Features * * * * * * Regulated Converters Reinforced Insulation for 250VAC Working Voltage Clearance and Creepage Distance: 8mm 5kVAC I/P to O/P 2MOPP Isolation 2A Patient Leakage Current Industry Standard Pinout 2:1 and 4:1 Wide Input Range Description Selection Guide Part Number Input Voltage Range (VDC) Output Voltage (VDC) Output Current (mA) REM3-xx3.3S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 3.3 REM3-xx05S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 5 REM3-xx12S/ 4.5-9 / 9-18 / 18-36 / 36-75 12 REM3-xx15S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 15 REM3-xx24S/ Efficiency typ. (%) Max. Capacitive Load (F) 1000 81 /82 /82 / 81 1050 600 84.5 / 84.5 / 84.5 / 84 780 250 85.5 / 87 / 87 / 87 130 200 87.5 / 87 / 87 / 86.5 100 4.5-9 / 9-18 / 18-36 / 36-75 24 125 85.5 / 87 / 87 / 86.5 39 REM3-xx05D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 5 300 83 / 83.5 / 83 / 83 430 REM3-xx12D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 12 125 86 / 87.5 / 86 / 86 75 REM3-xx15D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 15 100 86 / 86.5 / 86 / 86 56 REM3-xx3.3SW/ (3,4) REM3 3 Watt 5 2:1 & 4:1 DIP24 Single and Dual Output 5Y The REM3 series of medical grade regulated DC/DC converters features reinforced 5kVAC/1 minute isolation with low 2A leakage and are 60601-1 3rd Ed. certified for 250VAC continuous working. The compact DIP24 package offers tightly regulated single and dual outputs, even under no-load conditions. The outputs are short circuit and overload protected. The converters are available in two different pinning options and optionally with an external control pin for standby consumption as low as 12.5mW. The converters are fully certified to CB, IEC/EN and ANSI/AAMI standards and carry the CE and UL marks. (3,4) DC/DC Converter 9-36 / 18-75 3.3 1000 82 /81 1050 REM3-xx05SW/ (3,4) 9-36 / 18-75 5 600 84.5 / 84 750 REM3-xx12SW/ (3,4) 9-36 / 18-75 12 250 87 / 87 130 REM3-xx15SW/ (3,4) 9-36 / 18-75 15 200 87 / 86.5 100 REM3-xx24SW/ (3,4) 9-36 / 18-75 24 125 87 / 86.5 39 REM3-xx05DW/ (3,4) 9-36 / 18-75 5 300 83 / 83 430 REM3-xx12DW/ (3,4) 9-36 / 18-75 12 125 87 / 86 75 REM3-xx15DW/ 9-36 / 18-75 15 100 86 / 86 56 (3,4) (3,4) wa E314885 EAR rranty 2MOPP 250VAC Model Numbering REM 3- __ __ __ __ / __ __ Output Power CTRL Option (4) Input Voltage (1) Pinning (3) Output Voltage Input Type (2) IEC-60601-1 Certified ES-60601-1 Certifed EN-55011 Certified EN-55022 Certified Single or Dual Notes: Note1: for 4:1 Input Voltage Type add "W", see Note 2. 2:1 4:1 "W" xx= 4.5-9 Vin = "05" xx= 9-36Vin = xx= 9-18Vin = "12" xx= 18-75Vin = xx= 18-36Vin = "24" xx= 36-75Vin = "48" "24" "48" Note2: Blank for Standard 2:1 Input Voltage Range; W" suffix for 4:1 Input Voltage Range Note3: A" suffix for A pinning; C" suffix for C pinning, for more details refer to Package Style and Pinning Note4: CTRL" suffix for control pin option, for A pinning only, for C pinning not aviable Examples: REM3-0512D/A REM3-1215S/C REM3-4815SW/A/CTRL REM3-243.3SW/C www.recom-power.com = = = = 2:1 Input, 2:1 Input, 4:1 Input, 4:1 Input, 4.5-9Vin, 9-18Vin, 36-75Vin, 9-36Vin, 12Vout, 15Vout, 15Vout, 3.3Vout, pinout A", pinout C", pinout A" pinout C", without control pin without control pin with control pin without control pin REV.: 05/2016 Refer to Applications Notes MED-1 REM3 DC/DC Converter Series Specifications (measured at TA= 25C, nominal input voltage, full load and after warm-up) BASIC CHARACTERISTICS Parameter Condition 2:1 Absolute Maximum Input Voltage ( 3sec max.) 4:1 2:1 Typ. Max. 16VDC 25VDC 50VDC 100VDC 50VDC 100VDC 5Vin nom. 12Vin nom. 24Vin nom. 48Vin nom. 24Vin nom. 48Vin nom Under Voltage Lockout 4:1 Start-up Time Min. 5Vin nom. 12Vin nom. 24Vin nom. 48Vin nom. 24Vin nom. 48Vin nom 4VDC 8VDC 16VDC 33VDC 8VDC 16VDC 4.5VDC 9VDC 18VDC 36VDC 9VDC 18VDC constant resistive load, Power up or Remote ON/OFF 30ms DC-DC ON DC-DC OFF Open or 0-1.2VDC 2.2-12VDC Remote ON/OFF (referenced to -Vin Pin) Current of CTRL Pin -0.5mA 1mA Remote OFF Input Current 2.5mA Operating Frequency Range 135kHz Output Ripple and Noise (20MHz BW limited) 150kHz 10F/25V X7R MLCC for 3.3, 5Vout 10F/25V X7R MLCC for 12, 15Vout 4.7F/50V X7R MLCC for 24Vout 165kHz 30mVp-p 40mVp-p 50mVp-p Efficiency Table1: Efficiency Crosstable Graph1: Efficiency Factor vs. Load 1.2 Efficiency Crosstable (%) @ full load 1.0 12 24 48 24W 48W 3.3S 81 82 82 81 82 81 05S 84.5 84.5 84.5 84 84.5 84 12S 85.5 87 87 87 87 87 15S 87.5 87 87 86.5 87 86.5 24S 85.5 87 87 86.5 87 86.5 05D 83 83.5 83 83 83 83 12D 86 87.5 86 86 87 86 15D 86 86.5 86 86 86 86 Efficiency Factor Output Voltage Input Voltage 5 0.8 0.6 0.4 0.2 0.0 0 10 20 30 40 50 60 70 80 90 100 Load (%) continued on next page MED-2 REV.: 05/2016 www.recom-power.com REM3 DC/DC Converter Series Specifications (measured at TA= 25C, nominal input voltage, full load and after warm-up) Calculation Example: Efficiency (%) 100 90 choose your model: 80 REM3-1212D 70 * Efficiency from Table1 (= 87.5% @ max Load / nom Vin) 60 50 * Loading conditions in application (e.g. 50%) 10 20 30 40 50 60 70 80 90 * use Eff factor from Graph1 (= 0.98 @50%) Load (%) Calculation: Eff50% = Eff100% * Efffactor50% = 87.5 * 0.98 = 85.75% Vin = 12V Pout100% * 0.5 Iout = 50% PDIS50% = Pin50% - Pout50% = - (Pout100% * 0.5) = 1.75 - 1.5 = 0.25W Eff50% Eff100% = 87.5% Efffactor50% = 0.98 TOVER = Rth * PDIS50% = 18 * 0.25 = 4.5C Rth = 18C/W TCASEmax = 105C T = T - T = 105 - 4.5 = 100.5C AMBmax CASEmax 100 OVER REGULATIONS Parameter Condition Type Value Output Voltage Accuracy 1% Line Voltage Regulation Load Voltage Regulation Cross Regulation Transient Response LL to HL LL to HL Single Dual 0.2% no load to full load no load to full load Single Dual 0.2% 0.5% 1% asymmetrical load 25% / Full Load only Dual Output 5% 25% load step change 250s PROTECTIONS Parameter Condition Type Short Circuit Protection (SCP) Over Load Protection (OLP) Value continuous, auto-recovery % of Iout rated Hiccup mode, 150% typ. Single 3.3Vout 5Vout 12Vout 15Vout 24Vout 3.7VDC min. / 5VDC max. 5.6VDC min. / 7VDC max. 13.5VDC min. / 16VDC max. 18.3VDC min. / 22VDC max. 29.1VDC min. / 34.5VDC max. Dual 5Vout 12Vout 15Vout 5.6VDC min. / 7VDC max. 13.5VDC min. / 18.2VDC max. 17VDC min. / 22VDC max. Output Over Voltage Protection (OVP) continued on next page www.recom-power.com REV.: 05/2016 MED-3 REM3 DC/DC Converter Series Specifications (measured at TA= 25C, nominal input voltage, full load and after warm-up) Isolation Voltage I/P to O/P working voltage 5kVAC / 1 minute 250VAC / continuous Means of Protection 2MOPP Leakage Current 240VAC, 60Hz 2A Medical Device Classification Type CF applied device (design to meet) Internal Clearance Creepage I/P to O/P I/P to O/P External Clearance I/P to O/P, Single Output I/P to O/P, Dual Output I/P to O/P, Single Output I/P to O/P, Dual Output Creepage 8mm 8mm >19.72mm >14.64mm >19.72mm >14.64mm Isolation Capacitance 12pF typ. / 17pF max. Insulation Grade Notes: Note5: Reinforced Insulation This Power module is not internally fused. A input line fuse must be always used. Recomended Fuse: 2:1 Input Voltage Fuse (slow blow) 4:1 Input Voltage Fuse (slow blow) 5V T1.25A 24V T0.63A 12V T0.63A 48V T0.315A 24V T0.315A 48V T0.315A ENVIRONMENTAL Parameter Condition Relative Humidity 5% to 95% RH Temperature Coefficient Thermal Impedance 0.02% / C natural convection (20LFM) max. Case Temperature Range max. Ambient Temperature Range MTBF (+25C) Value 18C / W -40C to +105C see calculation example according to MIL-HDBK-217F, full load 6444 x 103 hours SAFETY AND CERTIFICATIONS Certificate Type (Safety) CB Medical Safety Report / File Number E314885-A6 1409015 Standard IEC-60601-1 Medical Report + ISO14971 Risk Assessment ANSI/AAMI E314885-A6 ES60601-1 CAN/CSA Medical E314885-A6 C22.2 No. 60601-1:08 Certificate Type (Others) Conditions Standard / Criterion EMI Standard (7) Conducted CISPR 11, EN55011, Class A, B Radiated CISPR 11, EN55011, Class A, B Conducted and Radiated FCC18 Air 8kV; Contact 6kV EN61000-4-2, Criteria A Radiated Immunity 10V/m EN61000-4-3, Criteria A Fast Transient (6) 2kV EN61000-4-4, Criteria A 2kV EN61000-4-5, Criteria A 20Vr.m.s EN61000-4-6, Criteria A ESD Surge (6) Conducted Immunity continued on next page MED-4 REV.: 05/2016 www.recom-power.com REM3 DC/DC Converter Series Specifications (measured at TA= 25C, nominal input voltage, full load and after warm-up) Power Frequency Magnetic Field 10A/m EN61000-4-8, Criteria A Thermal Shock MIL-STD-810F Vibration MIL-STD-810F Notes: Note6: An external input filter capacitor is required if the model has to meet EN61000-4-4 or/and EN61000-4-5. Recommended components: 5Vin aluminium capacitor (Nippon Chemi-con KY series, 1000F/25V) and a reverse diode (Vishay V10P45) to connect in parallel 12Vin, 24Vin aluminium capacitor (Nippon Chemi-con KY series, 470F/50V) 48Vin aluminium capacitor (Nippon Chemi-con KY series, 330F/100V) Note7: The whole REM3 series can meet EMI Class A with no external filter. And Class B only with external components. EMC Filter Suggestion for Class B C1 DC/DC Converter L1 +VIN C2 -VIN MODEL C1 (8) C2 (8) L1 (8) REM3-05xxS_D 22F/16V MLCC 22F/16V MLCC 33mH WE 744825433 REM3-12xxS_D REM3-24xxS_D REM3-24xxS_D/W 4.7F/50V MLCC 4.7F/50V MLCC 500H WE 744223 REM3-48xxS_D REM3-48xxS_D/W 2.2F/100V MLCC 1F/100V MLCC 200H WE 744272221 Notes: Note8: The component values can be adapted according to customers` applictation. DIMENSION and PHYSICAL CHARACTERISTICS Parameter Type Case Material Potting Material Value non-conductive black plastic silicone (UL94-V0) Package Dimension (LxWxH) 31.80 x 20.30 x 10.40mm Package Weight 14g continued on next page www.recom-power.com REV.: 05/2016 MED-5 REM3 DC/DC Converter Series Specifications (measured at TA= 25C, nominal input voltage, full load and after warm-up) Dimension Drawing (mm) "A" Pinning (Standard) MEDICAL GRADE DC/DC CONVERTER 20.3 MEDICAL GRADE DC/DC CONVERTER REM3-Series REM3-Series 31.8 3.8 3.8 10.4 10.4 31.8 15.24 2.54 2 15.24 20.32 0.6 dia. Pin Connections Pin # Single Dual 1 CTRL* CTRL* 2 -Vin -Vin 11 NC -Vout 14 +Vout +Vout 16 -Vout Com 22 +Vin +Vin 23 +Vin +Vin 11 12 1 Bottom View 24 23 2 5.08 2.54 2.54 Pin Connections Pin # Single Dual 1 +Vin +Vin 11 No Pin Com 12 -Vout No Pin 13 +Vout -Vout 15 No Pin +Vout 23 -Vin -Vin 24 -Vin -Vin 20.3 "C" Pinning 15 13 1 2 20.32 0.6 dia. 11 Bottom View 23 22 16 14 * If dont choose CTRL option, there is no pin on the corresponding pin number Tolerance: xx.x= 0.5mm xx.xx= 0.25mm NC= not connected Tolerance: xx.x= 0.5mm xx.xx= 0.25mm PACKAGING INFORMATION Parameter Type Value Packaging Dimension (LxWxH) Tube 255 x 21.8 x 16.5mm Packaging Quantity 7pcs Storage Temperature Range -55C to +125C 16.50.75 11.50.75 0.50.2 Tube Dimension Drawing (mm) 110.75 21.80.75 The product information and specifications are subject to change without prior notice. RECOM products are not authorized for use in safety-critical applications (such as life support) without RECOM's explicit written consent. A safety-critical application is defined as an application where a failure of a RECOM product may reasonably be expected to endanger or cause loss of life, inflict bodily harm or damage property. The buyer shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized use of RECOM products in such safety-critical applications. MED-6 REV.: 05/2016 www.recom-power.com Mouser Electronics Authorized Distributor Click to View Pricing, Inventory, Delivery & Lifecycle Information: RECOM: REM3-1212S/C REM3-2415DW/C REM3-4812S/A/CTRL REM3-4812SW/A/CTRL REM3-4812SW/A REM32415SW/C REM3-4815D/A REM3-4824SW/A REM3-4812DW/A/CTRL REM3-2415SW/A/CTRL REM3-4815D/C REM3-2412SW/A/CTRL REM3-2424SW/C REM3-4815DW/A REM3-1212S/A/CTRL REM3-0512S/A/CTRL REM34805S/A/CTRL REM3-4805S/C REM3-4812D/A/CTRL REM3-0515D/A/CTRL REM3-2415D/C REM3-243.3SW/C REM3-1205D/C REM3-0512D/C REM3-2415S/C REM3-2415S/A REM3-2412S/A REM3-1215S/A REM3-1224S/A REM3-0505D/C REM3-4805SW/A/CTRL REM3-2412DW/C REM3-2415DW/A/CTRL REM3-243.3S/C REM31205S/A REM3-1212S/A REM3-2405D/A REM3-2415D/A REM3-2405DW/A REM3-2412DW/A/CTRL REM30505S/A/CTRL REM3-4812D/C REM3-053.3S/C REM3-123.3S/A REM3-1205S/C REM3-2415S/A/CTRL REM30524S/A REM3-4805DW/A/CTRL REM3-483.3S/C REM3-1215S/C REM3-2405D/C REM3-2415D/A/CTRL REM3483.3SW/A/CTRL REM3-4805SW/A REM3-0505S/A REM3-0512S/A REM3-2424S/A/CTRL REM3-4812S/A REM31212D/A REM3-1212D/A/CTRL REM3-4805SW/C REM3-2412SW/A REM3-2415DW/A REM3-0515S/A/CTRL REM3-2424SW/A REM3-123.3S/A/CTRL REM3-4815D/A/CTRL REM3-4805DW/C REM3-2415SW/A REM30524S/C REM3-4812D/A REM3-2412D/A REM3-4805D/A REM3-243.3SW/A REM3-4805S/A REM3-1224S/C REM3-1205D/A REM3-1215D/C REM3-2405S/C REM3-4812S/C REM3-2405DW/C REM3-2412DW/A REM34812DW/A REM3-4824S/A REM3-2412S/A/CTRL REM3-2412S/C REM3-1215S/A/CTRL REM3-4815DW/C REM34815S/A REM3-053.3S/A REM3-2424S/C REM3-4824S/C REM3-1215D/A REM3-483.3SW/A REM34815DW/A/CTRL REM3-053.3S/A/CTRL REM3-483.3S/A REM3-0515S/A REM3-4815S/A/CTRL REM3-0512D/A