6100A 6180A MODELS Ideal for : 21CFR Part II Applications Furnaces, NADCAP Water/waste water Power generation Sterilizers, autoclaves Features : Colour touchscreen display USB `plug & play' Up to 48 Universal Inputs Up to 96MB non-volatile Flash memory 125ms Parallel sampling Compact Flash or Secure Digital Card Modbus master Ethernet TCP/IP Paperless Graphic Recorders Specification Sheet The 6000 Series offer unrivalled input accuracy with a 125ms total sample rate for up to 48 input channels. Input channels are freely configurable to suit your process requirements. Each instrument has an intuitive, touch screen display to enable operators to clearly view process data in varying formats. All have onboard Flash data storage capability, Ethernet communication and choice of Compact Flash or SD Card. Data is stored in a tamper-resistant binary format that can be used for secure, long term records of your process. The 6000 Series is truly designed for todays networked world and can be accessed via a Local Area Network, dial-up connection, Intranet or Internet. Available Features 6100A 6180A 5.5" 1/4 VGA 12.1" XGA Channels 18 48 Relays 16 36 Display Events Inputs Groups Auditor Features Virtual channels* 24 (6 per option card) 6 standard (12 optional) Lite or Full available 36, 96, 128 Timers Fitted as standard Alarms 4 per channel Batch Bridge- Remote Optional Lite as standard (Full optional) viewing software Screen builder Security 24 (optional) Unlimited unique user names with configurable access permissions and passwords Configuration software Standard Review/Quickchart Lite Standard software Standard views Vertical and horizontal trending, Vertical and Horizontal Bargraphs, Circular Trend and Numeric Values * Virtual channels can be configured as Maths, Totalisers, Counters or Comms *EUROTHERM FLEXIBLE SOLUTIONS* Data Logging and Archiving Dynamic Host Configuration Protocol (DHCP) The 6000 Series recorders have internal Flash memory for secure data storage. They are also able to accept various removable media types (compact flash, SD card or USB memory stick). Data stored within the internal memory can be archived to the removable media on demand or at preset intervals. The 6000 will give indication of how long its internal memory and that of the removable media installed will last according to the configuration of the recorder. Dynamic Host Configuration Protocol, the successor to BootP, allows a 6000 Series host to obtain Network parameters, such as IP address, Subnet Mask, default gateway and DNS server address dynamically. The implementation of DHCP on the 6000 Series significantly reduces the overhead for maintaining a network of instrumentation. All 6000s have Ethernet capability. The 6000 can be configured to archive to the removable media and / or over Ethernet. Archiving files over Ethernet effectively gives a secure, infinite archiving capacity. Approximate duration for continuous recording of one Group of six channels, high compression: Archive Media 0.125s 0.5s 1s Sample Rate 5s 10s 30s 60s 32Mb Internal Flash (approx. 4 million samples) 2.83 days 11.3 days 22.6 days 113 days 226 days 1.86 yrs 3.7 yrs 96Mb Internal Flash (approx. 12 million samples) 8.5 days 33.98 days 67.9 days 339 days 1.86 yrs 5.5 yrs 11 yrs 64Mb CF/SD Card or USB memory stick (approx. 8 million samples) 5.66 days 22.6 days 45.3 days 226 days 1.2 yrs 3.7 yrs 7.4 yrs 256Mb CF/SD Card or USB memory stick (approx, 32 million samples) 22.6 days 90.6 days 181 days 1Gb CF/SD Card or USB memory stick (approx. 125 million samples) Ethernet (FTP Server) 88 Days TECHNICAL SPECIFICATION Recorder Environmental performance Temperature limits Operation: Storage: Humidity limits Operation: Storage: Protection Bezel and display: Sleeve: 6100A Portable case option: Shock Vibration (10 to 150Hz) Altitude 0 to +50C -20 to 60C 5% to 80% RH 5% to 90% RH IP65 IP20 IP21 BS EN61010 BSEN60873, Section 9,18 <2000 metres. Approvals Electromagnetic compatibility CE, cUL (EMC) UL file number e57766 Emissions and immunity BS EN61326 2.4 yrs 4.9 yrs 14.8 yrs 20 yrs Electrical safety (BS EN61010) 354 days 1.9 yrs 9.6 yrs 19 yrs 58 yrs 116 yrs Infinite Installation cat. II; Pollution degree 2 INSTALLATION CATEGORY II The rate impulse voltage for equipment on nominal 230V mains is 2500V. POLLUTION DEGREE 2 Normally, only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation shall be expected Time Synchronisation (SNTP) The 6000 Series support Simple Network Time Protocol which, when enabled, updates the instrument time every 15 minutes from the configured SNTP server. The unit can also act as a Unicast SNTP server on the network, allowing client instruments to synchronise with the 6000 to a resolution of one millisecond. Batch Recording Up to six user-defined fields can be used to enter batch specific data. Field Descriptor Operator entered batch information - up to 20 characters - up to 60 characters The user can choose to log any number of the given fields on start and / or stop of a batch. The information will appear on the chart as a message and cannot be separated from the process data to which it relates. Auditor Features Designed to meet the requirements of the FDA Regulation 21 CFR Part 11 for Electronic Records and Signatures, this software option provides the 6000 Series with additional security such as password ageing, electronic signatures and time stamped audit trail. Modbus Master Allows users to view data from multiple instruments connected either by a local Network connection using Modbus TCP, or a Serial connection using Modbus RTU, Event Input The Event Input option offers six isolated event input circuits per board fitted. Triggered externally these discrete inputs can be used to initiate internal actions within the 6000 Series Paperless Graphic Recorder. For example they could be used to remotely start or stop a Batch. ASCII Printer Output (Reports) Fitted as standard the ASCII text printer option provides the 6000 Series with the ability to generate up to 10 simple reports that can be directed to a Serial ASCII text printer. Reports, triggered by an event/job can be configured to contain parameters such as time and date, batch names, process values and user defined messages. Physical Panel mounting Panel mounting angle 6100A Bezel size: Panel cutout dimensions: Depth behind bezel rear face: Weight: 6180A Bezel size: Panel cutout dimensions: Depth behind bezel rear face: Weight: Operator interface Type Size and resolution DIN43700 45 144 x 144mm. 138 x 138mm (both -0/+1mm) 246.5mm (284 LTC) 3kg max. (5kg if fitted in portable case) 292 x 292mm 281 x 281mm (both -0/+1mm) 261mm 7kg max. Colour TFT LCD with cold cathode backlight, fitted with resistive, analogue, Touch-Panel Model 6100A: 1/4VGA (320 x 240 pixels) 5.5" Model 6180A: XGA (1024 x 768 pixels) 12.1" Power requirements Supply voltage Standard: 85 to 265V ac; 47 to 63Hz or 110 to 370V dc Low voltage option: 20 to 42V RMS;45 to 400Hz or 20 to 54V dc Power (Max) 60VA (Inrush current 36A) Fuse type None Interrupt protection: Standard: Holdup >200msec, at 240V ac, with full load Low voltage option: 20msec at 20V dc or RMS, with full load Back-up Battery Type Support time (RTC) Replacement period Stored data Ethernet communications Type (IEEE802.3) Protocols MODBUS, SMTP, ICMP Cable Type: Maximum length: Termination: Poly-carbonmonofluoride/lithium (BR2330) Part No. PA261095 1 year min. with recorder unpowered 3 years Time; date; values for totalisers, counters and timers; batch data; Fvalue, Rolling average, Stopwatch etc. 10/100baseT Ethernet. TCP/IP, FTP, DHCP, BOOTP, SNTP, CAT5 100 metres RJ45 Serial Communications Option No of Ports Protocol Isolation (dc to 65Hz BS EN61010) Terminals to ground Transmission standard 2 ASCII (typical applications: Input of ASCII string inputs from Barcode readers, Credit card readers etc.) ASCII printer support Modbus RTU Master and Slave Installation category II; Pollution degree 2 50V RMS or dc (basic insulation) EIA232 or EIA485 (software selectable) Thermocouple data Temperature scale Bias current Cold junction types CJ error CJ rejection ratio Upscale/downscale drive ITS 90 0.05 nA Off, internal, external, remote 1C max with inst. at 25C 50:1 minimum High, low or none selectable for each thermocouple channel Additional error: 0.01C (typ.) if high or low selected Types and ranges See Table 3 T/C Type Overall range (C) B 0 to +1820 Input Board General Input types Input type mix Maximum number of inputs A/D conversion method Input ranges Termination Noise rejection (48 to 62 Hz) Maximum common mode voltage Maximum series mode voltage dc Volts, dc millivolts, dc milliamps (with shunt), Thermocouple, 2/3-wire RTD Contact closure (not Channels 1, 7, 13, 19, 25, 31, 37, 43) >60 ms Freely configurable. 6 per board >16 bits, 2nd order delta sigma See Table1 and Table 2 below. Edge connector / terminal block Common mode: >140dB (channel to channel and channel to ground). Series mode: >60dB. 250 Volts continuous 45mV at lowest range; 23.74 Volts peak at highest range. Isolation Channel to channel: Channel to common electronics: Channel to ground: Dielectric strength (BS EN61010) Channel to channel: Channel to ground: Insulation resistance Input impedance Over voltage protection Open circuit detection Recognition time Minimum break resistance Update/archive rates Input/Relay-output sample rate Trend update Archive sample-value Display value DC Input ranges Shunt 300V RMS or dc (double insulation) 300V RMS or dc (double insulation) 300V RMS or dc (basic insulation) (1 minute type tests) 2500V ac 1500V ac >10M at 500 V dc 38mV, 150 mV, 1 V ranges: >10M; 20V range: 65.3k 50 Volts peak (150V with attenuator) 57nA max. 500msec 10M 8Hz 8Hz maximum Latest value at archive time Latest value at display update time (8Hz) C D E G2 J K L N R S T U NiMo/NiCo Ni/NiMo Platinel Pt20%Rh/ Pt40%Rh High Range -38mV 38mV -150mV 150mV -1V 1V -20V 20V Max linearisation error 0to400C=1.7C 400to1820C=0.03C 0.12C 0.08C 0.03C 0.07C 0.02C 0.04C 0.02C 0.04C 0.04C 0.04C 0.02C 0.08C 0.06C 0.14C 0.02C 0.07C Table 3 Thermocouple types and ranges Resistance inputs Ranges (including lead resistance) Influence of lead resistance Error: Mismatch: Temperature scale Accuracy and resolution RTD types and ranges RTD Type Overall range (C) Cu10 Cu53 JPT100 Ni100 Ni120 Pt100 Pt100A Pt1000 -20 to +400 -70 to 200 -220 to +630 -60 to +250 -50 to +170 -200 to +850 -200 to +600 -200 to +850 0 to 150, 0 to 600, 0 to 6k Negligible 1/ ITS90 See Table 2 See Table 4 Standard Max linearisation error General Electric Co. RC21-4-1966 JIS C1604:1989 DIN43760:1987 DIN43760:1987 IEC 751 Eurotherm Recorders SA IEC 751 0.02 C <0.01C 0.01 C 0.01 C 0.01 C 0.01 C 0.09 C 0.01 C Table 4 RTD types and ranges Externally mounted resistor modules 0.1% of input 0.2% of input Resolution Typical error (instrument at 20C) Maximum error (Instrument at 20C) Worst case temp Performance 1.4V 5.5V 37V 720V 0.035% I/P + 0.031% range 0.035% I/P + 0.028% range 0.035% I/P + 0.024% range 0.097% I/P + 0.027% range 0.085% I/P + 0.052% range 0.084% I/P + 0.039% range 0.084% I/P + 0.029% range 0.448% I/P + 0.033% range 80ppm of I/P per C 80ppm of I/P per C 80ppm of I/P per C 443ppm of I/P per C Table 1 Voltage ranges - accuracy and resolution Low Range High Range Resolution Typical error (instrument at 20C) Maximum error (Instrument at 20C) Worst case temp Performance 0 0 0 150 600 6k 5m 22m 148m 0.027% I/P +0.034% range 0.027% I/P +0.035% range 0.030% I/P +0.028% range 0.042% I/P + 0.110% range 0.042% I/P + 0.065% range 0.045% I/P + 0.035% range 35ppm of I/P per C 35ppm of I/P per C 35ppm of I/P per C Table 2 Resistance ranges - accuracy and resolution IEC 584.1 0 to +2300 Hoskins 0 to +2495 Hoskins -270 to +1000 IEC 584.1 0 to +2315 Hoskins -210 to +1200 IEC 584.1 -270 to +1372 IEC 584.1 -200 to +900 DIN43710:1985 (To IPTS68) -270 to +1300 IEC 584.1 -50 to +1768 IEC 584.1 -50 to +1768 IEC 584.1 -270 to +400 IEC 584.1 -200 to +600 DIN43710:1985 -50 to +1410 ASTM E1751-95 0 to +1406 Ipsen 0 to +1370 Engelhard 0 to +1888 ASTM E1751-95 Additional error due to shunt Additional error due to attenuator Performance 6100A/6180A See Table 1 Low Range Standard Analogue Output Board Three 25V nominal 20mA per output Installation category II; Pollution degree 2 Channel to channel: 100V RMS or DC (double insulation) Channel to ground: 100V RMS or dc (basic insulation) Fuse (20mm Type T) Supply voltage = 110/120V ac: 100mA Supply voltage = 220/240V ac: 63mA Relay Output Board General Maximum number of relay boards 6100A 4 (max no of relay outputs = 16) 6180A 9 (max no of relay outputs = 36) Number of relays per board 3 per C/O 4 per N/C 4 per N/O Estimated mechanical life 30,000,000 operations Update rate See `Update rates' in `Recorder Specification' above AC load ratings Derating The figures give below are for restive loads. for reactive or inductive loads, de-rate in accordance with Graph 1, in which F1 = Actually measured results on representative samples F2 = Typical values according to experience Contact life = Resistive contact life x reduction factor Maximum switching power 500VA Maximum contact voltage 250V providing this does not cause the maximum switching power (above) to be exceeded Maximum contact current 2 Amps providing this does not cause the maximum switching power (above) to be exceeded DC load ratings Maximum switching power Maximum contact voltage/ Current See Graph 2 for operating volt/amp envelope See Graph 2 for examples Max. DC load breaking capacity 300 200 resistive load DC Voltage (Volts) F2 F1 0.6 0.5 0.4 0.3 1 0.8 0.6 0.4 Power factor (cos ) 100 50 40 30 20 inductive load (L/R = 20msec) 10 0.1 0.2 0.5 1 2 5 DC current (Amps) 0.2 Graph 1 Derating curves for ac loads 20 Graph 2 DC load switching curves Event Input Number of inputs Maximum No. of boards 6 discrete inputs 6100A 4 6180A 4 Isolation Event input to ground: 100V RMS or dc (basic insulation) Event input to Event input: 0V Recognition levels `Active': -30V to +0.8V `Inactive': +2 to +30V Maximum frequency 8Hz Minimum pulse width 62.5ms Contact resistance Event: Active if resistance <35K Inactive if resistance >200K Status not defined if 35K < resistance <200K between input terminal and `C' terminal Current sink (voltage I/P) 10mA Safety Isolation Event input to ground Event input to Event input 100V RMS or dc (basic insulation) 0V Portable Portable option 6100A is available as a portable unit with either Thermocouple, General or HTM2010 connections. 6180A is available with optional carry handle and feet for portability Max. No Option of Inputs** Slots** 6100A Relays Serial Transm'r Comms PSU Event Analogue I/P O/P General 18 4 Yes* Yes Yes* Yes* Yes* Thermocouple 18 0 No Yes No Yes* No HTM2010 18 0 No Yes No No ** Mutually exclusive * Requires one option slot PORTABLE MECHANICAL INSTALLATION 430mm 190mm 384.8mm 40.5mm 153mm Transmitter PSU Isolated, 6100A recorder only Number of outputs Output voltage Maximum current Isolation (dc to 65Hz BS61010) F 1 0.9 0.8 0.7 240mm 15mm View on right hand side 300mm 38mm 174mm Installation category II; Pollution degree 2 300V RMS or dc (double insulation) 150V RMS or dc (basic insulation) Reduction Factor F 0 to 10V (source 5mA max.) 0 to 20mA (max. load 1K) 8Hz 250msec (10% to 90%) 0.024% of hardware range See table Performance in instrument at 20C Range Accuracy Temperature drift 0 to 10V 0.1% of range 0.12mV +0.022% of reading per C 0 to 20mA 0.1% of range 1A + 0.03% of reading per C Safety isolation Isolation (dc to 65 Hz; BS EN61010) Output channel-to OP channel Output channel to ground Installation category II; Pollution degree 2 Relay to Relay: 300V RMS or dc (double insulation) Relay to ground: 300V RMS or dc (basic insulation) Four Two 190mm General Max. number of output boards Number of outputs per board Output ranges Voltage: Current: Update rate Step response Linearity Performance Safety isolation Isolation (dc to 65Hz; BS EN61010) Top view MECHANICAL INSTALLATION INPUT BOARD WIRING 6100A 1 2 Standard terminal cover: 246.5 mm 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 Channels 1, 7, 13, 19, 25, 31, 37, 43 Channels 2, 8, 14, 20, 26, 32, 38, 44 Channels 3, 9, 15, 21, 27, 33, 39, 45 V+ V- I V+ V- I V+ V- I Channels Channels Channels 4, 10, 16, 22, 28, 5, 11, 17, 23, 29, 6, 12, 18, 24, 30, 34, 40, 46 35, 41, 47 36, 42, 48 Cold junction V+ V- I V+ V- I V+ V- I Long terminal cover: 288 (closed); 415mm (open) 24.75 mm 144 mm 211.5 mm Safety Earth Access Flap Max number of boards = 4 for small frame or 9 for large frame 137 mm 102 mm Side elevation (RHS) 35 mm Front view 106.9 mm 144 mm RELAY BOARD WIRING 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 2 Option Boards Relay 1 Relay 2 Relay 3 Relay 1 Relay 2 Relay 3 nc c no nc c no nc c no nc c no nc c no nc c no Three changeover relays Panel thickness = 3 to 25 mm. Optimum thickness depends on panel material 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 2 Option Boards x Rly 1 Rly 2 Rly 3 Rly 4 Rly 1 Rly 2 Rly 3 Rly 4 c nc c nc c nc c nc c nc c nc c nc c nc Panel cutout 138mm x 138mm (+1 - 0) Four normally closed relays 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 137 mm or 5.44 x 5.44 inches (+0.04 - 0.00) View on underside 2 Option Boards Rly 1 Rly 2 Rly 3 Rly 4 Rly 1 Rly 2 Rly 3 Rly 4 c no c no c no c no c no c no c no c no Minimum recommended inter-unit spacing Side clamps x = 15 mm (0.6 inch) y = 10 mm (0.4 in) Top/bottom clamps x = 10 mm (0.4 in) y = 15 mm (0.6 inch) Four normally open relays NC c y NO Contacts shown in power off/alarm state ISOLATED TRANSMITTER POWER SUPPLY WIRING Panel clamping (6100A only) Vertical Maximum installed angle a = b = 45 degrees max Long terminal cover Recorder rear terminal panel User terminal block Red Black - 6180A Red Black R Voltages generated by various currents for R = 100 and R =250 Red I V+ V- Red 250 0V 1.0 V 5.0V Black Transmitter Power Supply PCB Channel input 100 0 mA 0V 4 mA 0.4 V 20 mA 2.0 V Black a b - + Transmitter - + - + Only one channel shown, for clarity. + Other transmitters R (minimum spec) = 1.4W, 1%. 260.90 mm 292 mm 28.34 mm 211 mm Safety Earth EVENT INPUT BOARD WIRING 279 mm 292 mm Side elevation 216 mm 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 Front view 1 2 3 4 5 6 C 1 2 3 4 5 6 C Event input Number 1 2 3 4 5 Event input Number 6 C 1 2 3 4 5 6 C Behaviour undefined for +0.8 < Vin < +2 V Access Flap Input 6 shown; inputs 1 to 5 identical +2 to +30 V +0.8 to -30 V x Panel thickness range = 6 to 25 mm. Optimum thickness depends on panel material User 0V Panel cutout 281mm x 281mm (+1 - 0) Contact closure inputs Minimum recommended inter-unit spacing Side clamps x = 25 mm (1 inch) y = 12.5 mm (0.5 in) 279 mm or 11.07 x 11.07 inches (+0.04 - 0.00) ANALOGUE OUTPUT WIRING View on underside Top/bottom clamps x = 12.5 mm (0.5 in) y = 25 mm (1 inch) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 2 Option Boards y I+ MAXIMUM INSTALLED ANGLE a = b = 45 degrees max a b Panel mounting technique V+ - I+ Channel 2 V+ - I+ Channel 1 V+ - I+ Channel 2 I+ V+ - + V+ - Channel 1 Vertical Voltage inputs - Voltage outputs V INPUT BOARD SIGNAL WIRING V+ V- + I V+ V- - I V+ V- DC V (-20V to +20V) DC mV Thermocouples V+ V- I V+ V- I V+ V- I V+ V- I Attenuator assembly Shunt assembly + I - + - Minimum contact = 60ms DC V (-200 to +200V ) DC mA Shunt assembly 100 = LA246779UK10 250 = LA246779UK25 3-wire resistance thermometer 2-wire resistance thermometer Potentiometer Contact Closure (not Channels 1,7,13 etc.) Attenuator assembly LA244180 REAR TERMINAL CONNECTIONS Serial Comms Pin 1 2 3 4 5 6 7 8 9 EIA232 NC Rx Tx DTR Signal ground NC NC NC 5V via 1500 EIA485 (5-wire) EIA485 (3-wire) RxA Link to pin 7 NC NC NC NC NC NC Signal ground Signal ground RxB Link to pin 8 TxA TxA/RxA TxB TxB/RxB 5V via 1500 5V via 1500 usb1 usb2 1 2 3 4 5 1 2 3 4 5 6 7 8 9 6 7 8 9 Serial 1 Serial 2 Low supply voltage option 6180A +V or ac OPTION BOARD 1 2 OPTION BOARD 3 4 5 6 OPTION BOARDS 7 8 OPTION BOARDS 9 OPTION BOARDS 43 - 48 1 INPUT CHANNELS INPUT CHANNELS INPUT CHANNELS INPUT CHANNELS 22 37 - 42 1 31 - 36 22 1 1 6100A 1 13 - 18 1 1 OPTION BOARDS 1 INPUT CHANNELS 1 INPUT CHANNELS 1 2 22 3 4 22 7 - 12 22 1-6 22 1 22 7 - 12 1 1-6 1 19 - 24 22 0V or ac 6100A 22 25 - 30 Earth 22 22 INPUT CHANNELS Channels 13 - 18 (if fitted )occupy option board slots 3 and 4 22 OPTION BOARDS OPTION BOARDS INPUT CHANNELS INPUT CHANNELS ORDERING CODE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 0 Model 6100A 100mm TFT 1/4VGA Display . . . . . . . . . . . . . . . . . . . . . . . . . .6100A 6180A 180mm TFT XGA Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6180A 1 Number of channels 0 Input channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U00 6 Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U06 12 Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U12 18 Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U18 24 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U24 30 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U30 36 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U36 42 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U42 48 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U48 2 Not Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXX 3 Case Options Panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .PANEL General Portable (6100A only) . . . . . . . . . . . . . . . . . . . . . . . . . . .PORTGEN Type T Thermocouple Portable (6100A only) . . . . . . . . . . . . . . .PORTTTC Type K Thermocouple Portable (6100A only) . . . . . . . . . . . . . . .PORTKTC Type J Thermocouple Portable (6100A only) . . . . . . . . . . . . . . . .PORTJTC Voltage Input only Portable (6100A only) . . . . . . . . . . . . . . . . . .PORTVTC HTM2010/2030 Portable Test kit (6100A only) . . . . . . . . . . . . . .HTM2010 Carry Handle (Bezel Colour Silver, 6180A only) . . . . . . . . . . . . . . . . . . .CH Panel Mounting with Heavy Duty Case clamps (6180A only) . . . . . .PMHD Panel Mounting with Thin panel mounting kit (6180A only) . . . . . .PMTN 4 Lock Media lock not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOLCK Electronic Lock Fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .LOCK 5 Bezel Colour Silver including portable options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .SLV Eurotherm Green . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .GRN Black . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BLK 6 Power Supply 90- 264 Vac (110-370Vdc) 45 -65 Hz . . . . . . . . . . . . . . . . . . . . . . . . . . .VH 20 - 42 Vac (20 - 54Vdc) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .VL 240Vac 45 -65 Hz with low inrush current to meet NAMUR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .VH-NMR 24Vdc with low inrush current to meet NAMUR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .VL-NMR 7 24V Isolated Transmitter Power Supply Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NONE 110 - 120Vac 3channel TPS (6100A only) . . . . . . . . . . . . . . . . . . . .115TPS 220 - 240Vac 3 channel TPS (6100A only) . . . . . . . . . . . . . . . . . . . .230TPS 8 Non Standard Non Standard Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXXXX No logo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NLG 9 Internal Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32M Byte for history - approximately 4 million samples . . . . . . . . . .032M 96M Byte for history - approximately 12 million samples . . . . . . . . .096M 10 Removable Media Compact Flash and Front USB port . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CF Secure Digital and Front USB port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .SD 11 Memory Card Size Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMC 128M byte Card (CF or SD, dependant on removable media type) . .128M 256M byte Card (CF or SD, dependant on removable media type) . .256M 512M byte Card (CF or SD, dependant on removable media type) . .512M 1Gbyte Card (CF or SD, dependant on removable media type) . . . . .001G 12 USB Memory Stick Size Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMS 64M byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . . . . . .064MMS 256M byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . . . . .256MMS 512M byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . . . . .512MMS 1G byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .001GMS 13 Rear USB No rear USB ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0RUSB Two USB ports at rear of product . . . . . . . . . . . . . . . . . . . . . . . . . . .2RUSB 14 Serial Communication Ports Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0SRL Two EIA 232/422/485 Serial Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2SRL 15 Not Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXX 16 Calibration Certificates Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOCAL Calibration certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CAL 17 Changeover Relays Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00 3 changeover relays (1 option board) . . . . . . . . . . . . . . . . . . . . . . . . . . .03 6 changeover relays (2 option boards) . . . . . . . . . . . . . . . . . . . . . . . . . . .06 9 changeover relays (3 option boards) . . . . . . . . . . . . . . . . . . . . . . . . . . .09 12 changeover relays (4 option boards) . . . . . . . . . . . . . . . . . . . . . . . . . .12 15 changeover relays (5 option boards, 6180A only) . . . . . . . . . . . . . . .15 18 changeover relays (6 option boards, 6180A only) . . . . . . . . . . . . . . .18 21 changeover relays (7 option boards, 6180A only) . . . . . . . . . . . . . . .21 24 changeover relays (8 option boards, 6180A only) . . . . . . . . . . . . . . .24 27 changeover relays (9 option boards, 6180A only) . . . . . . . . . . . . . . .27 18 Normally Closed Relays Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00 4 Normally Closed relays (1 option board) . . . . . . . . . . . . . . . . . . . . . . .04 8 Normally Closed relays (2 option boards) . . . . . . . . . . . . . . . . . . . . . . .08 12 Normally Closed relays (3 option boards) . . . . . . . . . . . . . . . . . . . . . .12 16 Normally Closed relays (4 option boards) . . . . . . . . . . . . . . . . . . . . . .16 20 Normally Closed relays (5 option boards, 6180A only) . . . . . . . . . . .20 24 Normally Closed relays (6 option boards, 6180A only) . . . . . . . . . . .24 28 Normally Closed relays (7 option boards, 6180A only) . . . . . . . . . . .28 32 Normally Closed relays (8 option boards, 6180A only) . . . . . . . . . . .32 36 Normally Closed relays (9 option boards, 6180A only) . . . . . . . . . . .36 19 Normally Open Relays Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00 4 Normally Open relays (1 option board) . . . . . . . . . . . . . . . . . . . . . . . .04 8 Normally Open relays (2 option boards) . . . . . . . . . . . . . . . . . . . . . . . .08 12 Normally Open relays (3 option boards) . . . . . . . . . . . . . . . . . . . . . . .12 16 Normally Open relays (4 option boards) . . . . . . . . . . . . . . . . . . . . . . .16 20 Normally Open relays (5 option boards, 6180A only) . . . . . . . . . . . .20 24 Normally Open relays (6 option boards, 6180A only) . . . . . . . . . . . .24 28 Normally Open relays (7 option boards, 6180A only) . . . . . . . . . . . .28 32 Normally Open relays (8 option boards, 6180A only) . . . . . . . . . . . .32 36 Normally Open relays (9 option boards, 6180A only) . . . . . . . . . . . .36 20 Event Inputs Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00 06 Event Inputs (1 board) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .06 12 Event Inputs (2 boards) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 18 Event Inputs (3 boards) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 24 Event Inputs (4 boards) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24 21 Analogue Outputs None . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00 2 Analogue Outputs (1 option board) . . . . . . . . . . . . . . . . . . . . . . . . . . .02 4 Analogue Outputs (2 option boards) . . . . . . . . . . . . . . . . . . . . . . . . . . .04 6 Analogue Outputs (3 option boards) . . . . . . . . . . . . . . . . . . . . . . . . . . .06 8 Analogue Outputs (4 option boards) . . . . . . . . . . . . . . . . . . . . . . . . . . .08 22 Quantity of Shunts Enter quantity required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ._ _ 23 Shunt Value Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOS 100 ohm shunts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100 250 ohm shunts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250 24 Quantity of Attenuators (100:1) Enter quantity required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ._ _ 25 Warranty Standard warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXXX Extended warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .WL005 26 Bridge Bridge Lite (supplied as standard) . . . . . . . . . . . . . . . . . . . . . . . . . . . .BLITE Bridge Full . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BFULL 27 Review and Quickchart Review and Quickchart Lite (supplied as standard) . . . . . . . . . . . . . .RLITE Review and Quickchart Full . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .RFULL 28 Auditor Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOADT Audit Trail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ALITE Auditor Full . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AFULL 29 Security Manager Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOSM Security Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .SECMAN 30 Groups 6 Groups (supplied as standard) . . . . . . . . . . . . . . . . . . . . . . . . .06GROUP 12 Groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12GROUP 31 Maths, Totalisers and Counters Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC00 36 Virtual Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC36 96 Virtual Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC96 128 Virtual Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC128 32 Batch Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOBTCH Batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BATCH 33 Screen Builder Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOSB Advanced Screen Builder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ADSB 34 Master Communications Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMSTR Modbus Master Communications . . . . . . . . . . . . . . . . . . . . . . . . .MODBUS Eurotherm: International sales and service Understanding and providing local support is a key part of Eurotherm's business. Complementing worldwide Eurotherm offices are a whole range of partners and a comprehensive technical support team... a soothing melody to ensure you get a service you will want to go back to. AUSTRALIA Sydney Eurotherm Pty. Ltd. Telephone (+61 2) 9838 0099 Fax (+61 2) 9838 9288 E-mail info@eurotherm.com.au FRANCE Lyon Eurotherm Automation SA Telephone (+33 478) 664500 Fax (+33 478) 352490 E-mail ea@automation.eurotherm.co.uk IRELAND Dublin Eurotherm Ireland Limited Telephone (+353 1) 469180 Fax (+353 01) 4691300 E-mail info@eurotherm.ie SWEDEN Malmo Eurotherm AB Telephone (+46 40) 384500 Fax (+46 40) 384545 E-mail info@eurotherm.se AUSTRIA Vienna Eurotherm GmbH Telephone (+43 1) 7987601 Fax (+43 1) 7987605 E-mail eurotherm@eurotherm.at GERMANY Limburg Eurotherm Deutschland GmbH Telephone (+49 6431) 2980 Fax (+49 6431) 298119 E-mail info@regler.eurotherm.co.uk SWITZERLAND Freienbach Eurotherm Produkte (Schweiz) AG Telephone (+41 55) 4154400 Fax (+41 55) 4154415 E-mail epsag@eurotherm.ch BELGIUM & LUXEMBURG Huy Eurotherm S.A/N.V. Telephone (+32) 85 274080 Fax (+32 ) 85 274081 E-mail sales@eurotherm-belgium.be HONG KONG & CHINA Eurotherm Limited Aberdeen Telephone (+85 2) 28733826 Fax (+85 2) 28700148 E-mail eurotherm@eurotherm.com.hk ITALY Como Eurotherm S.r.l Telephone (+39 31) 975111 Fax (+39 31) 977512 Telex 380893 EUROTH I E-mail info@eurotherm.it BRAZIL Campinas-SP Eurotherm Ltda. Telephone (+5519) 3707 5333 Fax (+5519) 3707 5345 E-mail eurothermltda@eurothermltda.com.br Guangzhou Office Telephone (+86 20) 8755 5936 Fax (+86 20) 8755 5831 DENMARK Copenhagen Eurotherm Danmark A/S Telephone (+45 70) 234670 Fax (+45 70) 234660 E-mail info@eurotherm.se FINLAND Abo Eurotherm Finland Telephone (+358) 22506030 Fax (+358) 22503201 Beijing Office Telephone (+86 10) 6762 0936 Fax (+86 10) 6762 0931 Shanghai Office Telephone (+86 21) 6352 6406 Fax (+86 21) 6352 7351 INDIA Chennai Eurotherm India Limited Telephone (+91 44) 24961129 Fax (+91 44) 24961831 E-mail sales@eurothermdel.com KOREA Seoul Eurotherm Korea Limited Telephone (+82 31) 2738507 Fax (+82 31) 2738508 E-mail help@eurotherm.co.kr NETHERLANDS Alphen a/d Ryn Eurotherm B.V. Telephone (+31 172) 411752 Fax (+31 172) 417260 E-mail sales@eurotherm.nl NORWAY Oslo Eurotherm A/S Telephone Oslo (+47 67) 592170 Fax (+47 67) 118301 E-mail info@eurotherm.se UNITED KINGDOM Worthing Eurotherm Limited Telephone (+44 1903) 268500 Fax (+44 1903) 265982 E-mail info@eurotherm.co.uk Web www.eurotherm.co.uk U.S.A Leesburg VA Eurotherm Inc. Telephone (+1 703) 443 0000 Fax (+1 703) 669 1300 E-mail info@eurotherm.com Web www.eurotherm.com ED43 SPAIN Madrid Eurotherm Espana SA Telephone (+34 91) 6616001 Fax (+34 91) 6619093 E-mail ventas@iberica.eurotherm.co.uk (c) Copyright Eurotherm Limited 2005 Invensys, Eurotherm, the Eurotherm logo and Wonderware are trademarks of Invensys plc, its subsidiaries and affiliates. All other brands may be trademarks of their respective owners. All rights are strictly reserved. No part of this document may be reproduced, modified, or transmitted in any form by any means, nor may it be stored in a retrieval system other than for the purpose to act as an aid in operating the equipment to which the document relates, without the prior written permission of Eurotherm limited. Eurotherm Limited pursues a policy of continuous development and product improvement. The specifications in this document may therefore be changed without notice. The information in this document is given in good faith, but is intended for guidance only. Eurotherm Limited will accept no responsibility for any losses arising from errors in this document. Part No. HA029073 Issue 2 6000 Specification sheet Printed in England 12.05