PP019 and PP020 Passive Probes Operator’s Manual
© 2017 Teledyne LeCroy, Inc. All rights reserved.
Unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited.
Customers are permitted to duplicate and distribute Teledyne LeCroy documentation for
internal educational purposes.
Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Other product or brand
names are trademarks or requested trademarks of their respective holders. Information in this
publication supersedes all earlier versions. Specifications subject to change without notice.
This electronic product is subject to disposal and recycling regulations
that vary by country and region. Many countries prohibit the disposal of
waste electronic equipment in standard waste receptacles. For more
information about proper disposal and recycling of your Teledyne LeCroy
product, visit teledynelecroy.com/recycle.
928605-00 Rev A
July, 2017
Operator’s Manual
1
Terms and Symbols
The following symbols appear on the product or in its documentation:
WARNING
High Voltage. Risk of electric shock or burn.
WARNING or
CAUTION
Condition or practice could result in injury or loss
of life, or damage to equipment.
Introduction
Teledyne LeCroy PP019 and PP020 are passive high impedance voltage
oscilloscope probes designed and calibrated for use with general purpose
oscilloscopes having an input impedance of 1 MΩ shunted by 16 pF.
However, they may be compensated for use with instruments having an
input capacitance from 10 to 30 pF for PP019, and 8 to 20 pF for PP020.
They are compatible with oscilloscopes that automatically detect probe
attenuation and adjust their readout accordingly.
To avoid electric shock, keep fingers behind the probe’s finger
guard during use.
Rated for indoor use only. Do not operate where conductive
pollutants may be present.
PP019/PP020 Passive Probe
2
Specifications
Model Attn.
Input
Impedance B/W
Rise
Time Length
Compensation
Range
R (MΩ) C (pF) (MHz) (ns) (m) (pF)
PP019
10x 10 12 250 1.4 1.2 10 - 30
PP020
10x 10 11 500 0.7 1.2 8 - 20
Attenuation Ratio .................. 10:1 ±1%
Max. Input Voltage ....... Circuits not directly connected to Mains: 500 Vrms,
......................................... 1500 V transient overvoltage
......................................... Measurement Category II*: 400 Vrms
To reduce risk of fire or electric shock, do not exceed the voltage or
category ratings of the probe. See voltage derating curve on p.4.
Pollution Degree ..................... Pollution Degree 2*
Operating Altitude ................. Up to 3000 meters
Max. Operating Temp ......... 0 to 50 °C
Humidity .................................... 5% to 95% RH (10 to 30 °C)
5% to 75% RH (30 to 40 °C)
5% to 50% RH (above 40 °C)
RH not controlled below 10 °C
* As defined in IEC/EN 61010-031. See definitions below.
Definitions:
Measurement Category II (CAT II) refers to measurements performed on circuits directly connected to
utilization points (socket outlets and similar) of the low-voltage mains installation.
Pollution Degree 2 refers to an operation environment where normally only dry non-conductive
pollution occurs. Temporary conductivity caused by condensation must be expected.
Operator’s Manual
3
PP019 Input Impedance Profile
PP020 Input Impedance Profile
PP019/PP020 Passive Probe
4
Voltage vs. Frequency Derating Curve
EC Declaration of Conformity
The product conforms to the applicable European Union
requirements per Low Voltage Directive (LVD) 2014/35/EU.
Compliance was demonstrated to the following specification as listed in
the Official Journal of the European Communities:
IEC/EN 61010-031:2015 Safety requirements for electrical equipment for
measurement, control and laboratory use Part 031: Safety requirements for
handheld probe assemblies for electrical measurement and test.
EU RoHS Compliance
The probe and accessories conform to the 2011/65/EU RoHS2 Directive.
Disposal
The probe complies with the applicable European Union
requirements to Directives 2012/19/EU and 2013/56/EU on Waste
Electrical and Electronic Equipment (WEEE) and Batteries.
Operator’s Manual
5
General Safety Information
Observe generally accepted safety procedures in addition to those
listed here to avoid personal injury or damage to equipment. The
overall safety of any system incorporating this accessory is the
responsibility of the assembler of the system.
Connect only to grounded instruments. Use only with compatible Teledyne
LeCroy oscilloscopes that have their BNC input connected to an earth
ground. Do not connect the probe reference lead to any point which is at a
potential other than earth ground.
Connect and disconnect properly. Connect probe to the oscilloscope before
connecting it to the test circuit. Disconnect the probe input and reference
lead from the test circuit before disconnecting from the oscilloscope. Do
not connect/disconnect probes while connected to a voltage source.
Do not overload. Do not apply any potential to the probe leads that exceeds
the maximum rating of the probe. Observe all terminal ratings on the
oscilloscope before connecting. Consult the oscilloscope product manual
for further ratings information.
Always comply with the Voltage vs. Frequency Derating Curve.
Do not remove probe casing. Touching exposed connections may result in
electric shock.
Use indoors only within operational environment listed. Do not use in wet or
explosive atmospheres. Keep product surfaces clean and dry.
Handle with care. Probe tips are extremely sharp and may puncture skin or
cause other bodily injury if not handled properly.
Keep fingers behind the finger guard of probe body and accessories.
PP019/PP020 Passive Probe
6
Do not operate with suspected failures. Before each use, inspect the probe
and accessories for any potential damage such as tears or other defects in
the probe body, cable jacket, accessories, etc. If any part is damaged, cease
operation immediately and sequester the probe from inadvertent use.
Cleaning
Clean only the exterior of the probe, cables, and accessories. Use a
soft cotton cloth moistened with a mild detergent and water solution.
Do not allow any portion of the probe to be submerged at any time.
Dry the probe and accessory thoroughly before attempting to make
any voltage measurement.
Do not subject the probe to solvents or solvent fumes as these can
cause deterioration of the probe body, cables, and accessories.
Service
Refer all repairs to qualified service personnel. Contact Teledyne LeCroy at
the following address to arrange to return the probe for service:
Teledyne LeCroy, Inc.
700 Chestnut Ridge Road
Chestnut Ridge, NY, 10977-6499
Ph: 800-553-2769 or 845-425-2000
Fax: 845-578-5985
contact.corp@teledynelecroy.com
For the most complete and up-to-date list of sales and service centers by
country, visit teledynelecroy.com/support/contact.
Operator’s Manual
7
Probe Compensation
Proper compensation of the probe is required to assure amplitude accuracy
of the waveform being measured by matching the probe to the
oscilloscope’s input capacitance. Compensation should be adjusted
whenever the probe is connected to or transferred between oscilloscopes.
Low Frequency (PP019 and PP020)
1. Apply a 1 kHz square wave to the probe or connect to the
oscilloscope’s calibrator output.
2. Adjust the single LF trimmer on the PP019/20 BNC Compensation
Box until you achieve a flat-topped square wave as in the center of the
second illustration below.
LF trimmer on PP019 and PP020 BNC Compensation Box.
Over-compensated, correct, and under-compensated waveforms
PP019/PP020 Passive Probe
8
High Frequency (PP020 Only)
High frequency should rarely require adjustment. However, if adjustment is
required:
1. Apply a 1 MHz square wave to the probe (< 0.7 ns rise-time).
2. Remove the plastic caps from the two, HF trimmers on the PP020
BNC Compensation Box.
HF trimmers on PP020.
3. Adjust HF2 then HF1 until you achieve a flat-topped square wave as in
the center illustration below.
Over-compensated, correct, and under-compensated waveforms
Operator’s Manual
9
Replaceable Parts
Item
Description
Quantity
1
Sprung Hook, 5 mm, Black
1
2
GND Lead w/Alligator Clip
1
3
BNC Adapter, 5 mm
1
4
Replacement Tip, Black
1
5
Deluxe Trimmer Tool
1
6
IC Tip Insulator, 5 mm, Black
1
7
Tip Insulator, 5 mm, Black
1
8
Identifier Rings
1
9
Probe Tip GND
1
Standard Accessory Replacement Kit part number is PKIT4-5MM-101.
PP019/PP020 Passive Probe
10
Warranty
NOTE: THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED
OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF
MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR
USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT OR OTHERWISE. THE
CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION AND INSURANCE
CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY. TELEDYNE
LECROY WILL RETURN ALL PRODUCTS UNDER WARRANTY WITH TRANSPORT
PREPAID.
Teledyne LeCroy warrants this oscilloscope accessory for normal use and operation
within specification for a period of one year from the date of shipment. Spare parts,
replacement parts and repairs are warranted for 90 days.
In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace
any assembly returned within the warranty period to the Customer Service
Department of an authorized service center. However, this will be done only if the
product is determined by Teledyne LeCroy’s examination to be defective because of
workmanship or materials, and the defect is not caused by misuse, neglect, accident,
abnormal conditions of operation; or, damaged by attempted repair or modifications
by a non-authorized service facility.
The customer will be responsible for the transportation and insurance charges for the
return of products to the service facility. Teledyne LeCroy will return all products under
warranty with transportation charges prepaid.
This warranty replaces all other warranties, expressed or implied, including, but not
limited to, any implied warranty of merchantability, fitness, or adequacy for any
particular purposes or use. Teledyne LeCroy shall not be liable for any special,
incidental, or consequential damages, whether in contract or otherwise.