a
Mounting Considerations for ADXL Series Accelerometers
by Mike Shuster, Bob Briano and Charles Kitchin
ONE TECHNOLOGY WAY
•
P.O. BOX 9106
•
NORWOOD, MASSACHUSETTS 02062-9106
•
617/329-4700
As with all accelerometers, optimum performance
depends on proper mounting of the device. It must
be mounted so that the sensor is properly coupled to
the object for which acceleration is to be measured.
Also the effects of mechanical resonances must be
minimized.
All mechanical structures, no matter how ”solid,“ have
mechanical resonances that result in an output signal at
the resonant frequency that is larger than expected. Me-
chanical resonances cannot be eliminated completely.
The key to successfully reducing any serious problems
is to keep the resonant frequencies above the frequency
band of interest or to attenuate them to acceptable lev-
els. Structures that seem benign at rest often mechani-
cally resonate when vibrated at high frequency.
The ADXL Series of accelerometers are intended for PC
board mounting, and careful mechanical design of the
PC board is important. Resonances can be difficult to
determine before prototyping, but some precautions
can be taken.
1. Replace or firmly attach any individual components
that are flexible such as large through-hole compo-
nents. Surface mount resistors and capacitors are
preferred because they will not vibrate on their leads.
2. Wire-wound resistors and electrolytic capacitors can
change values when they are vibrated. Again, ce-
ramic surface mount components are best.
3. Mechanically couple a potentially problematic struc-
ture to one that has a high resonant frequency. An
example of this is to firmly attach PC boards to thick
bases.
4. Couple together several structures such that the com-
bination has a dampened response: PC boards, card
cages, etc.
5. Cables inside moving structures should be tied down
firmly. Cables between a moving structure and a sta-
tionary one should be flexible and have a service
loop.
AN-379
APPLICATION NOTE
6. Cavities, such as housings, can resonate. Filling cavi-
ties with suitable potting materials or epoxy is a good
solution for this, and also prevents components from
vibrating.
7. Keep all dimensions as small as possible because, in
general, smaller dimensions mean that the reso-
nances are at higher frequencies.
8. Use silicon grease or other materials on mating sur-
faces to improve coupling between components that
are screwed together.
Many applications do not have strict requirements for
being resonance free, but in some cases resonances can
be critical to a successful system. The best way to test a
design for resonances is to shake it, using a calibrated
shaker, over the frequency band of interest using a refer-
ence accelerometer attached at critical points.
Figure 1 illustrates the ideal response of the ADXL50
which is achievable with proper fixturing or potting ma-
terial. The sensor response is flat from dc to the
dominant pole set by the demodulator capacitor, typi-
cally at 1.3 kHz. Figure 2 shows the response of the
ADXL05.
NORMALIZED SENSITIVITY – dB
FREQUENCY – Hz
6
–21
1 10k10 100 1k
3
0
–3
–6
–9
–12
–15
–18
–24
Figure 1. Ideal Response of the ADXL50