* Some options may alter transducer performance and/or mechanical characteristics.
www.trans-metrics.com email: sales@trans-metrics.com
trans metrics 5325 Naiman Pkwy. • Solon, Ohio 44139 • Phone: 440-248-2229 • 888-782-2229 • Fax: 440-248-7780
trans metrics, Division of United Electric Controls
Selected options appear in the 'OP-
TION / PRICING' Guide. If a desired
option is not listed, or if multiple options
are required, please call the factory to
discuss your requirements.
Ordering
Specify series (B020
see back
),
circuit type (A, B, C, D, L, T), pres-
sure connection (B, C, G, H, etc...),
electrical connection (A, B, etc...), op-
tions (xxx), pressure range (xxxxx psi
-
can be units other than psi
) pressure
reference ( a - absolute).
Examples: B020CGA 12-16 psia,
B020CIC 700 - 1100 mbar absolute,
B020THD 27-32 in Hg absolute
See the B020 'OPTION / PRICING'
Guide for detailed information.
Applications
The Trans Metrics' B020 pressure
transducer is made specifically for
barometric pressure applications which
require a high output signal combined
with high reliability, durability, and ac-
curacy.
Design
B020 models incorporate four major
design elements that allow them to
measure pressure under the above
conditions: a beam and diaphragm sen-
sor (for low pressure capabilities),
bonded foil strain gages configured in
a Wheatstone bridge (for accuracy,
reliability, and stability), integral elec-
tronics (for amplified output), and stain-
less steel construction (for durability and
corrosion resistance).
Pressure References
B020 models are available in an ab-
solute pressure reference.
Output/Circuit Types
Three output types (0-5 VDC, 1-5
VDC, 4-20mA) and six circuit types (A,
B, C, D, L, T) are available for stan-
dard B020 transducers. Circuits are
two, three or four wire (circuit diagrams
and specifications can be found on the
back).
Calibration and Compensation
Trans Metrics individually calibrates
and temperature compensates each
B020. Transducers are calibrated
against standards traceable to the Na-
tional Institute of Standards and Tech-
nology. A certificate of calibration is
supplied with each unit.
Options/Modifications*
Numerous options and modifica-
tions are available and are designated
by an option number (xxx) after the
model number.
DESCRIPTION:
Barometric Pressure Transducer
B020 Series
FEATURES:
High accuracy
12-16 PSIA pressure range (others available)
0-5 VDC, 1-5 VDC or 4-20 mA output
2, 3 or 4-wire circuitry
Integral electronics
Absolute pressure reference
Field-adjustable span and zero
Reverse polarity protection
Over-pressure stop (optional)
Stainless steel construction
CE mark (amplified voltage units only)
Printed in U.S.A. B020 02/2002 Page 1 of 3
CIRCUITS/DIMENSIONS:
SPECIFICATIONS:
Specifications are subject to change without notice
www.trans-metrics.com email: sales@trans-metrics.com
trans metrics 5325 Naiman Pkwy. • Solon, Ohio 44139 • Phone: 440-248-2229 • 888-782-2229 • Fax: 440-248-7780
Model: B020
Pressure Ranges: 12-16 PSIA (others available)
Output Ranges: 0-5 VDC ±0.5% (A, B, C, L)
(Field-Adjustable ±5% typical)
1-5 VDC ±0.5% (D)
4-20 mA ±0.5% (T)
Zero Balance: ±0.5% FSO (A, B, C, D, L, T)
(Field-Adjustable ±5% typical)
Static Error Band: ±0.10% FSO (Max.)
(BSL - Nonlinearity, Hysteresis,
and Nonrepeatability combined)
Thermal Zero Shift: ±0.005% FSO/°F
±0.009% FSO/°C
Thermal Sensitivity Shift: ±0.005%/°F
±0.009%/°C
Operating Temperature -30°F to 180°F
Range:
-35°C to 82°C
Compensated Temperature 0°F to 160°F
Range:
-20°C to 70°C
Max. Safe Exposure Temp.: +250°F,
+125°C
Current Consumption (Typical):25 mA (A, B, C, D)
4-20mA (T)
2 mA (L)
Excitation: 15 to 32 VDC (A)
9 to 28 VDC (B, C, D)
4 to 28 VDC (L)
9 to 40 VDC (T)
(see load vs.excitation curve)
Important Note: T type
circuit 4-20 mA has over-
scale limit of 34mA
Rise Time (10 to 90%):Less than 1 ms typical
Resolution: Continuous
Proof Pressure: 2 times rated pressure range
5 times rated pressure range
(With Overpressure stop)
Burst Pressure: 10 times rated pressure range
Pressure Cavity Volume: 0.1 in3,
1.6 cm
3
(does not
include threaded area)
Material: 15-5 PH stainless steel (sensor)
300 series stain. steel (housing)
Weight: Approximately 14 oz. or
400 g
Identification: Model, range, serial #, connec-
tions, manufacturer & country
of origin are inscribed on the
case.
Operational Details: Reference B020 instruction
manual.
B020 SERIES
3.15 in
8.0 cm
2.00 in
5.1 cm
0.75 in
0.5 in
1.3 cm
0.93 in
2.4 cm
0.75
1.9 cm
Power
Supply
Readout
Transducer
low
high
- signal
Rd/A
Bk/D
Wt/C
+ signal
Gn/B
E F
cal/test, use optional
B020 A, B
(4-wire)
Circuit
Power
Supply
Readout
Transducer
common
high
signal
Rd/A
Bk/C&D
B cal/test
(C/D)
, use opt.
B020 C, D & L
(3-wire)
Circuit
Power
Supply
Readout
Transducer
common
A
B
-
+
cal/test, use optional
B020 T Circuit
(2-wire)
Or/F
+_
1.03 in
2.6 cm
Pressure Connection
*
Electrical Connection
See
OPTIONS/PRICING
guide for availability of pressure and electrical connections
**
010203040
1600
1400
1200
1000
800
600
400
200
Excitation, VDC (V+)
Load Resistance, (R
L
)
9v
1550
R
L
max. = (V
+
) - 9V
20mA
Operating
Range
Page 2 of 3
OPTIONS / PRICING:
Trans Metrics' B020 series model numbers are constructed as a series of numbers and letters that identify the accuracy, electrical
circuit, pressure connection, electrical connection, and any options or features which may be unique to a particular pressure
transducer.
The model number below features a B020 series pressure transducer, 0-5 VDC output, 7/16-20 female thread and a
PTIH-10-6P electrical connection. Any other options selected would be assigned a three (3) digit number which would be added
to the end of the model number. In this example, 017 represents 0-10 VDC signal output.
Pressure selections should be specified including the pressure range, units and pressure reference.
AC
E
D
B
The chart below will assist you in selecting a transducer configuration and obtaining pricing information.
example:
F G H
12-16 psi a
B020 C G A 017
m
o
d
e
l
Pressure
Reference
H
Units
G
Pressure
Range
F
Options
E
Electrical
Connection
D
Pressure
Connection
C
Circuit Type
B
Series/Accuracy
A
r
a
n
g
e
B020 SERIES
If you need an option not listed above, please consult factory for availability
SEB Static Error Band
BSL: Best Straight Line
www.trans-metrics.com email: sales@trans-metrics.com
trans metrics 5325 Naiman Pkwy. • Solon, Ohio 44139 • Phone: 440-248-2229 • 888-782-2229 • Fax: 440-248-7780
Printed in U.S.A. B020 02/2002 Page 3 of 3
M ode l Se lectio ns
ASeries / Accuracy
B020 B020 Series with ±0.10% SEB BSL
BCircuit Type
A 0-5 VDC Signal O utput (Differential, Exc = 15-32)
B 0-5 VDC Signal O utput (Differential, Exc = 9-28)
C 0-5 VDC Signal Ou tput (Single E nded)
D 1-5 VDC S ignal Ou tput
L 0-5 VD C Signal O utput (Low C urrent)
T 4-20 m A Signal Output
CPressure Connection
B 7/16-20 37° male (for ¼" tub e)
C 7/16-20 fem. M S33649-4 (for ¼" tube)
G 7/16-20 SAE fema le (for ¼ " tube)
H 1/4-18 N PT fem ale
I 1/4-18 NPT ma le
J 1/8-27 NPT ma le
V 1/4 in. VCR fitting (fem ale)
W 1/4 in. VCR fitting (male)
DElectrical Connection
A PTIH-10-6P (mate #80002, sold sep arately $29)
B P CIH -10-6P (mate #80001, so ld separately $48)
C Cable 1 me ter 28 AWG PVC
D Cable 1 meter 24 AW G Teflon®
F Flying Leads 1 meter 24 AWG Teflon®
I M ini-Hirschma nn (DIN 43650-C, m ate in clude d)
Model Selections (continued)
ECom mon Options / M odifications
017 0 - 10 VDC FS O (C Circuit)
178 1 - 6 VDC FSO (D circuit)
253 Over-pressure stop
### A dditional cable lengths / types
### Special wiring (specify)
Pressure Selections
FPressure R ange*
min. 13.5 to 15.5 psia
< > we accomo date any range in between
ma x. 10 to 16 psia
GU n its [ Availa b le P re s su re R an g e]
psi 12-16 p sia, others available
bar 700 -1,100 m bar absolute, others available
kg/cm ² 0.85 - 1.13kg/cm², others available
KPa 70-110 KPa , others available
in Hg 27-31 inHg absolute, o thers available
oth er consult factory
HPressure R eferenc e
absolute Reference to a vacuum