Fluke 975 AirMeter specifications
Feature Range Display resolution Accuracy
Measured specifications
Temperature -20 °C to 50 °C
(-5 °F to 122 °F)
0.1 °F
(0.1 °C)
± 0.9 °C/± 1.62°F from 40 °C to 50 °C
± 0.5 °C/± 1.00°F from 5 °C to 40 °C
± 1.1 °C/± 1.98 °F from -20 °C to 5 °C
Relative humidity 10 % to 90 % RH
non-condensing
1 % ± 2 % RH (10 % RH to 90 % RH)
Air velocity 0.50 fpm to 3000 fpm
0.25 m/sec to
15 m/sec
1 fpm
(0.005m/sec)
± 4% or 4 fpm*
± 4 % or 0.02 m/sec* whichever is greater
*Accuracy specification only valid for velocity readings above 50 fpm.
CO20 to 5000 ppm 1 ppm Warm up time 1 min (5 minutes for full specification)
2.75 % + 75 ppm
CO 0 to 500 ppm 1 ppm ± 5 % or ± 3 ppm, whichever is greater, @ 20 °C and
50 % RH
Calculated specifications
Dew point temperature -44 °C to 50 °C
(-47 °F to 122 °F)
0.1 °C
(0.1 °F)
± 1 °C when temp: -20 °C to 50 °C RH: 40 % to 90 %
± 2 °C when temp: -20 °C to 50 °C RH: 20 % to 40 %
± 4 °C when RH: 10 % to 20 %
Wet bulb temperature -16 °C to 50 °C
(3 °F to 122 °F)
0.1 °C
(0.1 °F)
± 1.2 °C when RH: 20 % to 90 % temp: -20 °C to 50 °C
± 2.1 °C when RH: 10 % to 20 %
Volume flow rate (in a duct) 0 to 3,965 M3/m
(0 to 140,000 cfm)
0.001 M3/min
(1 cfm)
N/A: The volume flow calculation will be a simple average
of the data points times the duct area
% outside air
(based on temperature)
0 to 100 % 0.1 % N/A
% outside air (based on CO2) 0 to 100 % 0.1 % N/A
General specifications
Operating temperature:
CO and CO2 sensors
-20 °C to 50 °C (-4 °F to 122 °F)
Operating temperature:
All other functions
-20 °C to 50 °C (-4 °F to 122 °F)
Storage temperature -20 °C to 50 °C (-4 °F to 122 °F)
Humidity 10 % to 90 % (non-condensing)
Altitude Up to 2000 m (6562 ft)
Shock and vibration MIL-PRF-28800F: Class 2
Multi-language interface English, French, Spanish, Portuguese and German
Battery Rechargeable Li-Ion (primary), three-AA (backup)
Data logging 25,000 records (continuous), 99 records (discrete)
Warranty Two-years
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved.
Printed in U.S.A. 9/2006 2694296 D-EN-N Rev B
Fluke. Keeping your world
up and running.™
Ordering Information
Fluke-975 AirMeter™
Fluke-975V AirMeter™ with Velocity
Included
AA alkaline batteries (3)
Users manual (with safety information)
Calibration cap
Hard carrying case
FlukeView® Forms software
Power adapter
International power plugs
Air velocity probe (Fluke 975V only)
Optional accessories
ToolPak™ Magnetic Meter Hanging Kit
Fluke-975CK AirMeter Calibration Kit
Fluke-975VP AirMeter Air Velocity Probe
Fluke 975V pictured