Fluke 8808A
Digital Multimeter
Extended Specifications
Making measurements is
as simple as pushing a
button
The Fluke 8808A 5.5 digit
multimeter has a broad range
of functions, measuring volts,
ohms and amps with a basic
V dc accuracy of 0.01 %. It is
remarkably easy to use, even by
unskilled operators, because it
makes the measurements you
perform most often extremely
easy and fast to do.
Six setup buttons on the
8808A front panel operate like
a car radio’s station presets.
Simply set up the meter for a
common measurement, then
press shift followed by a setup
button (S1 to S6) to save the
setup. Now each time you
perform that measurement, you
simply press the appropriate
setup key. It’s that easy!
The setup buttons eliminate
the need to follow complex work
instruction sheets. Operators no
longer need to press multiple
buttons to set up a measurement
function and range, test limits, or
enter other parameters to make
a measurement.
Eliminate production
mistakes
The Fluke 8808A 5.5 digit
multimeter dependably performs
the most common measurements
required by today’s applications.
Features at a glance
5.5 digit resolution
Basic V dc accuracy of 0.01 %
Dual display
Dedicated dc leakage current measurement
2x4 ohms 4-wire measurement technique
Six dedicated buttons for fast access to instrument setups
Hi.Lo limit compare for Pass/Fail testing
Fluke 45 remote command emulation
Whether you are performing
functional tests or making
critical measurements on test
points, using the limit compare
mode with pass/fail indicators
eliminates production mistakes,
especially those where results
are “on the edge.”
The 8808A display has
built-in enunciators that clearly
show the operator whether a
test passes or fails. The pass/fail
indicators take the guesswork
out of testing: the result is either
within limits or it’s out.
2 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
2 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
General Specifications
Voltage
100V Setting ..............................................................90 V to 110 V
120V Setting .............................................................108 V to 132 V
220V Setting .............................................................198 V to 242 V
240V Setting .............................................................216 V to 264 V
Frequency ...................................................................47 Hz to 440 Hz
Power Consumption..................................................15 VA peak (10 W average)
Dimensions
Height..........................................................................88 mm (3.46 in)
Width...........................................................................217 mm (8.56 in)
Depth ..........................................................................297 mm (11.7 in)
Weight.........................................................................2.1 kg (4.6 Ib)
Display
Vacuum Fluorescent Display, segment
Environment
Temperature
Operating................................................................0 rC to 50 rC
Storage....................................................................-40 rC to 70 rC
Warm Up.................................................................½ hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating................................................................Uncontrolled (< 10qC)
<90 % (10 qC to 28 qC)
<75 % (28 qC to 40 qC)
<45 % (40 qC to 50 qC)
Storage ...................................................................-40 rC to 70 rC <95 %
Altitude
Operating ...............................................................2,000 Meters
Storage ........................................................................12,000 Meters
Vibration .....................................................................Complies with MIL-PRF-28800F Class 3
Safety
Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V
EMC
Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
Triggering
Trigger Delay ............................................................400 ms
External Trigger Delay .............................................<2 ms
External Trigger Jitter ..............................................<1 ms
Trigger Input .............................................................TTL Levels
Trigger Output............................................................5 V max
Math Functions
Min/max, relative, hold, compare and dB functions
Electrical
Input Protection ........................................................1000 V all ranges
Overrange ...................................................................10 % on the largest ranges of all functions except continuity and diode
test
Remote Interfaces
RS-232C
Warranty
One year
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
Electrical Specifications
Specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
Maximum Input ..........................................................1000 V on any range
Common Mode Rejection..........................................120 dB at 50 or 60 Hz r0.1 (1 k: unbalance)
Normal Mode Rejection.............................................80 dB at Slow Rate
A/D Nonlinearity ........................................................15 ppm of range
Input Bias Current .....................................................<30 pA at 25 rC
Settling Considerations.............................................Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast Input Impedance
200 mV 199.999 mV 1 PV 10 PV 10 PV >10 G:[1]
2 V 1.99999 V 10 PV 100 PV 100 PV >10 G:[1]
20 V 19.9999 V 100 PV 1000 PV 1000 PV 10 M:r1 %
200 V 199.999 V 1 mV 10 mV 10 mV 10 M:r1 %
1000 V 1000.00 V 10 mV 100 mV 100 mV 10 M:r1 %
Notes:
[1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M:.
Uncertainty [1]
90 days 1 year
Range
23 °C r 5 °C 23 °C r 5°C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005
2 V 0.01 + 0.002 0.015 + 0.003 0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004 0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
Notes:
[1] Uncertainty given as r (% of reading + % of range)
4 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
4 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
AC Voltage Specifications
AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range, and for 50 kHz to 100 kHz, add 0.13 % of range.
Maximum Input ..........................................................750 V rms or 1000 V peak or 8 x 107 Volts-Hertz product
Measurement Method ...............................................AC-coupled true-rms. Measures the ac component of input with up to
1000 V dc bias on any range.
AC Filter Bandwidth..................................................20 Hz – 100 kHz
Common Mode Rejection..........................................60 dB at 50 Hz or 60 Hz (1 k: unbalance)
Maximum Crest Factor .............................................3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) ..............Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
Only applies for non-sinusoid signals
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Input Impedance
200 mV 199.999 mV 1 PV 10 PV 10 PV
2 V 1.99999 V 10 PV 100 PV 100 PV
20 V 19.9999 V 100 PV 1000 PV 1000 PV
200 V 199.999 V 1 mV 10 mV 10 mV
750 V 750.00 V 10 mV 100 mV 100 mV
1 M: r2 % shunted by
<100 pf
Uncertainty [1]
90 days 1 year
Range Frequency
23 °C r 5 °C 23 °C r 5 °C
Temperature
Coefficient/°C
Outside 18 – 28 °C
200 mV 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
2 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
20 V 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
200 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
750 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
Notes:
[1] Uncertainty given as r (% of reading + % of range)
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 5
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 5
Resistance
Specifications are for 4-wire resistance function, or 2-wire resistance with REL. If REL is not used, add 0.2 : for 2-wire
resistance plus lead resistance.
Measurement Method ...............................................Current source referenced to LO input
Max Lead Resistance (4-wire ohms) ......................10 % of range per lead for 200 :, 2 k: ranges. 1 k: per lead on all
other ranges.
Input Protection .........................................................1000 V on all ranges
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Current Source
200 : 199.999 : 0.001 : 0.01 : 0.01 : 0.8 mA
2 k: 1.99999 k: 0.01 : 0.1 : 0.1 : 0.8 mA
20 k: 19.9999 k: 0.1 : 1 : 1 : 0.08 mA
200 k: 199.999 k: 1 : 10 : 10 : 0.008 mA
2 M: 1.99999 M: 10 : 100 : 100 : 0.9 PA
20 M: 19.9999 M: 100 : 1 k: 1 k: 0.16 PA
100 M: 100.000 M: 1 k: 10 k: 10 k: 0.16 PA || 10 M:
Uncertainty [1]
90 days 1 year
Range
23 °C r 5 °C 23 °C r 5 °C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 : 0.02 + 0.004 0.03 + 0.004 0.003 + 0.0006
2 k: 0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
20 k: 0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
200 k: 0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
2 M: 0.03 + 0.003 0.04 + 0.004 0.004 + 0.0005
20 M: 0.2 + 0.003 0.25 + 0.003 0.01 + 0.0005
100 M: 1.5 + 0.004 1.75 + 0.004 0.2 + 0.0005
Notes:
[1] Uncertainty given as r (% of reading + % of range)
DC Current
Input Protection ........................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses.
Shunt Resistance .......................................................0.01 : for 2 A and 10 A ranges
1 : for 20 mA and 200 mA
Burden voltage < 5 mV for 200 PA and 2 mA range.
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Burden Voltage
200 PA 199.999 PA 0.001 PA 0.01 PA 0.01 PA <5 mV
2 mA 1999.99 PA 0.01 PA 0.1 PA 0.1 PA <5 mV
20 mA 19.9999 mA 0.1 PA 1 PA 1 PA <0.05 V
200 mA 199.999 mA 1 PA 10 PA 10 PA <0.5 V
2 A 1.99999 A 10 PA 100 PA 100 PA <0.1 V
10 A 10.0000 A 100 PA 1 mA 1 mA <0.5 V
6 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
6 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
Uncertainty [1]
90 days 1 year
Range
23 °C r 5 °C 23 °C r 5 °C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 PA 0.02 + 0.005 0.03 + 0.005 0.003 + 0.001
2 mA 0.015 + 0.005 0.02 + 0.005 0.002 + 0.001
20 mA 0.03 + 0.02 0.04 + 0.02 0.005 + 0.001
200 mA 0.02 + 0.005 0.03 + 0.008 0.005 + 0.001
2 A 0.05 + 0.02 0.08 + 0.02 0.008 + 0.001
10 A 0.18 + 0.01 0.2 + 0.01 0.008 + 0.001
Notes:
[1] Uncertainty given as r (% of reading + % of range)
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs
from 1 % to 5 % of range, add an additional error of 0.1 % of range.
Input Protection .........................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses
Measurement Method ...............................................AC-coupled True RMS
Shunt Resistance .......................................................0.01 : for 2 A and 10 A ranges
1 : for 20 mA and 200 mA
AC Filter Bandwidth..................................................20 Hz – 100 kHz
Maximum Crest Factor ..............................................3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) ..............Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
Only applies to non-sinusoid signals
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Burden Voltage
20 mA 19.9999 mA 0.1 PA 1 PA 1 PA <0.05 V
200 mA 199.999 mA 1 PA 10 PA 10 PA <0.5 V
2 A 1.99999 A 10 PA 100 PA 100 PA <0.1 V
10 A 10.0000 A 100 PA 1 mA 1 mA <0.5 V
Uncertainty [1]
90 days 1 year
Range Frequency
23 °C r 5 °C 23 °C r 5 °C
Temperature
Coefficient/°C
Outside 18 – 28 °C
20 mA 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
200 mA 20 Hz - 45Hz 0.8 + 0.05 1 + 0. 06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
2 A 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
10 A 20 Hz - 45Hz 1 + 0.1 1.25 + 0.12 0.015 + 0.005
45 Hz - 2 kHz 0.35 + 0.1 0.5 + 0.12 0.015 + 0.005
Notes:
[1] Uncertainty given as r (% of reading + % of range)
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 7
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 7
Frequency
Gate Time ...................................................................131 ms
Measurement Method ...............................................AC-coupled input using the ac voltage measurement function.
Settling Considerations.............................................When measuring frequency after a dc offset voltage change, errors
may occur. For the most accurate measurement, wait up to 1 second to
allow input blocking RC time constant to settle.
Measurement Considerations ..................................To minimize measurement errors, shield inputs from external noise
when measuring low voltage, low frequency signals.
Uncertainty
90 days 1 year
Range Frequency
23 °C r 5 °C 23 °C r 5 °C
Temperature
Coefficient/°C
Outside 18 – 28 °C
20 Hz – 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz – 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz – 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
100 mV to
750 V[1,2]
200 kHz – 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
Notes:
[1] Input > 100 mV
[2] Limited to 8* 107 V Hz
Continuity
Continuity Threshold ................................................20 :
Test Currents ..............................................................1 mA
Response Time .........................................................100 samples/sec with audible tone
Rate..............................................................................Fast
Maximum Reading.....................................................199.99 :
Resolution...................................................................0.01 :
Diode Test
Response Time...........................................................100 samples/sec with audible tone
Rate..............................................................................Fast
Maximum Reading.....................................................1.9999 V
Resolution...................................................................0.1 mV
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2007-2008 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 5/2008 3053965 D-EN-N Rev C
Pub_ID: 11288-eng Rev 02
Fluke. Keeping your world
up and running.®
Ordering information
Models Description
8808A 120V 5.5 Digit Multimeter
8808A 220V 5.5 Digit Multimeter
8808A 100V 5.5 Digit Multimeter
8808A 240V 5.5 Digit Multimeter
8808A/SU includes
8808A package plus, FlukeView Forms basic software, USB to RS-232
interface adapter cable.
8808A/SU 120V 5.5 Digit Multimeter, SW USB Cable Kit
8808A/SU 220V 5.5 Digit Multimeter, SW USB Cable Kit
8808A/SU 100V 5.5 Digit Multimeter, SW USB Cable Kit
8808A/SU 240V 5.5 Digit Multimeter, SW USB Cable Kit
8808A includes
Meter, TL71 test leads, line cord, spare line fuse, statement of cal prac-
tices, WEEE information sheet, Warranty statement, Getting Started
guide (English, French, German, Spanish, Italian, Simplified Chinese,
Japanese), CD Rom with user manual (English).
8 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications