9190A Ultra-Cool
Field Metrology Well
Ultra-cool dry-block calibrator
with best-in-class stability Technical Data
The Fluke Calibration 9190A Ultra-Cool Field Metrology
Well is the most accurate and stable, cold temperature
dry-block on the market. It’s ideal for applications that
pli
and calibration of RTDs, thermocouples, thermometers,
and sensors used with process control equipment such
as medical freezers, laboratory refrigerators, cold rooms,
blood banks, sterilizers (autoclaves), and freeze dryers.
demand strict quality control and regulatory process com-
ance. These applications include on-location validation
Great for cleanroom environments
Calibration baths are the most stable and uniform temperature
sources available, but they aren’t a good fit for cleanrooms. The
size of a bath limits its portability; it uses fluids that produce fumes
that can easily be spilled. The 9190A Ultra-Cool Field Metrology
Well is a great alternative. Its wide temperature range brackets the
coldest and highest temperature ranges required for pharmaceuti-
cal, biomedical and food processing applications. The 9190A is
small and lightweight, making it easy to transport. There are no
fluids so cleanrooms stay clean. And the 9190A cooling and heat-
ing times are faster than a calibration bath—that means calibration
work gets done more quickly.
An accurate temperature source is critical for
dependable process measurements
Unreliable process measurements can have a damaging impact on
business, leading to poor product quality, recalls, fines, waste, and
lost profits. Ultimately, measurements are only as good as the tem-
perature sources used to calibrate the measurement equipment.
The 9190A Ultra-Cool Field Metrology Well incorporates the
best technology and design expertise gained from decades of dry-
block development experience. The 9190A conforms to EURAMET
cg-13, guidelines for best measurement practices for temperature
block calibrators. As a result, the 9190A specifications for accuracy,
stability, axial (vertical) uniformity, radial (well-to-well) uniformity,
loading, and hysteresis have been thoroughly and carefully defined
and tested. With a 9190A Ultra-Cool Field Metrology Well, you can
be confident you’re using the most accurate and stable ultra-cool
dry-block calibrator available. And that will have a positive impact
on your business.
accuracy and stability uncertainty
process
control
Increased
measurement
reliability
Business
results
Calibration equipment Lower
calibration
Improved
Accurate and stable calibr ation equipment impacts business results (e.g. quality , efficiency, less waste)
2 Fluke Calibration 9190A Ultra-Cool Field Metrology Well
Features at a glance
Wide temperature range
From –95 °C to 140 °C
Excellent accuracy
Accuracy using built-in refer-
ence thermometer readout:
± 0.05 °C full range
Display accuracy: ± 0.2 °C full
range
Best-in-class stability
± 0.015 °C full range
Fast cooling time
23 °C to –90 °C: 80 minutes
23 °C to –95 °C: 90 minutes
140 °C to 23 °C: 60 minutes
Portability
Weighs only 16 kg (35 lb)
Built-in front and back handles
for easy two-handed carry
Best measurement
practices
Conforms to EURAMET cg-13
guidance on measurement
practices for temperature block
calibrators
Ergonomics
Two-handed carry; built-in
handles on front and back
All controls and inputs on the
front panel
Stability indicator
Shows whether block tem-
perature is stable and within
limits defined by the user
PC interface
USB and RS-232 support
remote operation from a
computer
Dual heater zone
Active heater zone control
compensates for differential
temperatures between zones
Minimizes axial temperature
gradients
Ramp-and-soak set-points
Define up to eight set-point
temperatures; automatically
ramp to and “soak” at each
set-point
Specify soak time, number of
cycles, and direction
Fast time to temperature and
small footprint
Compact, efficient free piston
Stirling cooler technology
(FPSC)
Optional process
features
Reference sensor control
Set-point temperature control
gets passed from the
internal sensor to external
reference PRT
9190A controls the well tem-
perature based on its
measurement and position
inside the well
Helps minimize the effects of
the axial gradient when
reference PRT is aligned with
short sensors
4-20 mA connectors
Power a 4-20 mA transmitter
from the drywell
Supplies 24 V dc loop power
4-wire PRT/RTD input
Reads a 4-wire, 3-wire, or
2-wire PRT/RTD
Accuracy ± 0.02 °C
4-20 mA circuit fuses
Thermocouple input
Accepts common thermo-
couples with mini-jack
termination
Types: J, K, T, E, R, S, M, L, U,
N, C
Reference thermometer input
Smart connector input accepts
–A terminated reference PRTs
Accuracy ± 0.01 °C at –95 °C
Specifications
Base unit specifications
Temperature range at 23 °C –95 °C to 140 °C (–139 °F to 284 °F)
Display accuracy ± 0.2 °C full range
Accuracy with external reference [3] ± 0.05 °C full range
Stability ± 0.015 °C full range
Axial uniformity at 40 mm (1.6 in) ± 0.05 °C full range
Radial gradient ± 0.01 °C full range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
± 0.006 °C full range
(versus display with one 6.35 mm probe)
± 0.25 °C at –95 °C
± 0.10 °C at 140 °C
Operating conditions 0 °C to 35 °C, 0 % to 90 %
RH (non-condensing) < 2000 m altitude
Environmental conditions for all 13 °C to 33 °C
specifications except temperature range
Immersion (well) depth 160 mm (6.3 in)
Well diameter 30 mm (1.18 in)
Heating time [1] –95 °C to 140 °C: 40 min
Cooling time [1] 23 °C to –90 °C: 80 min
23 °C to –95 °C: 90 min
140 °C to 23 °C: 60 min
Stabilization time [2] 15 min
Resolution 0.01
Display LCD, °C or °F user selectable
Size (H x W x D) 480 mm x 205 mm x 380 mm (18.8 in x 8.0 in x 14.9 in)
Weight 16 kg (35 lb)
Power requirements 100 V to 115 V (± 10 %) 50/60 Hz, 575 W
200 V to 230 V (± 10 %) 50/60 Hz, 575 W
System fuse ratings 115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V
4–20 mA Fuse (-P model only) 50 mA F 250 V
Computer interface RS-232, USB Serial, and 9930 Interface-It temperature
calibration software included
Safety IEC 61010-1, Installation Category II, Pollution degree 2
Electromagnetic environment IEC 61326-1: Basic
Refrigerants R32 (Difluoromethane)
< 20 g, ASHRAE safety group A2L
R704 (Helium)
< 20 g, ASHRAE safety group A1
3 Fluke Calibration 9190A Ultra-Cool Field Metrology Well
-P Specifications
Built-in reference thermometer readout
accuracy (4-wire reference probe) [3]
± 0.010 °C at –95 °C
± 0.013 °C at –25 °C
± 0.015 °C at 0 °C
± 0.020 °C at 50 °C
± 0.025 °C at 140 °C
Reference resistance range 0  to 400 
Reference resistance accuracy [4] 0  to 42 : ± 0.0025 
42  to 400 : ± 60 ppm of reading
Reference characterizations ITS-90, CVD, IEC-751, resistance
Reference measurement capability 4 wire
Reference probe connection 6-pin din with INFO-CON technology
Built-in RTD thermometer readout accuracy NI-120: ± 0.015 °C at 0 °C
PT-100 (385): ± 0.02 °C at 0 °C
PT-100 (3926): ± 0.02 °C at 0 °C
PT-100 (JIS): ± 0.02 °C at 0 °C
RTD resistance range 0  to 400 
Resistance accuracy [4] 0  to 25 : ± 0.002 
25  to 400 : ± 80 ppm of reading
RTD characterizations PT-100 (385),(JIS),(3926), NI-120, resistance
RTD measurement capability 2-wire, 3-wire, and 4-wire RTD with jumpers only
RTD connection 4-terminal input
Built-in TC thermometer readout accuracy [5] Type J: ± 0.70 °C at 140 °C
Type K: ± 0.75 °C at 140 °C
Type T: ± 0.60 °C at 140 °C
Type E: ± 0.60 °C at 140 °C
Type R: ± 1.60 °C at 140 °C
Type S: ± 1.60 °C at 140 °C
Type M: ± 0.65 °C at 140 °C
Type L: ± 0.65 °C at 140 °C
Type U: ± 0.70 °C at 140 °C
Type N: ± 0.75 °C at 140 °C
Type C: ± 1.00 °C at 140 °C
TC millivolt range 10 mV to 75 mV
Voltage accuracy 0.025 % of reading +0.01 mV
Internal cold junction compensation accuracy ± 0.35 °C (ambient of 13 °C to 33 °C)
TC connection Miniature connectors (ASTM E1684)
Built-in mA readout accuracy 0.02 % of reading + 0.002 mA
mA range Cal 4-22 mA, Spec 4-24 mA
mA connection 2 terminal input
Loop power function 24 V dc loop power
Built-in electronics temperature coefficient ± 0.005 % of range per °C
(0 °C to 13 °C, 33 °C to 50 °C)
Notes:
[1] For ambient temperature of 23 °C.
[2] Time from when the SETPOINT is reached to when the unit is with in stability specification.
[3] The temperature range may be limited by the reference probe connected to the readout.
The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe
characterization errors.
[4] Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05
 to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
[5] The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.
4 Fluke Calibration 9190A Ultra-Cool Field Metrology Well
5 Fluke Calibration 9190A Ultra-Cool Field Metrology Well
9190A Ultra-Cool Field Metrology Well
Model Description
9190A-X Ultra-Cool Field Metrology Well, –95 °C to 140 °C, with
9190-INSX
9190A-X-P Ultra-Cool Field Metrology Well, –95 °C to 140 °C, with
9190-INSX, with Process Electronics
‘X’ in the above model numbers to be replaced with A, B, C, D, E, and F as
appropriate for the desired insert. See the inserts illustration and listing
below.
Recommended accessories
Model Description
9190-INSA Insert “A” 9190, imperial miscellaneous holes
9190-INSB Insert “B” 9190, imperial comparison holes
9190-INSC Insert “C” 9190, 0.25 inch holes
9190-INSD Insert “D” 9190, metric comparison holes
9190-INSE Insert “E” 9190, metric miscellaneous holes with 0.25
inch hole
9190-INSF Insert “F” 9190, metric comparison miscellaneous holes
with 0.25 inch hole
9190-INSY Insert “Y” 9190, custom insert
Custom insert is based on a maximum of eight holes.
Please contact your local sales representative if you have special
requirements.
9190-INSZ Insert “Z” 9190, blank
Ordering Information
Fluke Calibration
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2013 Fluke Calibration.
Specifications subject to change without notice.
Printed in U.S.A. 1/2013 4264972A D-EN-N
Pub-ID 12027-eng
Modification of this document is not permitted
without written permission from Fluke Calibration.
Fluke Calibration.
Precision, performance, confidence.
9190A-DW-156 - Ultra-Cool Field Metrology Well, –95 °C to
140 °C, with 9190-INSX and no inserts.
9190A-DW-P-156 - Ultra-Cool Field Metrology Well, –95 °C
to 140 °C, with 9190-INSX, Process Electronics and no
inserts.