Data Sheet 40.1001 Page 1/4
06.03 / 00398114
M. K. JUCHHEIM GmbH & Co
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM 20 2TT, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO PROCESS CONTROL INC.
885 Fox Chase, Suite 103
Coatesville PA 19320, USA
Phone: 610-380-8002
1-800-554-JUMO
Fax: 610-380-8009
E-mail: info@JumoUSA.com
Internet: www.JumoUSA.com
Pressure transmitter
JUMO MIDAS
Type 401001
Brief description
Pressure transmitters are used for measuring the relative (gauge) pressure in liquids and
gases. The pressure transmitter incorporates a thick-film strain gauge as a measuring de-
vice. The pressure sensor has an aluminium-oxide (Al2O3) ceramic base material. The
pressure is converted into an electrical signal.
Technical data
Reference conditions
to DIN 16 086 and IEC 770/5.3
Ranges
see order details
Overload limits
for ranges
0 — 40 bar 3 x full scale
ranges
0 — 60 to 0 — 100 bar 2 x full scale
Bursting pressure
ranges 0 — 40 bar ≤ 5 x full scale
ranges
0 — 60 to 0 — 100 bar 3 x full scale
Parts in contact with medium
standard: st. steel, Mat. Ref. 1.4305,
(AI2O3) 96%
seal: Viton®(FPM) or Isolast®(FFPM)
Output
4—20 mA
2-wire burden ≤ (UB-10 V) / 0.02A
0.5 — 4.5 V burden ≥ 20 kΩ
1 — (5)6 V burden ≥ 10 kΩ
0 — 10 V burden ≥ 10 kΩ
Burden error
< 0.5% max.
Zero offset
≤ 0.3% of full scale
Thermal hysteresis
≤ ± 0.8% of full scale
Ambient temperature error
within range -20 to +85°C
(compensated temperature range)
zero: ≤ 0.02%/°C typical,
≤ 0.04%/°C max.
span: ≤ 0.02%/°C typical,
≤ 0.04%/°C max.
Deviation from characteristic
≤ 0.5% of full scale
(limit point adjustment)
Hysteresis
≤ 0.2% of full scale
Repeatability
≤ 0.1% of full scale
Response time
≤ 3 msec max.
Stability per year
≤ 1% of full scale
Supply
10 — 30 V DC (for output 4 — 20 mA
and 1 — (5)6 V)
5 V DC (for output 0.5 — 4.5 V)
11.5 — 30 V DC (for output 0 — 10 V)
Ripple: the voltage spikes must not go
above or below the values specified for the
supply
max. current drawn: approx. 25 mA
Supply voltage error
≤ 0.02% per V
(nominal supply voltage 24 V DC)
ratiometric with supply 5 V DC (±0.5 V)
Permissible ambient temperature
for version with plug:
-20 to +125°C
for version with attached cable:
-20 to +100°C
Storage temperature
-40 to +125°C
for version with attached cable
-20 to +100°C
Permissible temperature of medium
-30 to +125°C
Electromagnetic compatibility (EMC)
to EN 61 326
Mechanical shock
(to IEC 68-2-27)
100 g/1 msec
Mechanical vibration
(to IEC 68-2-6)
20 g max. at 15 — 2000 Hz
Protection
with terminal box
IP65 to EN 60 529
(diameter of connecting cable
5 mm min., 7 mm max. )
with connecting cable or
circular connector M 12 x 1
IP67 to EN 60 529
Housing
stainless steel, Mat. Ref. 1.4305
EPDM
Pressure connection
see order details;
other connections on request
Type 401001/000-xxx-xxx-xx-xxx-61
Type 401001/000-xxx-xxx-xx-xxx-36