TB-20780 December 2008 Page 1 of 2
Ground Gard Verifier
Operation and Maintenance Instructions
ESD Systems.com • 432 Northboro Road Central • Marlboro, MA 01752 • (508) 485-7390 • Fax (508) 480-0257 • Website: ESDSystems.com
Figure 1. SPI-20780 Ground Gard Verifier
Description
The SPI-20780 Ground Gard Verifier can be used to verifiy the test
ranges of the Ground Gard 4.5, Ground Gard 5 and Ground Gard
5.5.
Packaging
1 Ground Gard Verifier
4 Banana Plug to Banana Plug Cords
1 Velcro
® Set
Installation
To connect the Ground Gard Verifier to the Ground Gard 4.5:
See Figure 3
To connect the Ground Gard Verifier to the Ground Gard 5:
See Figure 4
To connect the Ground Gard Verifier to the Ground Gard 5.5:
See Figure 5
Note: A
Velcro
® Set is included with the Ground Gard Verifier if
you wish to mount the unit to a specific location.
TECHNICAL BULLETIN TB-20780
Made in America
© 2008 DESCO INDUSTRIES INC.
Employee Owned
Figure 2. Ground Gard Verifier Features and Components
Power Adapter Cord
Ground Gard 4.5
SPI-20780
Remote Remote
Figure 3. Connecting the Ground Gard Verifier to the Ground
Gard 4.5
Power
Adapter Cord
Ground Gard 5
SPI-20780
Remote Remote
Figure 4. Connecting the Ground Gard Verifier to the Ground
Gard 5
Operation
Testing the Remote Modules
Connect the power adapter to an appropriate power source.
Make sure that the switch located on the face of the verifier
is toggled to LEFT WS / RIGHT WS. If connected properly,
all the LEDs on the Ground Gard Monitor should be
illuminated green.
To test the resistance of the left remote module, press the
black button labeled LEFT WS on the verifier. If the remote
module is properly calibrated, the monitor ’s audible alarm will
sound and the LEFT WRIST STRAP LED will illuminate red.
If the LED remains green, locate the trim pot underneath the
respective remote module. Hold down the verifier ’s LEFT
WS button and carefully adjust the trim pot in a counter-
clockwise direction until the monitor indicates a FAIL
condition. If properly calibrated, the monitor should now
indicate a FAIL condition when the verifier ’s LEFT WS button
is pressed.
To test the resistance of the right remote module, press the
black button labeled RIGHT WS on the verifier. If the remote
module is properly calibrated, the monitor ’s audible alarm will
sound and the RIGHT WRIST STRAP LED will illuminate red.
If the LED remains green, locate the trim pot underneath the
respective remote module. Hold down the verifier ’s RIGHT
WS button and carefully adjust the trim pot in a counter-
clockwise direction until the monitor indicates a FAIL
condition. If properly calibrated, the monitor should now
indicate a FAIL condition when the verifier ’s RIGHT WS
button is pressed.
Testing the Ground Monitor (Ground Gard 5 / 5.5 only)
Make sure that the switch located on the face of the verifier
is toggled to LEFT WS / RIGHT WS. If connected properly,
all three LEDs on the Ground Gard Monitor should be
illuminated green.
Disconnect the cords from the banana jacks on both remote
modules. Next, toggle the switch located on the face of the
verifier to OPEN GND / HI RESIST. Press the red button
labeled OPEN GND on the tester, and if the monitor is
functioning properly, its audible alarm should sound and the
GROUND MONITOR LED should illuminate red.
Press the red button labeled HI RESIST on the tester, and
the monitor ’s audible alarm should sound and the GROUND
MONITOR LED should illuminate amber.
If any of these conditions are not met, contact the
manufacturer for repair.
TB-20780 Page 2 of 2
ESD Systems.com • 432 Northboro Road Central • Marlboro, MA 01752 • (508) 485-7390 • Fax (508) 480-0257 • Website: ESDSystems.com
Limited Warranty
SPI expressly warrants that for a period of one (1) year from the date
of purchase, SPI Ground Gard Verifiers will be free of defects in
material (parts) and workmanship (labor). Within the warranty period,
the product will be tested, repaired, or replaced at our option, free of
charge. Call our Customer Service Department at
909-664-9986
for a
Return Material Authorization (RMA) and proper shipping instructions
and address. Include a copy of your original packing slip, invoice, or
other proof of purchase date. Any unit under warranty should be
shipped prepaid to the SPI factory. Warranty repairs will take
approximately two weeks.
If your unit is out of warranty, call Customer Service at
909-664-9986
for a Return Material Authorization (RMA) and proper shipping
instructions and address. SPI will quote repair charges necessary to
bring your unit up to factory standards.
Warranty Exclusions
THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL
OTHER PRODUCT WARRANTIES, EXPRESSED AND IMPLIED,
INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED. The express
warranty will not apply to defects or damage due to accidents, neglect,
misuse, alterations, operator error, or failure to properly maintain, clean
or repair products.
Limit of Liability
In no event will SPI or any seller be responsible or liable for any injury,
loss or damage, direct or consequential, arising out of the use of or the
inability to use the product. Before using, users shall determine the
suitability of the product for their intended use, and users assume all
risk and liability whatsoever in connection therewith.
Power Adapter Cord
Ground Gard 5.5
SPI-20780
Remote Remote
Figure 5. Connecting the Ground Gard Verifier to the Ground
Gard 5.5
© 2008 DESCO INDUSTRIES INC.