Temperature measuring transducer-MACX PL-T-UIREL-UP-2904901
Please be informed that the data shown in this PDF Document is generated from our Online Catalog. Please find the complete data in the user's
documentation. Our General Terms of Use for Downloads are valid
(http://phoenixcontact.com/download)
Freely programmable temperature transducer with analog output and 3 limit value relays, standard configuration,
resistance thermometer with 2, 3 or 4-conductor technology, thermocouples, electrical isolation, wide-range
power supply, screw connection, PLd
Product Features
Key commercial data
Packing unit 1 pc
Custom tariff number 85437090
Country of origin Germany
Technical data
Note
Utilization restriction EMC: class A product, see manufacturer's declaration in the download
area
Dimensions
Width 35 mm
Height 99 mm
Depth 114.5 mm
Ambient conditions
Ambient temperature (operation) -20 °C ... 65 °C
Ambient temperature (storage/transport) -40 °C ... 85 °C
Maximum altitude ≤ 2000 m
Permissible humidity (operation) typ. 5 % ... 95 % (non-condensing)
https://www.phoenixcontact.com/us/products/2904901
05/18/2015Page2/5
Temperature measuring transducer-MACX PL-T-UIREL-UP-2904901
Technical data
Ambient conditions
Noise immunity EN 61000-6-2 When being exposed to interference, there may be minimal
deviations.
Input data
Sensor types (RTD) that can be used Pt, Ni, Cu sensors: 2, 3, 4-wire
Sensor types that can be used (TC) B, E, J, K, N, R, S, T, L, U, CA, DA, A1G, A2G, A3G, MG, LG
Temperature measuring range -200 °C ... 850 °C
Input signal range 0 Ω ... 50 kΩ
Potentiometer resistance range 0 Ω ... 50 kΩ
Input signal range -1000 mV ... 1000 mV
Output data
Current output signal 4 mA ... 20 mA
Max. current output signal 22 mA
Load/output load current output ≤ 600 Ω (at 20 mA)
Behavior in the event of a sensor error according to NE 43 or freely configurable
Output name Relay output
Output description 1 SIL/PL
Contact type 2 PDT
Contact material AgSnO2, hard gold-plated
Maximum switching voltage 250 V AC (250 V DC)
Maximum inrush current 2 A (250 V AC)
2 A (28 V DC)
0.2 A (120 V DC)
Power supply
Supply voltage range 24 V ... 230 V AC/DC (-20 %/+10 %, 50/60 Hz)
Power consumption < 2.4 W
Connection data
Conductor cross section solid min. 0.2 mm²
Conductor cross section solid max. 2.5 mm²
Conductor cross section flexible min. 0.2 mm²
Conductor cross section flexible max. 2.5 mm²
Conductor cross section AWG min. 24
Conductor cross section AWG max. 14
Screw thread M3
Connection method Screw connection
General
https://www.phoenixcontact.com/us/products/2904901
05/18/2015Page3/5
Temperature measuring transducer-MACX PL-T-UIREL-UP-2904901
Technical data
General
Maximum transmission error 0.1 % (e.g. for Pt 100, 300 K span, 4 ... 20 mA)
Maximum temperature coefficient 0.01 %/K
Inflammability class according to UL 94 V0
Pollution degree 2
Surge voltage category II
Housing material PA 66-FR
Color yellow
Designation Input/output/power supply
Electrical isolation 300 Vrms (Rated insulation voltage (surge voltage category II; pollution
degree 2, safe isolation as per EN 61010-1))
2.5 kV (50 Hz, 1 min., test voltage)
Designation Input/output
Electrical isolation 375 V (Peak value in accordance with EN 60079-11)
Designation Input/power supply
Electrical isolation 375 V (Peak value in accordance with EN 60079-11)
Designation Input/switching output
Electrical isolation 375 V (Peak value in accordance with EN 60079-11)
Conformance CE-compliant
ATEX  II 3 G Ex nA nC ic IIC T4 Gc X
IECEx Ex nA nC ic IIC T4 Gc X
UL, USA / Canada UL 508 Listed
Class I, Div. 2, Groups A, B, C, D T6
Class I, Zone 2, Group IIC T6
Functional Safety (SIL) SIL 2
Classifications
eCl@ss
eCl@ss 5.1 27210107
eCl@ss 6.0 27210120
ETIM
ETIM 4.0 EC002653
ETIM 5.0 EC002653
Approvals
Approvals
https://www.phoenixcontact.com/us/products/2904901
05/18/2015Page4/5
Temperature measuring transducer-MACX PL-T-UIREL-UP-2904901
Approvals
Approvals
UL Listed / cUL Listed / GL / Functional Safety / cULus Listed
Ex Approvals
ATEX / IECEx / UL Listed / cUL Listed / cULus Listed
Approvals submitted
Approval details
UL Listed
cUL Listed
GL
Functional Safety
cULus Listed
Drawings
https://www.phoenixcontact.com/us/products/2904901
05/18/2015Page5/5
Temperature measuring transducer-MACX PL-T-UIREL-UP-2904901
Block diagram
Zone2
Div.2
IN OUT
Power
24V...230VAC/DC
3.3
3.1
1.2
3.2
1.1
1.3
4.3
4.1
4.2
5.3
5.1
5.2
GND
I
OUT
NC
PLC/DCS
passive
-
+
3.3
3.2
11
21
31
12
22
32
14
24
34
3.6
3.4
3.5
2.6
2.4
2.5
1.6
1.4
1.5
DI
DI
Sensor/Field
4.3
4.1
4.2
ϑ
RTD
4.3
4.1
4.2
5.3
5.1
5.2
ϑ
Poti
TC
-
+
ϑ
Phoenix Contact 2015 © - all rights reserved