MIT1525 < xxx ms/uF to discharge
from 15000 V to 50 V
Capacitance range 10 nF to 25 μF (dependant on
(above 500 V): measurement voltage)
Capacitance accuracy (23 °C): ±10% ±5 nF
Voltage output accuracy +4%, -0%, ±10 V nominal test
voltage (>200 V, 0 °C to 30 °C):
at 1 GΩ
Current measurement range: 0.01 nA to 6 mA
Current measurement
accuracy (23 °C): ±5% ±0.2 nA at all voltages
Interference (noise) rejection:
MIT515, MIT525 1 mA per 250 V up to a maximum of
3 mA
MIT1025 1 mA per 600 V up a maximum of
3 mA
MIT1525 4 mA
Voltmeter range: 30 V to 660 V ac or dc, 50/60 Hz
Voltmeter accuracy: ±3%, ±3 V
Timer range: Up to 99 minutes, 15 second
minimum setting
Memory capacity:
5½ hours continuous logging every 5
s.
or 33 logged PI tests
or 350 logged IR tests
MIT1525 11 hours continuous
logging every 5 s.
Test regimes:
MIT515 IR, IR(t), DAR, PI
MIT525, MIT1025 IR, IR(t), DAR, PI, SV, DD, ramp test
Interface: USB type B (device)
Real time output: USB, 1 reading/second (voltage,
current and resistance)
ENVIROMENTAL CONDITIONS
Altitude: 3000 m, CAT rating maintained
>2000 m*
* Test leads connected
4000 m (MIT1525 only)
Operating temperature -20 °C to 50 °C
Storage temperature: -25 °C to 65 °C
Humidity: 90% RH non-condensing at 40 °C
Ingress protection: IP65 (lid closed), IP40 (lid open)
GENERAL SPECIFICATION
Safety: Meets the requirements of IEC 61010-1, CATIV 600 V
EMC: Meets the requirements of IEC61326-1
Dimensions
MIT515, MIT525, MIT1025: L 315 mm x W 285 mm x H 181 mm
Weight: 4.5 kg
MIT1525: 360 mm x W 305 mm x H 194 mm
Weight: 6.5 kg
“spot test”, a 1 minute IR test and a 10 minute Polarisation
Index (PI) test, where PI is the ratio R10min/R1min and
temperature independent.
ELECTRICAL SPECIFICATIONS
Voltage input range: 85-265 V rms, 50/60 Hz, 60 VA
Battery: 11.1 V, 5.2 A hour, meets IEC
62133:2003
Battery life
MIT515, MIT525: Typical capacity 6 hours continuous
at 5 kV with a 100 MΩ load
MIT1025: Typical capacity 4.5 hours
continuous at 10 kV with a 100 MΩ
load
Battery charge time: 2.5 hours from deep discharge,
2 hours normal discharge
MIT1525: Typical capacity 6 hours continuous
at 15 kV with a 100 MΩ load
30 min. quick charge: 1 hour operation at 5 kV, 100 MΩ
Test voltages
MIT515, MIT525: 250 V, 500 V, 1000 V, 2500 V,
5000 V
MIT1025: 500 V, 1000 V, 2500 V, 5000 V,
10000 V
User defined test voltage:
MIT515, MIT525 100 V to 1 kV in 10 V steps,
1 kV to 5 kV in 25 V steps,
MIT1525: 1000 V, 2500 V, 5000 V,
10000 V, 15000 V
MIT1025 5 kV to 10 kV in 25 V steps
MIT1525 10 kV to 15 kV in 25 V steps
Accuracy (23 °C):
MIT515, MIT525 accuracy (23 °C)
5000 V 2500 V 1000 V 500 V 250 V
±5% to 1 TΩ 500 GΩ 200 GΩ 100 GΩ 50 GΩ
±20% to 10 TΩ 5 TΩ 2 TΩ 1 TΩ 500 GΩ
MIT1025 accuracy (23 ºC)
15 kV 10 k V 2500 V 5000 V 1000 V
±5% to 3 TΩ 2 TΩ 1 TΩ 500 GΩ 200 GΩ
±20% to 40 TΩ 20 TΩ 10 TΩ 2 TΩ 1 TΩ
Guard : 2% error guarding 500 kΩ leakage
with 100 MΩ load
Display range analogue: 100 kΩ to 10 TΩ
Display range digital:
MIT515, MIT525 10 kΩ to 10 TΩ
MIT1025 10 kΩ to 20 TΩ
Short circuit current: 3 mA nominal, max. power on all
loads outperforming many 5 mA
testers
Insulation alarm: 100 kΩ to 1 GΩ
Capacitor charge:
MIT515, MIT525 <3 s/μF at 3 mA to 5 kV
MIT1025 <5 s/μF at 3 mA to 10 kV
MIT1525 < x s/uF at 3 mA to 15 kV
Capacitor discharge:
MIT515, MIT525 <250 ms/μF to discharge from
5000 V to 50 V
MIT1025 <500 ms/μF to discharge from
10000 V to 50 V
MIT515, MIT525, MIT1025, MIT1525
5, 10 & 15 kV DC Insulation Resistance Testers