Quartzonix™ Pressure Standard Series 960
Functional Diagram
Digital Pressure Standard
ISO-9001 Certified
FEATURES
!Ranges: Barometric and
15 to 500 psia FS (105 to 3400 kPa)
!±0.0001% FS Resolution
!±0.01% FS Accuracy
!±0.01% FS/6 month Stability
!Shock, Acceleration and Vibration Resistant
!Available with Oil-filled Capillary Tube for Water Level
Measurement
APPLICATIONS
!Flight Testing
!Calibration Transfer Standards
!Metrology
!Windtunnel/Turbine Test
!Oceanography
Series 960 Quartzonix™ Pressure Standards are
designed for use as a precision pressure transducer
where the highest levels of traceable accuracy and
stability are required. Quartzonix™ pressure
standards use a patented monolithic quartz resonator
to achieve unparalleled accuracy and stability.
Pressure is measured via a change in the resonant
frequency of an oscillating quartz beam by pressure-
induced stress. Quartzonix™ pressure standards
produce an output frequency between 30 and 45 kHz
and can achieve a pressure resolution of ±0.0001%
FS. The units provide conformance to a calibration
curve of better than ±0.01% FS and have long-term
stability of ±0.01% FS over a six month period.
Precise thermal compensation is provided via an
integrated quartz temperature sensor used to measure
the operating temperature of the transducer. The low
mass design of the sensing elements allows the
transducer to be employed in applications which
experience high acceleration, shock and vibrational
loads.
- 2 -
Series 960 Specifications
@ 25ºC unless otherwise stated
Parameter 960 Units Comments
PNEUMATICS
Pressure Ranges 11 - 16 (76 - 110)
15 (105)
23 (160)
30 (210)
45 (310)
65 (450)
100 (690)
200 (1380)
300 (2070)
500 (3400)
psia (kPa) Barometric only
Proof Pressure11.5 x FS
Burst Pressure22.0 x FS
Pressure Media Media compatible with 316SS,
nickel and solder
Pressure Fitting c" compression
STATIC PERFORMANCE
Resolution3±0.0001 % FS
Pressure Hysteresis ±0.005 % FS
Static Accuracy4±0.01 % FS
Maximum Deviation
over Temperature5±0.015 %FS over specified temperature
range
Thermal Hysteresis ±0.005 % FS over specified temperature
range
Long Term Drift ±0.01 % FS 6 months, maximum
ELECTRICAL
Pressure Signal6
Temperature Signal 45-30
172 kHz
kHz nominal
+45 ppm/ºC
Output Signal 4volt
P-P square wave
capacitively coupled
Turn on Time 30 m sec
Power Requirements 8-15 VDC 2 mA @ 8V
Electrical Connection 8 ea. #30 AWG flying leads
Notes:
1Maximum applied pressure without causing a calibration shift.
2Maximum applied pressure without causing permanent damage to quartz sensing element.
3Achievable resolution using gate time of ½ second for frequency counter with 20.000 MHz time base. Increased or decreased resolution
may be obtained be lengthening or shortening gate times.
4Calibration conformance to a primary pressure standard after calibration.
5Maximum error at any pressure within calibrated temperature range using supplied calibration coefficients/equations.
6Nominal 10% frequency change with full scale pressure.
Specifications subject to change without notice.
- 3 -
Specifications Series 960
Interface Diagram
@ 25ºC unless otherwise stated
Parameter 960 Units Comments
ENVIRONMENTAL
Calibrated Temp Range 0 to 60
-20 to 70 (optional) ºC
Acceleration Sensitivity ±0.004 % FS/G worst axis
Vibration Sensitivity ±0.001 % FS/G 20G peak, 10 Hz - 2 kHz
Maximum Shock 200 G/10 msec ½ sine
PHYSICAL
Size 1.73 x 1.73 x 1.35 (4.4 x 4.4 x 3.4) in (cm)
Weight 9.5 (270) oz. (gms)
Note: Transducer calibration coefficients are stored within an on-board serial EEPROM.
Specifications subject to change without notice.
- 4 -
Series 960 Ordering/Part Number Information
No. 0301
Ordering Information:
PN: 960A-AAAABCDEFF Model 960 Pressure Standard, ±0.01% FS Accuracy
AAAA = Pressure Range
BARO, Barometric 11-16 psia (76-110 kPa)
0015, 15 psia (105 kPa)
0023, 23 psia (160 kPa)
0030, 30 psia (210 kPa)
0045, 45 psia (310 kPa)
0065, 65 psia (450 kPa)
0100, 100 psia (690 kPa)
0200, 200 psia (1380 kPa)
0300, 300 psia (2070 kPa)
0500, 500 psia (3400 kPa)
B = Pressure Fitting
1, c” Swagelok C = Electrical Termination
1, Flying Leads
2, 9 pin D-shell (For use with Model 9600 Pressure Gage)
D = Pressure Calibration
1, Standard E = Calibrated Temperature Range
1, 0 to 60ºC
2, -20 to 70ºC
FF = Specials
00, Standard
10, Oil-filled capillary