0.01 W to 100 W for 500 µA to 10 A ranges
U3402A
0.1 W to 10 W for 10 mA to 1.2 A ranges
0.01 W for 12 A range
Input protection U3401A
Front panel fuse 630 mA, 500 V; internal 25 A, 440 V
U3402A
mA input terminal: 1200 mADC or AC RMS. Protected with 1.25 A/500 V, IEC-127 sheet, FH fuse
12 A input terminal: 10 ADC or AC RMS continuous; or 12 ADC or AC RMS for 30 seconds maximum.
Protected with 15 A/600 V, breaking capacity 10,000 A FH fuse
Response time Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested
input signal at the same range
AC+DC
current
Measurement method AC+DC coupled to the fuse and current shunt, AC+DC coupled true rms measurement
(measures the AC component only)
Crest factor Maximum 3:0 at full scale
Measurement range Vdc and Vac are automatically set at the same range
Response time Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested
input signal at the same range
Resistance Measurement method U3401A: 2-wire ohms
U3402A: 2-wire ohms or 4-wire ohms
Open-circuit voltage U3401A: Limited to <+6 VDC
U3402A: Limited to <+5 VDC
Zeroing error 0.05 W or less (excluding test lead resistance) in each range when Relative operation is used
15 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Supplemental measurement specications (continued)
Measurement Specication
Resistance Input protection 500 V
Response time U3401A
Approximately 1.5 seconds for 5 MW and ranges below 5 MW
U3402A
Approximately 1.5 seconds for 12 MW and ranges below 12 MW
Approximately 5 seconds for 40 MW
Approximately 10 seconds for 120 MW
Approximately 23 seconds for 300 MW
Diode/
Continuity
Measurement method U3401A: 0.5mA +/- 0.2% constant current source, open circuit voltage limited to <6V.
U3402A: 0.5mA +/- 0.2% constant current source, open circuit voltage limited to <5V.
Test current Approximately 0.5 mADC
Open-circuit voltage U3401A: Limited to <+6 VDC
U3402A: Limited to <+5 VDC
Continuity threshold 10 W xed
Continuity level Approximately <+50 mVDC
Audible tone Continuous beep for continuity and single tone for normal forward-biased diode or semiconductor
junction
Input protection 500 V
Resistance/
Continuity
(U3402A only)
Measurement method 2-wire ohms
Test current Approximately 0.5 mADC
Open-circuit voltage Limited to <+5 VDC
Audible tone Continuous beep for continuity and single tone for normal forward-biased diode or semiconductor
junction
Zeroing error 0.05 W or less (excluding test lead resistances) in each range when Relative operation is used
Input protection 500 V
Frequency Measurement method Reciprocal counting technique. AC coupled input using AC voltage function
Crest factor Maximum 3:0 at full scale
Signal level 10% of range to full-scale input on all ranges; auto or manual range selection
Gate time 0.1 second or 1 period of the input signal, whichever is longer
Input impedance U3401A: 1 MW in parallel with <100 pF
U3402A: 1 MW ± 2% in parallel with <120 pF
Maximum input voltage U3401A: 1000 V RMS/1400 V PEAK
U3402A: 750 V RMS/1100 V PEAK
2x107 V-Hz product on any range, normal mode input
1x106 V-Hz product on any range, common mode input
Input protection 750 V RMS on all ranges
Response time Approximately 1.5 seconds when the displayed reading reaches 99.9% of frequency value
Noise
rejection
Common mode rejection
ratio (CMRR) for 1 kW
unbalanced LO lead
50/60 Hz ± 0.1%: DC >90 dB
Normal mode rejection
ratio (NMRR)
50/60 Hz ± 0.1%: >50 dB
16 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Supplemental measurement specications (continued)
Measurement Specication
dBm
operation
0 dBm 1 mW at 600 W reference impedance
Resolution U3401A
0.01 dB for all ranges
U3402A
Slow: 0.01 dB for all ranges
Medium: 0.01 dB for all ranges
Fast: 0.1 dB for all ranges
Reference impedance12 W2, 4 W 2, 8 W2, 16 W2, 50 W, 75 W, 93 W, 110 W, 124 W, 125 W, 135 W, 150 W, 250 W, 300 W,
500 W, 600 W, 800 W, 900 W, 1000 W, 1200 W, 8000 W
Math operation U3401A: dBm, Relative, Min/Max, Compare, Hold, Percentage
U3402A: dBm, Relative, Min/Max, Compare, Hold
I/O interface RS-232 (for calibration use only)
1. Reference impedance is displayed on the secondary display.
2. Reading is displayed in watts (audio power).
17 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
 

General characteristics
Power supply
100 V/120 V/220 V/240 V ± 10%
AC line frequency 50 Hz to 60 Hz
Power consumption
16 VA maximum
Input power option
Manual-ranging (100 VAC to
240 VAC ± 10%)
Fuse
U3401A
Terminal: 25 A, 440 V FB fuse; 0.63 A,
500 V FB fuse
Power line: 0.25 A, 250 V SB fuse, or
0.125 A, 250 V SB fuse
U3402A
Terminal: 15 A, 600 V FB fuse; 1.25 A,
500 V FB fuse
Power line: 0.25 A, 250 V SB fuse, or
0.125 A, 250 V SB fuse
Display
Highly visible vacuum uorescent
display (VFD)
Operating environment
Operating temperature from 0 °C to
+50 °C
Relative humidity up to 80% at 28 °C
RH
(non-condensing)
Altitude up to 2000 meters
Pollution degree 2
For indoor use only
Storage compliance
–20 °C to 60 °C
Relative humidity at 5% to 90% RH
(non-condensing)
Safety compliance
IEC 61010-1:2001/EN61010-1:2001
(2nd Edition)
Canada: CAN/CSA-C22.2 No. 61010-
1-04
USA: ANSI/UL 61010-1:2004
EMC compliance
IEC 61326-1:2005/EN61326-1:2006
Canada: ICES/NMB-001:2004
Australia/New Zealand: AS/NZS
CISPR11:2004
Shock and vibration
Tested to IEC/EN 60068-2
I/O connector
Output connectors
I/O interface
RS-232 (for calibration use only)
Dimensions (W × H × D)
255 mm × 105 mm × 305 mm (with bum-
pers)
215 mm x 87 mm x 282 mm (without
bumpers)
Weight
3.44 kg (with bumpers)
Warranty
Three years for U3401A/U3402A
Three months for standard shipped acces-
sories
Calibration cycle
One year
Warm-up time
At least 30 minutes
18 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
34138A Test Lead Kit U1161A Extended Test Lead Kit 34330A Current Shunt (30 A)
34133A Precision Electronic
Test Leads
11059A Kelvin Probe Set U34 00A-1CM Rack Mount Kit
Each U3400 Series includes these standard shipped accessories:
Quick Start Guide
Product Reference CD
Certicate of Calibration
34138A Test Lead Kit
AC Power Cord
Accessories
Ordering Information
19 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Some of the best equations express large ideas with elegant simplicity. The same is true with
the Keysight Technologies, Inc. U3401A and U3402A dual-display DMMs. When you need basic
capabilities with good performance at 4½ or 5½ digits, respectively, these affordable benchtop
units make the grade in educational, electronics and communications applications. For extra
credit, they also include a PC-grade lock slot for physical security. The U3400 Series: basic, good
capability in an elegantly simple DMM.
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. Keysight is a founding member of the AXIe consortium.
ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of
the PCI Industrial Computer Manufacturers Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the
Web inside your test systems. Keysight is a founding member of the LXI
consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure your
instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
www.keysight.com/find/low-cost-dmm
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada (877) 894 4414
Brazil 55 11 3351 7010
Mexico 001 800 254 2440
United States (800) 829 4444
Asia Pacic
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
Europe & Middle East
Austria 0800 001122
Belgium 0800 58580
Finland 0800 523252
France 0805 980333
Germany 0800 6270999
Ireland 1800 832700
Israel 1 809 343051
Italy 800 599100
Luxembourg +32 800 58580
Netherlands 0800 0233200
Russia 8800 5009286
Spain 0800 000154
Sweden 0200 882255
Switzerland 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom 0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-09-04-14)
20 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
This information is subject to change without notice.
© Keysight Technologies, 2010 - 2014
Published in USA, July 31, 2014
5990-3970EN
www.keysight.com