Serial Communications Accessories SITRANS RD200/300 Operating Instructions * 05/2013 SITRANS SIT SITRANS RD300RANS RD300 Safety Guidelines: Warning notices must be observed to ensure personal safety as well as that of others, and to protect the product and the connected equipment. These warning notices are accompanied by a clarification of the level of caution to be observed. Qualified Personnel: This device/system may only be set up and operated in conjunction with this manual. Qualified personnel are only authorized to install and operate this equipment in accordance with established safety practices and standards. Unit Repair and Excluded Liability: The user is responsible for all changes and repairs made to the device by the user or the user's agent. All new components are to be provided by Siemens Milltronics Process Instruments. Restrict repair to faulty components only. Do not reuse faulty components. WARNING: Cardboard shipping package provides limited humidity and moisture protection. This product can only function properly and safely if it is correctly transported, stored, installed, set up, operated, and maintained. This product is intended for use in industrial areas. Operation of this equipment in a residential area may cause interference to several frequency based communications. Note: Always use product in accordance with specifications. Copyright Siemens AG 2013. All Rights Reserved Disclaimer of Liability This document is available in bound version and in electronic version. We encourage users to purchase authorized bound manuals, or to view electronic versions as designed and authored by Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments will not be responsible for the contents of partial or whole reproductions of either bound or electronic versions. While we have verified the contents of this manual for agreement with the instrumentation described, variations remain possible. Thus we cannot guarantee full agreement. The contents of this manual are regularly reviewed and corrections are included in subsequent editions. We welcome all suggestions for improvement. Technical data subject to change. MILLTRONICS(R) is a registered trademark of Siemens Milltronics Process Instruments. Contact SMPI Technical Publications at the European Authorized Representative following address: Technical Publications Siemens AG Siemens AG Industry Sector Siemens Milltronics Process Instruments 76181 Karlsruhe 1954 Technology Drive, P.O. Box 4225 Deutschland Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com For a selection of Siemens Milltronics level measurement manuals, go to: www. siemens.com/processautomation. Under Process Instrumentation, select Level Measurement and then go to the manual archive listed under the product family. For a selection of Siemens Milltronics weighing manuals, go to: www. siemens.com/processautomation. Under Weighing Technology, select Continuous Weighing Systems and then go to the manual archive listed under the product family. (c) Siemens AG 2013 SITRANS RD300 Table of Contents SAFETY NOTES ------------------------------------------------------------ 1 The Manual ------------------------------------------------------------------------- 1 SAFETY INFORMATION ------------------------------------------------- 2 SPECIFICATIONS ---------------------------------------------------------- 3 RD200 RS-232 Serial Adapter ------------------------------------------------ 3 RD200 RS-485 Serial Adapter ------------------------------------------------ 3 RD200 USB Serial Adapter ---------------------------------------------------- 3 RD300 RS-232 Serial Adapter ------------------------------------------------ 4 RD300 RS-485 Serial Adapter ------------------------------------------------ 4 RD300 USB Serial Adapter ---------------------------------------------------- 4 RD200/300 USB to RS-232 Converter -------------------------------------- 5 RD200/300 RS-232 to RS-422/485 Converter ---------------------------- 5 RD200/300 USB to RS-422/485 Converter -------------------------------- 5 SERIAL COMMUNICATIONS OVERVIEW -------------------------- 6 RD200 RS-232 SERIAL ADAPTER ------------------------------------ 8 Description ------------------------------------------------------------------------- 8 Installation -------------------------------------------------------------------------- 9 Connections -------------------------------------------------------------------- 10 RD200 Serial Communication Address (serl)------------------------ 10 RD200 RS-485 SERIAL ADAPTER ----------------------------------- 11 Description ----------------------------------------------------------------------- 11 Installation ------------------------------------------------------------------------ 12 RD200 Serial Communication Address (serl)------------------------ 12 Connections -------------------------------------------------------------------- 15 RD200 TO USB SERIAL ADAPTER---------------------------------- 16 Description ----------------------------------------------------------------------- 16 Installation ------------------------------------------------------------------------ 16 RD300 RS-232 SERIAL ADAPTER ----------------------------------- 17 Description ----------------------------------------------------------------------- 17 Installation ------------------------------------------------------------------------ 18 Connections -------------------------------------------------------------------- 18 RD300 Serial Communication Address (serial) -------------------- 19 RD300 RS-485 SERIAL ADAPTER ----------------------------------- 20 Description ----------------------------------------------------------------------- 20 Installation ------------------------------------------------------------------------ 21 RD300 Serial Communication Address (serIAl) -------------------- 21 i SIT SITRANS RD300RANS RD300 Connections -------------------------------------------------------------------- 24 RD300 TO USB SERIAL ADAPTER---------------------------------- 25 Description ----------------------------------------------------------------------- 25 Installation ------------------------------------------------------------------------ 25 RD200/300 USB TO RS-232 CONVERTER ------------------------ 26 Description ----------------------------------------------------------------------- 26 Installation ------------------------------------------------------------------------ 26 RD200/300 RS-232 TO RS-422/485 CONVERTER --------------- 28 Description ----------------------------------------------------------------------- 28 Installation ------------------------------------------------------------------------ 29 Connections -------------------------------------------------------------------- 31 RD200/300 USB TO RS-422/485 CONVERTER ------------------ 33 Description ----------------------------------------------------------------------- 33 Installation ------------------------------------------------------------------------ 34 Connections -------------------------------------------------------------------- 35 TROUBLESHOOTING---------------------------------------------------- 37 ii SITRANS RD300 SAFETY NOTES Special attention must be paid to warnings and notes highlighted from the rest of the text by grey boxes. CAUTION: relates to the caution symbol on the product, and means that failure to observe the necessary precautions can result in electric shock. ! WARNING: relates to a caution symbol on the product, and means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage. WARNING: means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage. CAUTION: means that failure to observe the necessary precautions can result in considerable material damage. Note: Means important information about the product or that part of the operating manual. The Manual This manual provides descriptions for the Converters and Adapters for connection to the SITRANS RD200/300. The manual is designed to help you get an overview of each and to aid in the selection and setup of the correct converter or adapter for your applications. If you have any questions, comments, or suggestions about the manual contents, please email us at techpubs.smpi@siemens.com. For the complete library of Siemens manuals, go to www.siemens.com/processautomation. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 1 SIT SITRANS RD300RANS RD300 SAFETY INFORMATION ! CAUTION: Read complete instructions prior to installation and operation of the meter. WARNING: Risk of electric shock or personal injury. WARNING: Hazardous voltages exist within enclosure. Installation and service should be performed only by trained service personnel. Page 2 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 SPECIFICATIONS RD200 RS-232 Serial Adapter COMPATIBILITY CONNECTORS EIA-232 PC compatible 9-pin D subminiature connector (DB9) and RJ11 (adapter to meter) CABLE 7' (2.1 m) standard modular cable provided with adapter DISTANCE Adapter to RD200 meter: 7' (2.1 m) max; Adapter to computer: 50' (15 m) max; serial interface cable not provided POWER Powered by RD200 RJ11 connection STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) RD200 RS-485 Serial Adapter COMPATIBILITY CONNECTORS EIA-485 Removable screw terminal connector and RJ11 (adapter to meter) CABLE 7' (2.1 m) standard modular cable provided with adapter DISTANCE Adapter to RD200 meter: 7' (2.1 m) max; Adapter to computer: 3,937' (1,200 m) max POWER Powered by RD200 RJ11 connection STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) RD200 USB Serial Adapter COMPATIBILITY CONNECTORS USB 1.1, USB 2.0 RJ11, and USB Type B CABLE One 7' (2.1 m) standard modular cable and one 3.28' (1.0 m) USB A-B Male cable provided with adapter DISTANCE Adapter to RD200 meter: 7' (2.1 m) max; USB connection to PC: 10' (3 m) max DRIVER Windows 98/SE, ME, 2000, Server 2003/2008, XP 32/64Bit, Vista 32/64-Bit, Windows 7 32/64-Bit POWER USB Port STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 3 SIT SITRANS RD300RANS RD300 RD300 RS-232 Serial Adapter COMPATIBILITY CONNECTORS EIA-232 PC compatible 9-pin D subminiature connector (DB9) and RJ45 (adapter to meter) CABLE 6' (1.8 m) standard Cat5e cable provided with adapter DISTANCE Adapter to RD300 meter: 6' (1.8 m) max; Adapter to computer: 50' (15 m) max; serial interface cable not provided POWER Powered by RD300 meter M-Link connection STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) RD300 RS-485 Serial Adapter COMPATIBILITY CONNECTORS EIA-485 Removable screw terminal connector and RJ11 (adapter to meter) CABLE 6' (1.8 m) standard Cat5e cable provided with adapter DISTANCE Adapter to RD300 meter: 6' (1.8 m) max; Adapter to computer: 3,937' (1,200 m) max POWER Powered by RD300 meter M-Link connection STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) RD300 USB Serial Adapter COMPATIBILITY CONNECTORS USB 1.1, USB 2.0 RJ45, and USB Type B CABLE One 6' (1.8 m) standard Cat5e cable and one 3.28' (1.0 m) USB A-B Male cable provided with adapter DISTANCE Adapter to RD300 meter: 6' (1.8 m) max. USB connection: 10' (3 m) max DRIVER Windows 98/SE, ME, 2000, Server 2003/2008, XP 32/64Bit, Vista 32/64-Bit, Windows 7 32/64-Bit POWER USB Port STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) Page 4 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RD200/300 USB to RS-232 Converter COMPATIBILITY CONNECTORS DRIVERS DISTANCE POWER USB 1.1, USB 2.0, EIA-232 PC compatible 9-pin D subminiature connector (DB9) and USB Type A (R) Windows 98/2000/ME/XP USB connection: 10' (3 m) max; RS-232 connection: 50' (15m) max USB port RD200/300 RS-232 to RS-422/485 Converter COMPATIBILITY CONNECTORS DISTANCE EIA-232, EIA-422, and EIA-485 Screw terminal connector and DB9 RS-232 connection: 50' (15 m) max; RS-422/485 connection: 3,937' (1,200 m) max NUMBER OF UNITS Up to 31 RS-485 compatible devices POWER 9-12 VDC; 115 VAC/12 VDC adapter included STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) ISOLATION Isolated RS-232 to RS-422/485: 1500 VAC between data lines; 700 VDC input/output-to-power RS-232 to RS-422/485: 1500 VAC between data lines only RD200/300 USB to RS-422/485 Converter COMPATIBILITY CONNECTORS DRIVERS DISTANCE USB 1.1, USB 2.0, EIA-422, and EIA-485 Screw terminal connector and USB Type B (R) Windows 98/2000/ME/XP, Linux 2.4 & greater USB connection: 10' (3 m) max; RS-422/485 connection: 3,937' (1,200 m) max NUMBER OF UNITS Up to 31 RS-485 compatible devices POWER USB port STATUS INDICATION Separate LEDs for Power (P), Transmit (TX), and Receive (RX) ISOLATION Isolated USB to RS-422/485: 1500 VAC between data lines; 700 VDC input/output-to-power USB to RS-422/485: 1500 VAC between data lines only A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 5 SIT SITRANS RD300RANS RD300 SERIAL COMMUNICATIONS OVERVIEW RS-232, RS-422, and RS-485 are standard interfaces approved by the Electronic Industries Alliance (EIA) for connecting serial devices. In EIA terms, the device (such as meter) that connects to the interface is called a Data Communications Equipment (DCE) and the device to which it connects (such as a computer) is called a Data Terminal Equipment (DTE). The RS-422 standard was designed to replace the older RS-232 standard because it supports higher data rates and greater immunity to electrical interference. RS-485 is similar to RS-422 but can support multi-point connections per line because it uses lower-impedance drivers and receivers. Line drivers and receivers are used to exchange data between two or more points (nodes) on a serial communications network. Reliable data communications can be difficult in the presence of induced noise, ground level differences, and other hazards associated with installation of a network. When communicating at high data rates, or over long distances in real world environments, RS-232 is often inadequate. The differential data transmission of RS-422 and RS-485 offers superior performance in most applications. Differential signals can help nullify the effects of ground shifts and induced noise signals that can appear as common mode voltages on a network. RS-422 was designed for greater distances and higher baud rates than RS-232. In its simplest form, a pair of converters from RS-232 to RS-422 (and back again) can be used to form an "RS-232 extension cord". Data rates of up to 100 kbits/second and distances of 3,937' (1,200 m) can be accommodated with RS-422. RS-422 devices however cannot be used to construct a true multi-point network. A multi-point network consists of multiple drivers and receivers connected on a single bus, where any point (node) can transmit and/or receive data. RS-485 is an enhanced version of the RS-422 standard, which allows multiple drivers and receivers on the same two-wire or four-wire system. The RS-485 standard specifies up to 32 drivers and 32 receivers on a single bus, but with the introduction of "automatic" repeaters and high-impedance drivers/receivers, this number can be extended to hundreds of points (nodes) on a network. Page 6 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 The cabling used for an RS-422 or RS-485 serial communications network should always be a high quality cable such as Belden 8162 or Alpha 6203C. A two-wire system requires two twisted pairs, and a fourwire system requires three twisted pairs (the extra twisted pair is needed for the signal ground). The graphic below illustrates how to connect a general four-wire network (a four-wire network actually contains 5 wires). General Four-Wire Network Connection The graphic below illustrates how to connect a general two-wire network (a two-wire network actually contains 3 wires). Note that the RS-232 to RS-422/485 coverter and USB to RS422/485 converter have DIP switches that allow for two-wire connections without the need to externally wire the DO to the DI and the /DO to the /DI (see the converter section for complete details). General Two-Wire Network Connection A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 7 SIT SITRANS RD300RANS RD300 RD200 RS-232 SERIAL ADAPTER Description The RD200 RS-232 serial adapter converts the serial output of the RD200 meter to an unbalanced, full-duplex RS-232 signal. The RS-232 port has a female DB9 connector with pins 2 (RX output), 3 (TX input), and 5 (Signal Ground). Pins 7 (RTS) and 8 (CTS) are tied together, and pins 1 (CD), 4 (DTR), and 6 (DSR) are tied together. The adapter is powered by the RD200 meter. Baud rates are adjustable and handled by the RD200 (see the RD200 Instruction Manual for more details). The RD200 RS-232 serial adapter has three diagnostic LEDs: a Power (P) LED to show when the adapter is powered properly, a Transmit Data (TX) LED to show when the adapter is sending data out from the PC, and a Receive Data (RX) LED to show when the adapter is receiving data from the RD200 meter. Page 8 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 Installation The graphic below shows the connection of a RD200 meter to a PC using the RD200 RS-232 serial adapter. The RD200 RS-232 serial adapter has an RJ11 connector to connect the RD200 meter copy cable and a PC compatible 9-pin D subminiature connector (DB9). The DB9 can be connected directly to the PC or by using a standard serial extension cable. RS-232 Adapter Serial Cable (Not Provided) Modular Cable (Provided) RS-232 Adapter Connections A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 9 SIT SITRANS RD300RANS RD300 Connections A RD200 meter copy cable is provided to connect the RD200 meter to the RD200 RS-232 serial adapter. The graphic below shows the RD200 meter copy cable details. (WH) 1 1 (BK) 2 2 (R) 3 3 (G) 4 4 (Y) 5 5 (BL) 6 6 Standard Modular Cable Wiring RD200 Serial Communication Address (serl) For the RD200 meter the address may be programmed from 00 to 99 for PDC Protocol and from 1 to 247 for Modbus RTU. The RD200 meter is factory set to PDC Protocol at address 00. To change the RD200 meter address: 1. Press and hold Right arrow and Menu buttons for three seconds to access Advanced Features menu of the meter. 2. Press Up arrow until Serial (SErL) menu is displayed and press Enter/Ack, Addr is displayed. 3. Press Enter/Ack to change meter address using Right and Up arrow buttons. Press Enter/Ack to accept. 4. Press Menu button to exit and return to Run Mode. Page 10 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RD200 RS-485 SERIAL ADAPTER Description The RD200 RS-485 serial adapter converts the serial output of the RD200 meter to balanced, full or half-duplex RS-485 signals. The RD200 RS-485 serial adapter has a removable screw terminal connector for the 485 terminals which includes Transmit Data (DO) and (/DO), Receive Data (DI) and (/DI), and Signal Ground. Baud rates are adjustable and handled by the RD200 (see the RD200 Instruction Manual for more details). The RD200 RS-485 serial adapter has three diagnostic LEDs: a Power (P) LED to show when the adapter is powered properly, a Transmit Data (TX) LED to show when the adapter is sending data out from the PC side, and a Receive Data (RX) LED to show when the adapter is receiving data from the RD200 meter. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 11 SIT SITRANS RD300RANS RD300 Installation The graphic on page 13 shows the connection of a RD200 meter to a PC using the RD200 RS-485 serial adapter and a RS-232 to RS422/485 converter in an RS-422 network. The RD200 RS-485 serial adapter has an RJ11 connector to connect the RD200 meter copy cable and a screw terminal connector to connect to the RS-422 network. The graphic on page 14 shows the connection of several RD200 meters with RD200 RS-485 serial adapters to a PC using a RS-232 to RS-422/485 converter in an RS-485 network. RD200 Serial Communication Address (serl) When using more than one meter in a multi-drop or multi-point mode, each meter must be provided with its own unique address. For the RD200 meter the address may be programmed from 00 to 99 for PDC Protocol and from 1 to 247 for Modbus RTU. The RD200 meter is factory set to PDC Protocol at address 00. To change the RD200 meter address: 1. Press and hold Right arrow and Menu buttons for three seconds to access Advanced Features menu of the meter. 2. Press Up arrow until Serial (SErL) menu is displayed and press Enter/Ack, Addr is displayed. 3. Press Enter/Ack to change meter address using Right and Up arrow buttons. Press Enter/Ack to accept. 4. Press Menu button to exit and return to Run Mode. Page 12 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RS-422 or RS-485 Wiring Notes: 1. Termination resistors are optional and values depend on the cable length and characteristic impedance. Consult the cable manufacturer for recommendations. 2. Refer to RS-232 to RS-422/485 Converter documentation for further details. 3. Use shielded cable, twisted-pairs plus ground. Connect ground shield only at one location. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 13 SIT SITRANS RD300RANS RD300 RS-485 Two-Wire Multi-Drop Wiring Notes: 1. Termination resistors are optional and values depend on the cable length and characteristic impedance. Consult the cable manufacturer for recommendations. 2. Refer to RS-232 to RS-485 Converter documentation for further details. 3. Use shielded cable, twisted-pair plus ground. Connect ground shield only at one location. Page 14 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 Connections A RD200 meter copy cable is provided to connect the RD200 meter to the RD200 RS-485 serial adapter. The graphic below shows the RD200 meter copy cable details. (WH) 1 1 (BK) 2 2 (R) 3 3 (G) 4 4 (Y) 5 5 (BL) 6 6 Standard Modular Cable Wiring The table below details the wiring connections from the RD200 RS-485 serial adapter to an RS-422/485 serial converter (such as the RS-232 to RD-422/485 coverters or USB to RS-422/485 converters) for a four-wire network. RD200 RS-485 serial adapter to RS-422/485 Serial Converter Connections RS-422/485 Serial Converter RS-485 Adapter DO DI DO DI DI DO DI DO Connections for RD200 RS-485 Adapter to Serial Converter If the serial converter is configured for a two-wire network then the requirement to externally wire the DO to the DI and the /DO to the /DI on the RD200 RS-485 serial adapter screw terminal connector is needed. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 15 SIT SITRANS RD300RANS RD300 RD200 TO USB SERIAL ADAPTER Description The RD200 USB serial adapter allows for direct connection of a RD200 meter to the USB port of a PC. Installation The graphic below shows the connection of a RD200 Series meter to a PC using a RD200 to USB Serial Adapter. USB Serial Adapter USB Cable (Provided) Modular Cable (Provided) RD200 USB Adapter Connections Page 16 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RD300 RS-232 SERIAL ADAPTER Description The RD300 RS-232 serial adapter converts the serial output of the RD300 meter to an unbalanced, full-duplex RS-232 signal. The RS-232 port has a female DB9 connector with pins 2 (RX output), 3 (TX input), and 5 (Signal Ground). Pins 7 (RTS) and 8 (CTS) are tied together, and pins 1 (CD), 4 (DTR), and 6 (DSR) are tied together. The adapter is powered by the RD300 meter M-Link connection. Baud rates are adjustable and handled by the RD300 (see the RD300 Instruction Manual for more details). The RD300 RS-232 serial adapter has three diagnostic LEDs: a Power (P) LED to show when the adapter is powered properly, a Transmit Data (TX) LED to show when the adapter is sending data out from the PC side, and a Receive Data (RX) LED to show when the adapter is receiving data from the RD300 meter. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 17 SIT SITRANS RD300RANS RD300 Installation The graphic below shows the connection of a RD300 meter to a PC using the RD300 RS-232 serial adapter. The RD300 RS-232 serial adapter has an RJ45 connector to connect the Cat5e cable and a PC compatible 9-pin D subminiature connector (DB9). The DB9 can be connected directly to the PC or by using a standard serial extension cable. RS-232 Adapter Serial Cable (Not Provided) Cat5e Cable (Provided) RS-232 Adapter Connections Connections A Cat5e cable is provided to connect the RD300 meter to the RD300 RS232 serial adapter. Page 18 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RD300 Serial Communication Address (serial) For the RD300 meter the address may be programmed from 001 to 247. The RD300 meter is factory set to address 001. To change the RD300 meter address: 1. Press and hold the Menu button for three seconds to access Advanced Features menu of the meter. 2. Press Up arrow until Serial (SEriaL) menu is displayed and press Enter, Addres is displayed. 3. Press Enter to change meter address using Right and Up arrow buttons. Press Enter to accept. 4. Press Menu button to exit and return to Run Mode. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 19 SIT SITRANS RD300RANS RD300 RD300 RS-485 SERIAL ADAPTER Description The RD300 RS-485 serial adapter converts the serial output of the RD300 meter to balanced, full or half-duplex RS-485 signals. The RD300 RS-485 serial adapter has a removable screw terminal connector for the RS-485 terminals which includes Transmit Data (DO) and (/DO), Receive Data (DI) and (/DI), and Signal Ground. The adapter is provided by the RD300 meter M-Link connection. Baud rates are adjustable and handled by the RD300 (see the RD300 Instruction Manual for more details). The RD300 RS-485 serial adapter has three diagnostic LEDs: a Power (P) LED to show when the adapter is powered properly, a Transmit Data (TX) LED to show when the adapter is sending data out from the PC side, and a Receive Data (RX) LED to show when the adapter is receiving data from the RD300 meter. Page 20 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 Installation The graphic on page 22 shows the connection of a RD300 meter to a PC using the RD300 RS-485 serial adapter and a RS-232 to RS422/485 converter in an RS-422 network. The RD300 RS-485 serial adapter has an RJ45 connector to connect the Cat5e cable and a screw terminal connector to connect to the RS-422 network. The graphic on page 23 shows the connection of several RD300 meters with RD300 RS-485 serial adapters to a PC using a RS-232 to RS-422/485 converter in an RS-485 network. RD300 Serial Communication Address (serIAl) When using more than one meter in a multi-drop or multi-point mode, each meter must be provided with its own unique address. For the RD300 meter the address may be programmed from 001 to 247. The RD300 meter is factory set to address 001. To change the RD300 meter address: 1. Press and hold the Menu button for three seconds to access Advanced Features menu of the meter. 2. Press Up arrow until Serial (SEriaL) menu is displayed and press Enter, Addres is displayed. 3. Press Enter to change meter address using Right and Up arrow buttons. Press Enter to accept. 4. Press Menu button to exit and return to Run Mode. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 21 SIT SITRANS RD300RANS RD300 RS-422 or RS-485 Wiring Notes: 1. Termination resistors are optional and values depend on the cable length and characteristic impedance. Consult the cable manufacturer for recommendations. 2. Refer to RS-232 to RS-422/485 Converter documentation for further details. 3. Use shielded cable, twisted-pairs plus ground. Connect ground shield only at one location. Page 22 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RS-485 Two-Wire Multi-Drop Wiring Notes: 1. Termination resistors are optional and values depend on the cable length and characteristic impedance. Consult the cable manufacturer for recommendations. 2. Refer to RS-232 to RS-485 Converter documentation for further details. 3. Use shielded cable, twisted-pair plus ground. Connect ground shield only at one location. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 23 SIT SITRANS RD300RANS RD300 Connections A Cat5e cable is provided to connect the RD300 meter to the RD300 RS485 serial adapter. The table below details the wiring connections from the RD300 RS-485 serial adapter to an RS-422/485 serial converter (such as the RS-232 to RD-422/485 coverters or USB to RS-422/485 converters) for a four-wire network. RD300 RS-485 serial adapter to RS-422/485 Serial Converter Connections RS-422/485 Serial Converter RS-485 Adapter DO DI DO DI DI DO DI DO Connections for RS-485 Adapter to Serial Converter If the serial converter is configured for a two-wire network then the requirement to externally wire the DO to the DI and the /DO to the /DI on the RD300 RS-485 serial adapter screw terminal connector is needed. Page 24 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RD300 TO USB SERIAL ADAPTER Description The RD300 USB serial adapter RD300 to USB Serial Adapter allows for direct connection of a RD300 Series meter to the USB port of a PC. Installation The graphic below shows the connection of a RD300 Series meter to a PC using a RD300 to USB Serial Adapter. RD300 USB Adapter Connections A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 25 SIT SITRANS RD300RANS RD300 RD200/300 USB TO RS-232 CONVERTER Description The USB to RS-232 Converter allows for direct connection of a serial device to the USB port of a PC. Installation The graphic on page 27 shows the connection of a RD200 meter to a PC using a USB to RS-232 Converter, RS-232 serial adapter, and the SITRANS RD200/300 meter copy Cable. Page 26 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RS-232 Adapter RS-232 Converter Modular Cable (Provided) Connections for USB to RS-232 converter to serial device A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 27 SIT SITRANS RD300RANS RD300 RD200/300 RS-232 TO RS-422/485 CONVERTER Description The RS-232 to RS-422/485 converter converts unbalanced, full-duplex RS-232 signals to balanced, full or half-duplex RS-422 or RS-485 signals. The RS-232 port, configured as a DTE port, has a female DB9 connector with pins 2 (RX output), 3 (TX input), and 5 (Signal Ground). Pins 7 (RTS) and 8 (CTS) are tied together, and pins 1 (CD), 4 (DTR), and 6 (DSR) are tied together. The RS-485 terminal blocks support Transmit Data (DO) and (/DO), Receive Data (DI) and (/DI), and Signal Ground. A separate terminal block supports the power input (+12VDC) and power ground (GND). Baud rates are automatic and handled by the RS-232 to RS-422/485 converter. The RS-232 to RS-422/485 converter has three diagnostic LEDs: a Power (P) LED to show when the converter is powered properly, a Transmit Data (TX) LED to show when the converter is sending data out from the PC side, and a Receive Data (RX) LED to show when the converter is receiving data from the network side. The DIP switch SW1 is located between the screw terminal connectors and allows for system configurations. Page 28 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 Installation The DIP switch SW1 allows for several different options. Factory settings for the switch are shown below. 1 2 3 4 5 RS-232 to RS-422/485 converter Terminal Connectors and DIP Switch The TERM switch position adds an internal 120 ohm termination resistor when ON (up). Termination should be used on both ends of the network with high data rates and long wiring runs. To configure a two-wire network set both of the 2 WIRE switch positions to ON (up). For a four-wire network set both of the 2 WIRE switches to OFF (down). When the RS422 switch position is ON (up) it is configured for an RS-422 network. For RS-485 networks set the RS422 switch to OFF (down). The ECHO switch position allows the data being sent from the RS-232 port to be echoed back into the RS-232 port. In most networks, this is an undesired effect and the ECHO switch would be OFF (down). RS-232 to RS-422/485 converter DIP Switch Settings SW1 Switch Position ON Setting Function 1 Network Termination Included (120 ohm) 2 2-Wire Network Mode Connection (factory setting) 3 2-Wire Network Mode Connection (factory setting) 4 RS-422 Mode 5 ECHO Mode RS-232 to RS-422/485 converter DIP Switch Settings A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 29 SIT SITRANS RD300RANS RD300 The graphic below shows an example configuration of a two-wire network. Each device (such as an RD200 meter) connected to the same network must have a unique address. RS-485 Two-Wire Multi-Drop Wiring Notes: 1. Termination resistors are optional and values depend on the cable length and characteristic impedance. Consult the cable manufacturer for recommendations. 2. Refer to RD200 RS-485 serial adapter documentation for further details. 3. Use shielded cable, twisted-pair plus ground. Connect ground shield only at one location. Page 30 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 Connections To power the RS-232 to RS-422/485 converter, connect 12 VDC to the +12VDC and ground to the GND screw terminals from the supplied 115 VAC/12 VDC adapter. The RS-232 to RS-422/485 converter may be configured for either a four-wire or two-wire network. The table below details the wiring connections from the RS-232 to RS-422/485 converter to an RS-422/485 serial device in a four-wire network. RS-232 to RS-422/485 converter to RS-422/485 Serial Device Four-Wire Connections RS-232 to RS-422/485 Serial Converter RS-422/485 Serial Device DO DI DO DI DI DO DI DO A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 31 SIT SITRANS RD300RANS RD300 The table below details the wiring connections from the RS-232 to RS422/485 converter to an RS-422/485 serial device in a two-wire network when the DIP switches on the RS-232 to RS-422/485 converter have been set to the "2-Wire Mode". RS-232 to RS-422/485 converter to RS-422/485 Serial Device Two-Wire Connections RS-232 to RS-422/485 Serial Converter RS-422/485 Serial Device DO DATA DO DATA The RS-232 to RS-422/485 converter internally connects the /DI to the /DO and the DI to the DO when the DIP switches are set to "2-Wire Mode". Either the /DI or /DO could be used to connect to /DATA and either the DI or DO could be used to connect to DATA. Two RS-232 to RS-422/485 Serial Converters could be used as a serial extender to allow RS-232 serial devices to communicate over long distances. This would allow an RS-232 device to operate up to 3,937' (1,200 m). Page 32 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 RD200/300 USB TO RS-422/485 CONVERTER Description The USB to RS-422/485 converter converts USB to balanced, full or half-duplex RS-422 or RS-485 signals. The USB to RS-422/485 converter has a removable screw terminal connector for the RS-422/485 terminals which includes Transmit Data (DO) and (/DO), Receive Data (DI) and (/DI), and Signal Ground. Baud rates are automatic and handled by the USB to RS-422/485 converter. The USB to RS-422/485 converter has three diagnostic LEDs: a Power (P) LED to show when the converter is powered properly, a Transmit Data (TX) LED to show when the converter is sending data out from the PC side, and a Receive Data (RX) LED to show when the converter is receiving data from the network side. The DIP switch SW1 is located inside the case and allows for system configurations. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 33 SIT SITRANS RD300RANS RD300 Installation The DIP switch SW1 allows for several different options. Factory settings for the switch are shown below. 1 2 3 4 5 USB to RS-422/485 converter DIP Switch Location The termination switch position adds an internal 120 ohm termination resistor when ON (up). Termination should be used on both ends of the network with high data rates and long wiring runs. To configure a two-wire network set both of the 2-wire switch positions to ON (up). For a four-wire network set both of the 2-wire switches to OFF (down). When the RS-422 switch position is ON (up) and the RS-485 switch position is OFF (down) it is configured for an RS-422 network. When the RS-422 switch position is OFF (down) and the RS-485 switch position is ON (up) it is configured for an RS-485 network. USB to RS-422/485 converter DIP Switch Settings SW1 Switch Position ON Setting Function 1 Network Termination Included (120 ohm) 2 2-Wire Network Mode Connection (factory setting) 3 2-Wire Network Mode Connection (factory setting) 4 RS-422 Mode 5 RS-485 Mode (factory setting) USB to RS-422/485 converter DIP Switch Settings Page 34 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 Connections The USB to RS-422/485 converter is powered by the USB port. The USB to RS-422/485 converter can be configured for either a fourwire or two-wire network. The table below details the wiring connections from the USB to RS-422/485 converter to an RS-422/485 serial device in a four-wire network. USB to RS-422/485 converter to RS-422/485 Serial Device Four-Wire Connections USB to RS-422/485 converter Serial Converter RS-422/485 Serial Device DO DI DO DI DI DO DI DO The table below details the wiring connections from the USB to RS422/485 converter to an RS-422/485 serial device in a two-wire network when the DIP switches on the USB to RS-422/485 converter have been set to the "2-Wire Mode". USB to RS-422/485 converter to RS-422/485 Serial Device Two-Wire Connections USB to RS-422/485 converter Serial Converter RS-422/485 Serial Device DO DATA DO DATA The USB to RS-422/485 converter internally connects the /DI to the /DO and the DI to the DO when the DIP switches are set to "2-Wire Mode". Either the /DI or /DO could be used to connect to /DATA and either the DI or DO could be used to connect to DATA. A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 35 SIT SITRANS RD300RANS RD300 The graphic below shows an example configuration of a two-wire network. Each device (such as an RD200 meter) connected to the same network must have a unique address. USB Connection to PC Modular Cable to Meter (Provided) USB to RS-422/485 Converter DO DO RS-422/485 Adapter DI DI Adapter #1 Last Adapter R Shield Twisted-Pair DI DI DO DO DI DI DO DO R R = 120 Ohm termination resistor, typical RS-485 Two-Wire Multi-Drop Wiring Notes: 1. Termination resistors are optional and values depend on the cable length and characteristic impedance. Consult the cable manufacturer for recommendations. 2. Refer to RD200 RS-485 serial adapter documentation for further details. 3. Use shielded cable, twisted-pair plus ground. Connect ground shield only at one location. Page 36 COMMS ACCESSORIES - OPERATING INSTRUCTIONS A5E31979195 SITRANS RD300 TROUBLESHOOTING Symptom Check/Action Power LED is off RS-232 or RS-485 serial adapters: 1. Check modular cable connection 2. Check power to meter RS-232 to RS-422/485 converter: 1. Check wall plug adapter output 2. Check power connection polarity USB to RS-422/485 converter, USB serial adapters: 1. Check USB connections 2. Try different USB port 3. Check USB port with other device Check: 1. Serial adapter and cable 2. Serial protocol selected 3. Meter address and baud rate 4. RD Software address and baud rate 5. Check DIP switch setting on the RS-232 to RS-422/485 converter or USB to RS422/485 converter Check: 1. Serial adapter and cable 2. Serial protocol selected 3. Meter address and baud rate 4. RD Software address and baud rate 5. Check DIP switch setting on the RS-232 to RS-422/485 converter or USB to RS422/485 converter Remove all unnecessary cables and meters. Try getting the system to work with only one meter (to ease troubleshooting) and then expand the system one device at a time. Meter not communicating with SITRANS RD Software or other programs If only the TX (or DATA IN) data status LED is flashing when serial communications attempted If both data status LEDs (TX and RX) are off when trying to communicate Communications slow Random communication errors Other symptoms not described above RS-232 serial adapters, or RS-232 to RS-422/485 converter: 1. Check serial cable 2. Connect the DB9 directly to the PC 3. Try a different serial port Increase the baud rate 1. Increase the TX delay time 2. Decrease the baud rate Call your local Siemens representative for assistance A5E31979195 COMMS ACCESSORIES - OPERATING INSTRUCTIONS Page 37 Notes.fm Page 1 Thursday, October 11, 2001 8:48 AM Notes For more information www.siemens.com/level www.siemens.com/weighing Siemens AG Industry Sector 1954 Technology Drive P.O. Box 4225 Peterborough, ON Canada K9J7B1 Subject to change without prior notice A5E31979195 Rev. AA (c) Siemens AG 2013 email: techpubs.smpi@siemens.com *A5E31979195* Printed in Canada www.siemens.com/processautomation LIM7232SI_A 05/13