Fujian Lilliput Optoelectronics Technology Co., Ltd
Web: www.owon.com.cn /www.owontme.com E-mail: info@owon.com.cn Facebook: fb.me/owontech
Performance Specifications
Model: XSA1015TG
Frequency Specification
Frequency
Range
9kHz - 1.5 GHz
Resolution
1Hz
Frequency span
Range
0 Hz, 100 Hz to maximum frequency of device
Accuracy
± span /(swept points -1)
Internal reference
Reference frequency
10.000000 MHz
Reference frequency accuracy
±[(days from last calibrate × freq aging rate)+ temperature stability + initial
accuracy ]
Temperature stability
<2.5ppm
Aging rate
<1ppm/year
Readout
Marker frequency resolution
span/(the number of sweep points -1)
Uncertainty
±(freq indication × freq reference uncertainty +1%× span +10%× resolution
bandwidth + Marker Frequency Resolution)
Frequency counter
Resolution
1 Hz, 10 Hz, 100 Hz, 1 kHz
Accuracy
±(marker freq × freq reference uncertainty + counter resolution)
Bandwidth
Resolution bandwidth (-3 dB)
10Hz to 500kHz (in 1 to 10 sequence), 1MHz, 3MHz
Resolution filter shape factor
<5:1 nominal(Digital implement, similar to Gauss Pattern)
Accuracy
<5% nominal
Video bandwidth (-3 dB)
10Hz to 3MHz
XSA1000 Series
Spectrum Analyzer
+ Frequency Range from 9 kHz up to1.5 GHz
+ 150dBm Displayed Average Noise Level
+ Phase Noise -85dBc/Hz @1Gz and offset at 10KHz
+ Total Amplitude Accuracy <1.5dB
+ 10Hz Minimum Resolution Bandwidth (RBW)
+ EMI Pre-compliance Test Kit
+ Up to 1.5 GHz Tracking Generator Kit (Opt.)
+ 10.4 inches display
.……..….…...…..…..…..…..……………………………………………………………………………………………………………
…...………………..……………………………………………………………………………………………………………………………………………
Fujian Lilliput Optoelectronics Technology Co., Ltd
Web: www.owon.com.cn /www.owontme.com E-mail: info@owon.com.cn Facebook: fb.me/owontech
Amplitude Specification
Amplitude and electric level
Amplitude measurement range
DANL to +20 dBm, close the preamplifier
Reference electric level
-80 dBm to +30 dBm, 0.1dBm steps
Preamplifier
20 dB, nominal, 9 kHz~1.5 GHz
Input attenuator range
0~39 dB, 3 dB steps
Max input DC voltage
50 VDC
Max continuous power
30dBm, average continuous power
Displayed average noise level (DANL)
Input attenuation 0 dB, 1Hz resolution bandwidth
Preamp off
1 MHz~10 MHz -130dBm (typical)
10 MHz~1GHz -130dBm (typical)
1GHz~1.5 GHz -128 dBm (typical)
Preamp on
1 MHz~10 MHz -150dBm (typical)
10 MHz~1GHz -150dBm (typical)
1GHz~1.5 GHz -148 dBm (typical)
Phase noise
20 ~30 ℃,fc=1 GHz
Phase noise
<-85 dBc/Hz @10 kHz offset
<-100 dBc/Hz @100 kHz offset
<-110 dBc/Hz @1 MHz offset
Level display range
Log scale coordinate
1dB ~255dB
Linear scale coordinate
0 to reference level
level unit
dBm, dBuW, dBpW, dBmV, dBuV, W,V
Points
201~1001
Number of traces
5
Detectors
Positive-peak, negative-peak, sample, normal, RMS
Trace functions
Clear write, Max Hold, Min Hold, View, Blank, Average
Frequency response
20℃ ~30℃, 30%~70% relative humidity, 20 dB input attenuation, reference 50 MHz
Preamp off
±0.8 dB;
Preamp on
±0.9 dB;
Accuracy
Input Attenuation Switching
Uncertainty
20℃ - 30℃, fc=50 MHz, Preamplifier Off, 20dB RF attenuation, input signal 0~39 dB
±0.5 dB
Absolute Amplitude Uncertainty
20℃ - 30℃, fc=50 MHz, RBW=1 kHz, VBW=1 kHz, peak detector, 20 dB RF attenuation,
Preamplifier Off ±0.4 dB, input signal= -20dBm
Preamplifier On ±0.5 dB, input signal= -40dBm
.……..….…...…..…..…..…..……………………………………………………………………………………………………………
…...………………..……………………………………………………………………………………………………………………………………………
Fujian Lilliput Optoelectronics Technology Co., Ltd
Web: www.owon.com.cn /www.owontme.com E-mail: info@owon.com.cn Facebook: fb.me/owontech
Uncertainty
input signal range 0dbm~-50dbm
±1.5 dB
VSWR
input 10 dB RF attenuation, 1 MHz~1.5GHz
<1.5, nominal
Distortion and spurious response
Second harmonic distortion
fc 50 MHz, Preamp off, signal input -30 dBm, 0 dB RF attenuation, 20 to 30
-60dbc
Third-order intermodulation
fc 50 MHz
+13 dBm
1 dB Gain Compression
fc 50 MHz, 0 dB RF attenuation, Preamp off , 20 to 30
+7 dBm, nominal
Residual response
connect 50 Ω load at input port, 0 dB input attenuation, 20 to 30
<-85dBm, typical
Input related spurious
-30 dBm signal at input mixer, 20 to 30
<-60 dBc
Sweep time and triggering
Span range
100Hz≤SPAN≤3GHz 10ms to 3000s
zero sweep width 1ms to 3000s
Mode
Continue, single
Trigger
Free run, video, external
Tracking generator
Output frequency range
100 kHz~1.5 GHz
Output power level range
-30 dBm~0 dBm ,
Output power level resolution
1DB
Output flatness
+/-3 dB
Maximum safe reverse level
Average total power:30 dBm, DC : ±50 VDC
Inputs and Outputs
Front panel RF input connector
50 Ω, N-type female
Front panel track generator output
50 Ω, N-type female
10 M reference input
50 Ω, N-type female
Communication port
USB HOST, USB DEVICE, LAN, earphone port, and VGA
General techincal specification
Display
TFT LCD, 10.4 inches
Weight
5 kg
Working temperature
0~40
Storage temperature
-20 °C to +60 °C
Power
100V~240V 50/60Hz