Accessories
l
Instruction manual
l
4-conductor cable (with a connector; 2 m) A4372-02
INFRARED DETECTOR
Amplifier for infrared detector
Low noise amplifier for InGaAs, PbS, PbSe and MCT detector
C4159/C5185 series, C3757-02
■ Absolute maximum ratings (Ta=25 °C)
Parameter Value Unit
Operating temperature 0 to +40 °C
Storage temperature -20 to +70 °C
■ Amplifiers for photovoltaic detectors (Typ.)
Parameter C4159-01 C4159-04 C4159-05 C4159-02 C4159-03 Unit
Applicable detector
InSb (dewer type) *
1
(φ0.6 mm, 1 mm)
InSb (dewer type) *
1
(
φ
2 mm)
InAs
(P7163) InGaAs InGaAs -
Conversion impedance 108, 107, 106
(3 ranges switchable)
2 × 10
7
, 2 × 10
6
, 2 × 10
5
(3 ranges switchable)
108, 107, 106
(3 ranges switchable)
250 to 5 × 103
continuous
107, 106, 105
(3 ranges switchable)
V/A
Frequency response (amp. only, -3 dB)
DC to 100 k DC to 45 k DC to 15 k 20 to 120 M DC to 15 k Hz
Output impedance 50 50 50 50 50 Ω
Maximum output voltage (1 k
Ω
load)
+10 +10 +10 ±1.3 +10 V
Output offset voltage ±5 ±5 ±10 ±30 ±5 mV
Equivalent input noise current (f=1 kHz)
0.15 *20.55
0.15 (10
8
, 10
7
range)
0.65 (106 range) 30 2.5 pA/Hz1/2
Reverse voltage Impossible Internally
generated
Can be applied
from external unit
-
External power supply ±15 ±15 ±15 V
Current consumption *3+30, -10 Max. +90, -60 Max. ±15 Max. mA
Note) Output nose voltage = Equivalent input noise current × Conversion impedance
■ Amplifiers for photoconductive detectors (Typ.)
Parameter C5185 C5185-01 C3757-02 Unit
Applicable detector MCT (dewar type),
InSb (P6606 series)
MCT
(P3981/P2750 series) *4PbS, PbSe -
Input impedance 5 5 10000 kΩ
Voltage gain 66 (× 2000) 66 (× 2000) 40 (× 100) dB
Frequency response (amp. only, -3 dB)
5 to 250 k 5 to 250 k 0.2 to 10 k Hz
Detector bias current 5 mA, 10 mA, 15 mA
(3 ranges switchable)
0.1 mA, 0.5 mA, 1 mA
(3 ranges switchable) Internal bias -
Output impedance 50 50 50 Ω
Maximum output voltage (1 k
Ω
load)
±2.5 ±2.5 ±10 V
Equivalent input noise voltage (f=1 kHz)
2.6 1.2 40 nV/Hz1/2
External power supply ±15 ±15 ±15 V
Current consumption *3+60, -10 Max. +60, -10 Max. +15, -15 Max. mA
Note) Output noise voltage = Equivalent input noise voltage
× Voltage gain
*1: Amplifiers for multi-element detectors are separately provided.
*2: 0.65 pA/Hz1/2 when conversion impedance is set to 106 V/A
*3: Recommended DC power supply (analog power supply): ±15 V
Current capacity: More than 1.5 times the maximum current consumption
Ripple noise: 6 mVp-p or less
*4: Preamp for P3257-30/-31 available upon request
1