Passive Component LCR Meter Measures Inductance, Capacitance, and Resistance With secondary parameter Q (Quality), D (Dissipation), R (Resistance) Features: * Simultaneous 20,000/10,000 count backlit display of the primary parameter (L, C or R) with the secondary parameter of Q (quality), D (dissipation) or R (resistance) * Set Hi/Lo limits using absolute values or percentage limits * Parallel or Series equivalent circuit * Max/Min/Avg; Auto power off with disable * Automatic blown fuse indication * Built-in test fixture or use external test leads * Open and Short Zero removes unwanted stray impedances from the measurement * Relative mode, zero, or supplied reference * Complete with test leads, 9V battery, software, cable, and protective holster Built-in RS-232 with Windows(R) 95/98/NT/2000/XP compatible software. Specifications Inductance Capacitance Resistance DF (with C) Q Test Frequency Dimensions Weight Ranges 2000.0H, 20.000mH, 200.00mH, 2000.0mH 20.000H, 200.00H, 2000.0H, 10000H 2000.0pF, 20.000nF, 200.00nF, 2000.0nF 20.000F, 200.00F, 2000.0F, 10.000mF 20.000, 200.00, 2.0000k, 20.000k, 200.00k, 2.0000M, 10.000M 0 to 0.9999 0 to 9999 120Hz and 1kHz 7.56 x 3.58 x 2.1" (19.2 x 9.1 x 5.25cm) 12.87oz (365g) Basic Accuracy (0.7%rdg + Lx/10000 + 5 digit) (DF<0.5) (0.7%rdg + 5 digit) (DF<0.1) (0.5%rdg + 8 digit) (0.7%rdg + 100/Cx + 5 digit) (1.2%rdg + 100/Lx + 5 digit) Ordering Information: 380193 ........Passive Component LCR Meter 153117 ........117VAC adaptor 153220 ........230VAC adaptor www.extech.com Specifications subject to change without notice. 5/24/10 - R1 Copyright (c) 2007-2008 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form.