Specifications subject to change without notice.
Copyright © 2007-2008 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form.
www.extech.com 5/24/10 - R1
Measures Inductance, Capacitance, and Resistance
With secondary parameter Q (Quality), D (Dissipation), R (Resistance)
Passive Component LCR Meter
Ordering Information:
380193 ........Passive Component LCR Meter
153117 ........117VAC adaptor
153220 ........230VAC adaptor
Features:
•Simultaneous 20,000/10,000 count
backlit display of the primary parameter
(L, C or R) with the secondary parameter of
Q (quality), D (dissipation) or R (resistance)
•Set Hi/Lo limits using absolute values or
percentage limits
•Parallel or Series equivalent circuit
•Max/Min/Avg; Auto power off with disable
•Automatic blown fuse indication
•Built-in test fixture or use external test leads
•Open and Short Zero removes unwanted stray
impedances from the measurement
•Relative mode, zero, or supplied reference
•Complete with test leads, 9V battery, software,
cable, and protective holster
Built-in RS-232 with
Windows®95/98/NT/2000/XP
compatible software.
Specifications Ranges Basic Accuracy
Inductance 2000.0μH, 20.000mH, 200.00mH, 2000.0mH ±(0.7%rdg + Lx/10000 + 5 digit)
20.000H, 200.00H, 2000.0H, 10000H (DF<0.5)
Capacitance 2000.0pF, 20.000nF, 200.00nF, 2000.0nF ±(0.7%rdg + 5 digit)
20.000μF, 200.00μF, 2000.0μF, 10.000mF (DF<0.1)
Resistance 20.000Ω, 200.00Ω, 2.0000kΩ, 20.000kΩ, ±(0.5%rdg + 8 digit)
200.00kΩ, 2.0000MΩ, 10.000MΩ
DF (with C) 0 to 0.9999 ±(0.7%rdg + 100/Cx + 5 digit)
Q 0 to 9999 ±(1.2%rdg + 100/Lx + 5 digit)
Test Frequency 120Hz and 1kHz
Dimensions 7.56 x 3.58 x 2.1” (19.2 x 9.1 x 5.25cm)
Weight 12.87oz (365g)