Cover material
thermoplastic resin
Diaphragm material
AISI 316L, Monel (*)
(*) Maximum overrange pressure reduced to 5 MPa (50 bar)
Gaskets material
PTFE, Viton
Capsule filling
Silicone oil, Perfluorinated polyethers (Galden) (*)
(*) When used for oxygen measurements the operating
temperature limits are -20°C and + 40°C.
Surface protections
- carbon steel body and flange: zinc plating and chrome
passivation,
- AISI 316L body and flange: no protection
Process connections (see figure ref. D and E)
- on flanges : 1/2 in NPT-F
- on adapters: 1/4 in NPT-F
- center distance: 54 mm.
Pneumatic connections
- Air supply (in figure ref. A): 1/4 in NPT-F
- Output (in figure ref. B): 1/4 in NPT-F
Pressure gauge
Brass with stainless steel case (all stainless steel on request)
external diameter 51 mm; 0-200 kPa, 0-2 bar and 0-30 psi
indication on 82 mm/260° scale.
Air filter regulator
with copper or stainless steel piping, as specified.
Die cast aluminium alloy with light grey epoxy finish.
Net weight (maximum)
16 kg approx
Packing
expanded polythene box
SPAN
LIMITS
min. and max.
MAXIMUM
ZERO
ELEVATION
STATIC PRESSURE
LIMITS
Full vacuum and
1.2 and 7.5 kPa
12 and 75 mbar
0.6 and 1.2 kPa
6 and 12 mbar
5 in
with special
feedback bellows
(•) Equal to Maximum Working pressure as well as overrange limit (on either side).
RANGE
LIMITS
lover and upper (M)
-7.5 and +7.5 kPa
-75 and +75 mbar
-7.5 and +7.5 kPa
-75 and +75 mbar
MAXIMUM
ZERO
SUPPRESSION (S)
6.3 kPa
63 mbar
6.9 kPa
69 mbar
MEASURING
CAPSULE
(DIAPHRAGM DIA.)
7.5 kPa
75 mbar
7.5 kPa
75 mbar
7,5 MPa (•)
75 bar (•)
SPECIFICATIONS
Air supply
nom. 140 kPa (1.4 bar, 20 psi); min. 125 kPa (1.25 bar,
18 psi); max. 175 kPa (1.75 bar, 25 psi)
Output signal
20 to 100 kPa/0.2 to 1 bar, 3 to 15 psi or 0.2 to 1 kg/cm2
Static air consumption
350 Nl/h
Maximum output flow
- with rising output pressure: 30 Nl/min.
- with falling output pressure: 40 Nl/min.
Accuracy
± 0.5% F.S.D. (typical)
Thermal drift (for ambient temperature variation between
-20° C and + 65° C)
- span 1.2 to 2 kPa (12 to 20 mbar): 0.6%/10°C
- span 2 to 7.5 kPa (20 to 75 mbar): 0.4%/10°C
Static pressure effect
for variation of 2.5 MPa (25 bar): ≤ ± 0.25%
Maximum displacement
2.5 cm3
Degree of protection in accordance with IEC 529
IP55
Ambient temperature limits
-40 and + 120°C
Bodies material
Carbon steel, AISI 316L
Body bolts and nuts material
High tensile carbon steel;
AISI 316 Class A4-80 per ISO3506;
AISI 316 Class A4-50 (*) per ISO3506, in compliance
with NACE MR0175;
high tensile stainless steel, in compliance with
NACE MR0175.
(*) Maximum static pressure reduced to 4 MPa (40 bar)
The data were obtained from laboratory tests on standard instruments with:
carbon steel or AISI 316L bodies; AISI 316L measuring element with silicone oil filling; gasket: PTFE;
calibration span : 4 kPa - 40 mbar.