Hawk 3 Digital Panel Meter /
Controller
All parameters set from easy to understand front panel access
• One, two or four 5-amp relays optional
• Five user-selectable brightness levels
• 1/8 DIN, shallow depth case, 3.24”
• RS485 digital communications optional (H345)
• 12 or 24 DCV Excitation output optional, maximum 30mA @ 12V and 20mA @ 24V
• 4-20mA or 0-10 DCV analog retransmission optional
• NEMA 4X rated front panel
File # E199939
1
Rev. 04-15
Hawk 3 Digital Panel Meter /
Controller
Specications
DISPLAY
Type 7- segment, red LED, 4 or 5 digits
Height
0.56” (14.2mm)
Brightness 5 settings, user programmable
Decimal Point 4 or 5 position, user programmable
Overrange Indication Display ashes “EEEE” indicating Maximum
Value Exceeded (Example: H335)
Underrange Indication Display ashes “-EEE” indicating Minimum
Value Exceeded (Example: H335)
Alarm Indicators 4 LED indicators for up to four independent
setpoints
Linearization H335 2 points
H345 16 points
POWER REQUIREMENTS
AC 85 to 250 VAC or 120VAC @ 10VA
DC 9 to 36 DCV @ 10VA
Isolation 250V RMS MAX
Note: Each supply is shown at the maximum and minimum values
except the 120 VAC unit, which is allowed ± 10%
ACCURACY @ 25°C as % of rdg
4-1/2 digit
3-1/2 digit
DC Current
High (5A, 2A) ± 0.2% of reading
± 1 count
±
0.3% of reading
± 1 count
All others ± 0.05% of reading
± 1 count
±
0.1% of reading
± 1 count
DC Volts
High (600V) ± 0.1% of reading
± 1 count
±
0.2% of reading
± 1 count
All others ± 0.05% of reading
± 1 count
±
0.1% of reading
± 1 count
Resistance
All ranges ± 0.1% of reading
± 2 counts
±
0.1% of reading
± 2 counts
* AC Current
High (2A, 5A) ± 0.2% of reading
± 2 counts
±
0.3% of reading
± 2 counts
All others ± 0.1% of reading
± 2 counts
±
0.2% of reading
± 2 counts
* AC Volts
High (600V) ± 0.1% of reading
± 1 count
±
0.2% of reading
± 1 count
All others ± 0.05% of reading
± 1 count
±
0.1% of reading
± 1 count
* AC functions measured at 50 hz, include ± 1 count for each
additional 100 Hz above 50 Hz
Specications - Con’t
ENVIRONMENTAL
Operating Temperature 0 to 50ºC
Storage Temperature -10 to +60°C
Relative Humidity < 80% for the temp. up to 31ºC and decreasing
linearly to 50% relative humidity at 50°C
Ambient Temp 25°C
Temperature Drift ± 100 ppm /°C
Warmup time 10 minutes
NOISE REJECTION
NMRR 60 dB @ 50-60 Hz
CMRR 70 db @ 50-60 Hz
For indoor use to an altitude up to 200m
A TO D CONVERSION
Technique Successive approximation with oversampling
Sample Rate 10 conversions per second
Display Rate User Programmable from 1 - 420 updates /
minute (240 default)
RS-485 Specications (only available on H345)
2 wire / Half duplex, Baud rate: 9600 baud, 1ms delay per character,
32 Nodes Maximum on Bus.
Optically and magnetically isolated for ground loop elimination
MECHANICAL
Bezel 3.92” x 2.0” x 0.52” (99.8mm x 51.9mm x
132mm)
Depth 3.24” (82.3mm) behind panel
Panel cutout 3.62” x 1.77” (92mm x 45mm) 1/8 DIN
Weight 10 oz. (283.5 g)
Cover NEMA 4X Rated front panel
ELECTRICAL
Accuracy Listed as % of reading at 25°C.
Add 100ppm/°C to compensate for drift.
Tested at 50Hz, include +/-1 count for every
100Hz above 50 Hz
Transient Overvoltage Installation Category III, Pollution Degree 2
Analog Output Sampling Rate = 100 mSec.
Reaction Time 0 to Full Scale = 10 μSec
2
Rev. 04-15
_____
IIIIII
Basic Unit Power
Supply
Function/
Range
Output
Signal
5A Relay Excitation
Ordering Information -
Select From Each One Below
Basic Unit Function/Range Output Signal
H335 3-1/2 digit, Red LED 41 200 ACμA 0 None
H345 4-1/2 digit, Red LED 42 2 ACmA 1 4-20 DCmA
43 20 ACmA 2 0-10 DCV
Power Supply 44 200 ACmA 6 RS-485 (4-1/2 only)
1 120 ACV (3-1/2 only) 45 2 ACA
2 85-250 ACV (4-1/2 only) 46 5 ACA 5A Relays
3 9-36 DCV 0 None
4 85-250 ACV (3-1/2 only) 51 200 ACmV TRMS 1 One
52 2 ACV TRMS 2 Two
Function/Range 53 20 ACV TRMS 4 Four
11 200 DCmv 54 200 ACV TRMS
12 2 DCV 55 600 ACV* TRMS Excitation
13 20 DCV 0 None
14 200 DCV 61 200 ACμA TRMS 1 12 DCV
15 600 DCV * 62 2 ACmA TRMS 2 24 DCV
63 20 ACmA TRMS
21 200 DCμA 64 200 ACmA TRMS
22 2 DCmA 65 2 ACA TRMS
23 20 DCmA 66 5 ACA TRMS
24 200 DCmA
25 2 DCA 71 4-20 DCmA Process
26 5 DCA 72 0-10 DCV Process
31 200 ACmV 81 200 Ohm
32 2 ACV 82 2K Ohm
33 20 ACV 83 20K Ohm
34 200 ACV 84 200K Ohm
35 600 ACV *
* Awaiting UL approval
Hawk 3 Digital Panel Meter /
Controller
Hawk 3 Indicators can be congured by making an entry into each section. Example: H335-3-71-0-4-1
3
Rev. 04-15
Hawk 3 Digital Panel Meter /
Controller
Installation and Panel Cutout - H335, H340, H345
Mounting Requirements
The Hawk 3 Advanced Digital Controller 1/8 DIN meters
require a panel cutout of 1.77” (45mm) high by 3.62”
(92mm) wide.
To install the Hawk 3 meter into a panel cutout, remove
the clips from the side of the meter.
Slide the meter through the panel cutout, then slide
the mounting clips back on the meter. Press evenly to
ensure a proper t. Tighten screws.
Engineering Label Placement
To replace the engineering unit label, place the tip of
a ballpoint pen into the small hole at the base of the
engineering label in the bezel.
Slide the label up until it pops out. Grasp and remove.
Slide the new label half the distance in, then use the
ballpoint pen to slide it down into place.
Engineering Label
4
Rev. 04-15
Hawk 3 Digital Panel Meter /
Controller
Inputs
DC Voltage
Range
Resolution
4-1/2
Resolution
3-1/2
Input
Impedance
Overload
200 mV 10 μV .1 mV 1M Ω 5 DCV
2 V .1 mV 1 mV 1M Ω 5 DCV
20 V 1 mV 10 mV 1M Ω 300 DCV
200 V 10 mV .1 V 1M Ω 300 DCV
600 V .1 V 1 V 1M Ω 1K DCV
DC Current
Range
Resolution
4-1/2
Resolution
3-1/2
Input
Impedance
Overload
200 μA 10 nA .1 μA 1K Ω 4.5 mA DC
2 mA .1 μA 1 μA 100 Ω 45 mA DC
20 mA 1 μA 10 μA 10 Ω 200 mA DC
200 mA 10 μA .1 mA 1 Ω 600 mA DC
2 A .1 mA 1 mA .013 Ω 5.5 A DC
5 A 1 mA 10 mA .013 Ω 5.5 A DC
AC Current (same for TRMS @ 60 Hz)
Range
Resolution
4-1/2
Resolution
3-1/2
Input
Impedance
Overload
200 μA 10 nA .1 μA 1K Ω 4.5 mA DC
2 mA .1 μA 1 μA 100 Ω 45 mA DC
20 mA 1 μA 10 μA 10 Ω 200 mA DC
200 mA 10 μA .1 mA 1 Ω 600 mA DC
2 A .1 mA 1 mA .013 Ω 5.5 A DC
5 A 1 mA 10 mA .013 Ω 5.5 A DC
AC Voltage (same for TRMS @ 60 Hz)
Range
Resolution
4-1/2
Resolution
3-1/2
Input
Impedance
Overload
200 mV 10 μV .1 mV 200K Ω 5 DCV
2 V .1 mV 1 mV 200K Ω 5 DCV
20 V 1 mV 10 mV 1M Ω 300 DCV
200 V 10 mV .1 V 1M Ω 300 DCV
600 V .1 V 1 V 1M Ω 1K DCV
Resistance
Range
Resolution
4-1/2
Resolution
3-1/2
Input
Impedance
Overload
200 Ω 10 m Ω .1 Ω 1.2K Ω ± 5 DCV
2K Ω .1 Ω 1 Ω 12K Ω ± 5 DCV
20K Ω 1 Ω 10 Ω 121K Ω ± 5 DCV
200K Ω 10 Ω 100 Ω 1.2M Ω ± 5 DCV
5
Rev. 04-15
Mouser Electronics
Authorized Distributor
Click to View Pricing, Inventory, Delivery & Lifecycle Information:
Simpson:
H335-3-14-0-2-0 H335111020 H335134020 H335113020 H335146020 H335-1-71-012 H335171020 H335371012
H335-3-13-1-0-0 H335-1-15-0-0-0 H335415020 H335-1-35-0-0-0 H335481020 H335-1-81-0-2-0 H335-1-51-0-4-0
H335-3-11-0-0-0 H335-3-13-0-0-0 H335171000 H335181000 H335182000 H335435000 H345254200 H335146040
H335135020 H345-1-14-2-4-2 H345225000 H335171100 H345214601 H345213601 H345212601 H345324210
H335325100 H335-4-53-0-0-0 H335166100