Fluke 116 HVAC Multimeter with Temperature and Microamps Technical Data Compact true-rms meter for HVAC troubleshooting The Fluke 116 was specifically designed for the HVAC professional. It has everything needed in an HVAC meter including temperature and microamp measurements to quickly troubleshoot problems with HVAC equipment and flame sensors. Features include: * Built in thermometer for HVAC applications * Microamps to test flame sensors * LoZ: helps prevent false readings due to ghost voltage * Large white LED backlight to work in poorly lit areas * Resistance, continuity, frequency and capacitance * Min/Max/Average with elapsed time to record signal fluctuations * Compact ergonomic design for one-handed operation * Compatible with optional magnetic hanger (ToolPakTM) * CAT III 600 V safety rated General specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 C to 28 C, with relative humidity at 0 % to 90 %. The accuracy specifications take the form of: ( [ % of reading ] + [ counts ] ) Maximum voltage between any terminal and earth ground Surge protection Actual size Display Bar graph Operating temperature Storage temperature Battery Battery life R N10140 600 V 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 Digital: 6,000 counts, updates 4/sec 33 segments, updates 32/sec -10 C to + 50 C -40 C to + 60 C 9 volt Alkaline, NEDA 1604A/ IEC 6LR61 400 hours typical, without backlight Accuracy specifications Measurement Range Resolution Accuracy ([% of reading] + [counts]) DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 Ordering information 6.000 V 0.001 V 0.5 % + 2 Fluke-116 DC volts 60.00 V 0.01 V 600.0 V 0.1 V Auto volts 600.0 V 0.1 V 2.0 % + 3 (dc, 45 Hz to 500 Hz) 4.0 % + 3 (500 Hz to 1 kHz) AC millivolts1 true-rms 600.0 mV 0.1 mV 1.0 % + 3 (dc, 45 Hz to 500 Hz) 2.0 % + 3 (500 Hz to 1 kHz) AC volts1 true-rms Continuity Ohms 6.000 V 0.001 V 60.00 V 0.01 V 600.0 V 0.1 V 600 1 Beeper on < 20 off > 250 ; detects opens or shorts of 500 s or longer. 0.9 % + 2 600.0 0.1 6.000 k 0.001 k 60.00 k 0.01 k 600.0 k 0.1 k 6.000 M 0.001 M 0.9 % + 1 40.00 M 0.01 M 1.5 % + 2 2.000 V 0.001 V 0.9 % + 2 Capacitance 1000 nF 1 nF 10.00 F 0.01 F 100.0 F 0.1 F 9999 F 1 F 1.9 % + 2 100 F to 1000 F 1.9 % + 2 > 1000 F 5 % + 20 Lo-Z capacitance 1 nF to 500 F Temperature2 (Type-K thermocouple) -40 C to 400 C 0.1 C 1 % + 102 -40 F to 752 F 0.2 F 1 % + 182 10 % + 2 typical AC amps true-rms (45 Hz to 500 Hz) 600.0 A DC amps 600.0 A 0.1 A 99.99 Hz 0.01 Hz 999.9 Hz 0.1 Hz 9.999 kHz 0.001 kHz 50.00 kHz 0.01 kHz Hz (V or A Included TL75 Test Leads, 80BK Integrated Temperature Probe, holster, User's manual and 9 V battery (installed). 1.0 % + 3 (45 Hz to 500 Hz) 2.0 % + 3 (500 Hz to 1 kHz) Diode test input)3 HVAC Multimeter with Temperature and Microamps 0.1 A 1.5 % + 3 (2.5 % + 3 > 500 Hz) 1.0 % + 2 0.1 % + 2 Notes: 1All ac voltage ranges except Auto-V/LoZ are specified from 1 % to 100 % of range. Auto-V/LoZ is specified from 0.0 V. Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display nonzero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. AC volts is ac-coupled. Auto-V LoZ, and ac mV are dc-coupled. 2Temperature uncertainty (accuracy) does not include the error of the thermocouple probe. 3Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage. Frequency is dc coupled, 45 Hz to 5 kHz for ac current. Fluke Corporation 116 HVAC Multimeter with Temperature and Microamps Fluke.Keeping your world up and running.TM Fluke Corporation PO Box 9090, Everett, WA USA 98206 Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa (31 40) 2 675 200 or Fax (31 40) 2 675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com (c)2006 Fluke Corporation. All rights reserved. Specifications subject to change without notice. Printed in U.S.A. 8/2006 2634037 D-EN-N Rev B