Fluke 114, 115, 116 and 117 Digital Multimeters Extended specifications Technical Data General specifications (all models) Accuracy is specified for 1 year after calibration, at operating temperatures of 18 C to 28 C, with relative humidity of 0 % to 90 %. Maximum voltage between any terminal and earth ground 600 V Surge protection 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 W Fuse for A input 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 C to +50 C; Storage: -40 C to +60 C Humidity 0 % to 90 % to 35 C; 75 % to 40 C; 45 % to 50 C Temperature coefficient 0.1 x (specified accuracy/C) (< 18 C or > 28 C) Operating altitude 2,000 meters Battery 9 Volt Alkaline, NEDA 1604A/IEC 6LR61 Battery life Alkaline: 400 hours typical, without backlight Safety compliances ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1 Certifications UL, CSA, TUV, N10140 , VDE IP rating (dust and water protection) IP42 N10140 Accuracy specifications (all models) Accuracy ([% of Reading] + [Counts]) Model 0.1 mV 0.5 % + 2 114, 115, 116, 117 0.001 V 0.01 V 0.1 V 0.5 % + 2 114, 115, 116, 117 Function Range Resolution DC millivolts 600.0 mV DC volts 6.000 V 60.00 V 600.0 V DC, 45 Hz to 500 Hz Auto-V LoZ true-rms 1 600.0 V 0.1 V 500 Hz to 1 kHz 2.0 % + 3 4.0 % + 3 45 Hz to 500 Hz 500 Hz to 1 kHz 114, 116, 117 AC millivolts1 true-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 114, 115, 116, 117 AC volts1 true-rms 6.000 V 60.00 V 600.0 V 0.001 V 0.01 V 0.1 V 1.0 % + 3 2.0 % + 3 114, 115, 116, 117 Continuity 600 1 Ohms 600.0 6.000 k 60.00 k 600.0 k 6.000 M 40.00 M Beeper on < 20 , off > 250 ; detects opens or shorts of 500 s or longer 114, 115, 116, 117 0.1 0.001 k 0.01 k 0.1 k 0.001 M 0.01 M 0.9 % + 2 0.9 % + 1 0.9 % + 1 0.9 % + 1 0.9 % + 1 1.5 % + 2 114, 115, 116, 117 Diode test 2.000 V 0.001 V 0.9 % + 2 115, 116, 117 Capacitance 1000 nF 10.00 F 100.0 F 9999 F 1 nF 0.01 F 0.1 F 1 F 1.9 % + 2 1.9 % + 2 1.9 % + 2 100 F to 1000 F: 1.9 % + 2 > 1000 F: 5 % + 20 115, 116, 117 LoZ capacitance (power-up option) 1 nF to 500 F 10 % + 2 typical 115, 116, 117 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled. 1 info@Fluke-Direct.com Fluke-Direct .com 1.888.610.7664 Accuracy specifications (all models) cont. Accuracy ([% of Reading] + [Counts]) Model 0.1 C 0.2 F 1 % + 102 1 % + 182 116 6.000 A 10.00 A 20 A overload for 30 seconds maximum 0.001 A 0.01 A 1.5 % + 3 115, 117 AC Amps true-rms1 (45 Hz to 1 kHz) 600.0 A 0.1 A 1.5 % + 3 (2.5 % + 3 > 500 Hz) 116 DC amps 6.000 A 10.00 A 20 A overload for 30 seconds maximum 0.001 A 0.01 A 1.0 % + 3 115, 117 DC Amps true-rms 600.0 A 0.1 A 1.0 % + 2 116 Hz (V or A input)2 99.99 Hz 999.9 Hz 9.999 kHz 50.00 kHz 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 % + 2 115, 117 99.99 Hz 999.9 Hz 9.999 kHz 50.00 kHz 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 % + 2 116 Function Range Resolution Temperature (K-Type thermocouple) -40 C to 400 C -40 F to 752 F AC amps true-rms1 (45 Hz to 500 Hz) Hz (V input)3 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled. 2 AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC Amps Hz is dc-coupled and specified from 45 Hz to 5 kHz. Amps input burden voltage (typical): 6 A input 2 mV/A, 10 A input 37 mV/A. 3 Frequency is ac-coupled, 45 Hz to 50 kHz. 1 Frequency counter sensitivity (models 115, 116, 117) Typical sensitivity (rms sine wave) Input range Volts AC AC Amps (115, 117 only) 5 Hz to 45 Hz 45 Hz to 5 kHz 5 kHz to 10 kHz 10 kHz to 50 kHz 6V 60 V 600 V 0.2 V 2V 20 V 0.2 V to 0.3 V 2 V to 3 V 20 V to 30 V 0.3 V to 0.4 V 3 V to 4 V 30 V to 40 V 0.4 V to 1.0 V 4 V to 10 V 40 V to 100 V 6A 10 A N/A N/A 0.4 A 0.5 A N/A N/A N/A N/A Input characteristics (all models) Function Input impedance (nominal) Common mode rejection ratio(1 k unbalanced) Volts AC > 5 M < 100 pF > 60 dB at dc, 50 or 60 Hz Volts DC > 10 M < 100 pF > 100 dB at dc, 50 or 60 Hz ~3 k < 500 pF > 60 dB at dc, 50 or 60 Hz Auto-V LoZ Open circuit test voltage Ohms < 2.7 V dc Diode Test < 2.7 V dc Full scale voltage Normal mode rejection > 60 dB at 50 or 60 Hz Short circuit current To 6.0 M 40 M < 0.7 V dc < 0.9 V dc 2.000 V dc < 350 A < 1.2 mA Fluke.Keeping your world up and running.TM Fluke Corporation Fluke-Direct .com Fluke 11X Series Digital Multimeters Extended Specifications info@Fluke-Direct.com 1.888.610.7664