Application Note
From the Fluke Digital Library @ www.fluke.com/library
Basics of MRI
NMR is based on the fact that
the atoms of many chemical ele-
ments and their isotopes have
some degree of magnetic orien-
tation. Under normal conditions,
the individual magnetic fields of
individual atoms show a random
orientation, and the net field is
zero (Figure 1).
When placed in a very strong
static magnetic field, the mag-
netic fields of the individual
atoms line up, resulting in a net
magnetization (Figure 2). The
magnetic fields used to accom-
plish this are around 0.1 to 4 tesla
(1 T = 10000 gauss) with a ten-
dency to the ever stronger fields.
Once the nuclei are aligned
by this static magnetic field, an
alternating magnetic field in a
direction perpendicular to the
In modern medicine, Magnetic Resonance Imaging or
“MRI” is one of the techniques used to study the inner
structure of a living body on a non-invasive basis. The
technique is an application of Nuclear Magnetic
Resonance or NMR which is also used beyond the
medical field to study the structure of matter. The main
advantage of MRI over X-ray based techniques is in the
better imaging capability of soft tissue, where X-ray
techniques are more useful for study of the bone
structure, for instance in case of fractures.
main field will make the protons
precess around the axis of their
static orientation.
When this occurs, the atoms
turn out to have a resonance
frequency in the RF-range. For
instance the hydrogen atom (1H)
has a resonance frequency of
approximately 42.58 MHz when
placed in a static magnetic field
of 1 T.
Given this resonance, the
nuclei can be thought of as res-
onators that absorb energy from
the applied RF-field.
Nuclei of different elements
and different isotopes all have
different resonance frequencies.
For as long as the RF-field is
applied, all nuclei will resonate
in phase, but once this stimulus
ends the individual nuclei will
lose synchronism, or de-phase,
MRI adjustments with
a Fluke ScopeMeter®
190 Series
Figure 1: Normal randomly arranged
magnetic orientation of nuclei.
Figure 2: When placed in a strong magnetic
field, the magnetic orientation of nuclei align
with the external field.
and return to their original posi-
tion aligned with the static mag-
netic field. This process of
returning to the original orienta-
tion takes a length of time that
is again characteristic of each
different type of nucleus.
By measuring the resonance
frequencies and the decay times,
the individual elements present
in the body can be determined.
Oscilloscopes used
in MRI
For the proper functioning of the
MRI-equipment, it is essential
that the RF stimuli are switched
on and off in a well controlled
manner, and with good repro-
ducibility. During development
and adjustment of the equip-
ment, a proper representation of
these RF-bursts is most impor-
2 Fluke Corporation MRI adjustments with a Fluke ScopeMeter®190 Series
tant. For this purpose an oscillo-
scope is a most valuable piece of
test equipment, if it can display
the RF carrier frequency and is
capable of showing the envelopes
of the bursts properly. These
envelopes show you everything
that needs to be known about
the process of activating and
terminating the RF bursts.
Many oscilloscopes, however,
have difficulty working properly
in the vicinity of the MRI-equip-
ment because of the extremely
high static magnetic fields.
By nature, the CRT of the
oscilloscope is sensitive to mag-
netic influences as it is based
upon electrons travelling
through the vacuum of the CRT.
The Fluke ScopeMeters are all
able to operate as compact, high
performance oscilloscope that
are well equipped to acquire
and display electrical signals of
various frequencies and that are
not influenced by magnetic
fields (in contrast to a traditional
CRT-based oscilloscope) since
the display used is an LCD.
ScopeMeters used
in MRI
The Fluke ScopeMeters are
available in different bandwidth
and with different sample rates
(see table 1). Given the RF-
frequencies commonly used in
MRI of 42.58, 63.8 and 127
MHz, the Fluke 190 Series
ScopeMeters are the most
appropriate instruments for
test and alignment purposes.
All Fluke ScopeMeters are
equipped with a function called
‘Glitch Detect’. This function
activates a digital overscan
mechanism to find high-fre-
quency signal content even
when working at low speed
timebase settings. An example
of this is given in Figure 4.
At higher timebase speed, the
number of cycles that is con-
tained in an individual burst is
quite limited, and it becomes
more difficult to recognize the
overall envelope of the signal
(see Figure 5). Furthermore, at
higher RF-carrier frequencies,
the effects of aliasing may also
influence the quality of the dis-
played bursts.
Here, some of the additional
functions found in the ScopeMe-
ters come in handy to enhance
the quality of the displayed
waveform.
Envelope function
The Fluke 190B Series ScopeMe-
ters are equipped with a so-
called “Envelope”- function that
builds the envelope over succes-
sive waveform-acquisitions on
the display. See Figure 6. For
this, the extreme values (that is
the most positive and the most
negative value for each of the
horizontal positions over succes-
sive traces) are stored, and
these are overwritten only when
newer values are found in a
next trace that exceed the ear-
lier ones at the same position.
After just a few acquired
traces, the envelop function
shows a good representation of
the overall envelope of the
burst. On top of the envelope
pattern, the most recently
acquired waveform is then also
displayed.
Digital persistence
Next to the envelope function,
the Fluke 190C Series of Color
ScopeMeters also includes a so-
called ‘Digital Persistence’ mode.
In that mode, individual traces
are stored in a dedicated display
memory and each waveform
fades away slowly. The decay
time, or fade-out time, can be
selected for short, medium or
longer persistence, or it can be
switched to have an ‘infinite
persistence’.
Figure 3: When placed in a static magnetic
field B0under the influence of an alternating
field perpendicular to the static field, the
nuclei start precessing.
Fluke type number Bandwidth Max. sample rate (single shot)
Fluke 123 20 MHz 25 MS/s
Fluke 124 40 MHz 25 MS/s
Fluke 192B 60 MHz 500 MS/s
Fluke 196B, 196C, 196BM, 196CM 100 MHz 1 GS/s
Fluke 199B, 199C, 199BM, 199CM 200 MHz 2.5 GS/s
Table 1: ScopeMeter models and their bandwidth.
Figure 4: At lower timebase setting, the ‘glitch
detect’ function builds a solid envelope of a
high frequency burst (Fluke 199C).
Figure 5: A short duration burst is captured,
but the envelope is more difficult to identify
(Fluke 199C).
Figure 6: The envelope function of the Fluke
199B is used here with the same signal as in
Figure 5.
3 Fluke Corporation MRI adjustments with a Fluke ScopeMeter®190 Series
The resulting display shows a
number of successive wave-
forms, superimposed and at dif-
ferent intensity levels in which
the most recent ones stand out
more clearly. See Figure 7.
When used on burst signals,
as encountered in MRI-equip-
ment, this Digital Persistence
gives the display quality one
would expect from a traditional
high-frequency oscilloscope
with its phosphor ensuring the
slow trace fade-out.
An additional benefit of the
Digital Persistence mode is that
it automatically adapts the
waveform representation if the
overall amplitude drops, e.g. as
a result of an alignment made to
the equipment under test.
The envelop mode, on the
other hand, will continue to
show the maximum amplitude
that the envelope has ever
attained since the functions was
initiated. Should a new ampli-
tude of the waveform be seen or
expected, a simple key-press
(e.g. of the attenuator, 1 step up
and back) will delete the cre-
ated envelope and start building
a new one.
Conclusion
The Fluke 190 Series Scope-
Meters are well equipped to
capture and display the short-
duration bursts of RF-energy as
used in the MRI-medical imag-
ing equipment. On top of the
basic functionality of these high
performance handheld oscillo-
scopes, some additional func-
tions are built-in that help to
optimally display the waveforms
of such short duration bursts.
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-51
16
Web access: http://www.fluke.com
©2005 Fluke Corporation. All rights reserved.
Printed in U.S.A. 10/2005 2543288 A-EN-N Rev A
Fluke. Keeping your world
up and running.
Figure 7: The Fluke 199C is used here with
‘Digital Persistence’ to capture a natural enve-
lope of a short duration burst of an RF-carrier.
ScopeMeters for medical equipment
applications
The Fluke ScopeMeter 190M Series offer addi-
tional capabilities targeting the service engineer
dealing with Medical Imaging Equipment and
Video Display Systems. These models have the
following additional capabilities :
Current-over-time measurement giving
mAs read-out
Using the cursors you can measure directly the
amount of radiation produced by X-ray systems
or the total amount of charge supplied to
a system.
Extended video triggering
The 190M Series supports triggering on high-
resolution non-interlaced video systems.
Smart averaging
Smart averaging shows the averaged curve of
repeated signal parts, as well as incidental
deviations.
Extended offset
Allows vertical zoom-in for study of small details
of the larger amplitude signal or when a larger
DC-component is included.
Medical models available are the Fluke 199CM,
Fluke 199BM, Fluke 196CM and Fluke 196BM.
See techical data sheet for full details.