Digital Multimeters
Tektronix DMM4050 and DMM4040 Data Sheet
DMM4050
Features & Benets
Key Performance Specications
6.5 Digit Resolution
Basic VDC Accuracy of up to 0.0024% (1 yr.)
100 mV to 1000 V Voltage Range, with up to 100 nV Resolution
100 A to 10 A Current Range, with up to 100 pA Resolution
10 to 1 GOhm Range, with up to 10  Resolution
CAT I 1000 V, CAT II 600 V
Available Functions and Features
Volts, Ohms, and Amps Measurements
Diode and Continuity Testing
Frequency and Period Measurements
Temperature and Capacitance Measurements (DMM4050)
2×4 Ohms 4-wire Measurement Technique
TrendPlot™ Paperless Data Recorder Mode
Measurement Statistics
Histogram Mode
Connectivity
Front and Rear 2×4 Measurement Inputs
USB Host Port on Front Panel for Easy Storage of Measurement Data
and Instrument Settings
RS-232, LAN, and GPIB Ports on Rear Panel for Quick PC Connectivity
Includes USB to RS-232 Interface Adapter Cable
Includes National Instrument’s LabVIEW SignalExpress™ TE Limited
Edition for Connecting Your Bench
3-year Warranty
Data Sheet
TrendPlot display.
Feature-Rich Tools for Precision
Measurements
As the circuits in embedded system designs become more sophisticated
with tighter tolerances, you must measure a multitude of different
parameters with a high degree of accuracy to validate your design. The
Tektronix DMM4050 and DMM4040 6.5 digit bench multimeters pack many
different functions and analysis into one instrument, all with exceptional
precision and performance.
Typical multimeter measurements volts, ohms, and amps are made with
a basic VDC accuracy of up to 0.0024%, and resolution of 100 pA and 10
,ensuri
ng you have the performance you need for today’s demanding
designs. You can also use the DMM4050/4040 to measure frequency and
period, and to perform continuity and diode tests. For additional exibility,
the DMM4050 offers temperature and capacitance measurements. This
allows you to replace your temperature meter, capacitance meter, counter,
continuity tester, and traditional DMM with one versatile instrument, saving
bench space and cost.
Analyze Your Device with Graphical Display
Modes
With the unique dual display of the DMM4050/4040, you can measure
two different parameters of the same signal from one test connection. To
reveal signal quality issues like drift, intermittent transients, and stability,
you can view the data as a real-time trend plot or a histogram with the
DMM4050/4040’s graphical display mode, or you can use measurement
statistics to track how signal parameters are changing over time.
TrendPlot™ Paperless Recorder Mode
Depending on your test case, your signal parameters may change
from instant to instant. By taking multiple measurements over minutes,
hours, or days, you can quantify those changes. With TrendPlot™,
you can graphically plot the trend of a measured value over time, from
short time spans to extended periods of time. TrendPlot can be used
with measurements like DC voltage, DC current, frequency, resistance,
and temperature. AC voltage and current can be plotted as RMS
measurements.
A sample of a Min/Max/Avg/SD Statistics Report.
Histogram display.
Measurement Statistics
With integrated statistics processing, you can calculate both the average
and standard deviation of a measurement, as well as track the minimum
and maximum measured values, with the push of a button. Statistics can
be performed on DC voltage, AC voltage, AC voltage in dB, DC current,
AC current, resistance, capacitance, frequency, period, and temperature
measurements.
Histogram Plots
To graphically see the average and standard deviation of a set of
measurements, you can use the histogram function to see the distribution of
measurement results.
Designed to Make Your Work Easier
The DMM4050/4040 multimeter is designed with the ease-of-use and
familiar operation you expect from Tektronix.
Intuitive Operation
Dedicated front-panel buttons provide fast access to frequently used
functions and parameters, reducing setup time. You no longer need to
search through software menus to nd the function you need.
Easy Data Storage and PC Connectivity
A USB port on the front panel makes it easy to store measurement data
and instrument setups to a memory stick. Or connect to your PC with the
LAN, RS-232, or GPIB port on the back panel. A USB to RS-232 interface
adapter cable is included standard with the DMM4050/4040 to make it easy
to connect to your PC’s USB port.
Simple and Accurate 4-wire Measurements
Patented split terminal jacks for the 2×4 ohms function allow you to perform
4-wire measurements using only two leads instead of four. Special test lead
accessories are available to enable you to establish the connection. You get
excellent resolution and accuracy plus the convenience and ease of using a
single pair of leads.
2www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
Signal Express acquiring data from Tektronix DMM4050 and DPO3052.
Connect Your Bench for Intelligent Debug
Easily capture, save, and analyze measurement results from your
multimeter with the special Tektronix Edition of National Instrument’s
LabVIEW SignalExpress™ software. Every DMM4050 and DMM4040 ships
with a free copy of the Limited Edition version of SignalExpress for basic
instrument control, data logging, and analysis. The optional Professional
Edition offers over 200 built-in functions that provide additional signal
processing, advanced analysis, sweeping, limit testing, and user-dened
step capabilities.
SignalExpress supports the range of Tektronix bench instruments*1,
enabling you to connect your entire test bench. You can then access the
feature-rich tools packed into each instrument from one intuitive software
Signal Express used to export DMM4050 data into Excel.
interface. This allows you to automate complex measurements requiring
multiple instruments, log data for an extended period of time, time-correlate
data from multiple instruments, and easily capture and analyze your results,
all from your PC. Only Tektronix offers a connected test bench of intelligent
instruments to simplify and speed debug of your complex design.
Performance You Can Count On
In addition to industry-leading service and support, every DMM4050 and
DMM4040 multimeter comes backed with a three-year standard warranty.
*1NI LabVIEW Signal Express supports the following Tektronix instruments: MSO/DPO4000/3000/2000
Series oscilloscopes, TDS3000C/2000B/1000B Series oscilloscopes, AFG3000 Series arbitrary/function
generators, DMM4050/4040/4020 Series digital multimeters.
www.tektronix.com/dmm4050_4040 3
Data Sheet
Characteristics
General Specications
Voltage
Characteristic Description
100 V Setting 90 V to 110 V
120 V Setting 108 V to 132 V
220 V Setting 198 V to 242 V
240 V Setting 216Vto264V
Frequency 47 Hz to 440 Hz. Automatically sensed at power-on
Power Consumption 28 VA peak (12 Watt average)
Dimensions
Dimension mm in.
Height 88 3.46
Width 217 8.56
Depth 297 11.7
Weight kg lb.
Net 3.6 8.0
Shipping 5.0 11.0
Display
Vacuum Fluorescent Display, dot matrix
Environment
Characteristic Description
Temperature
Operating 0 °C to 55 °C
Storage -40 °C to 70 °C
Warm Up 1 hour to full uncertainty specications
Relative Humidity (noncondensing)
Operating <90% (0 °C to 28 °C)
<80%(2Cto4C)
<50%(4Cto5C)
Storage <95% (-40 °C to 70 °C)
Altitude
Operating 2,000 meters
Storage 12,000 meters
Vibration Complies with Mil-T-28800F Type III, Class 5 (Sine only)
Safety Designed to comply with IEC 61010-1:2000-1, UL
61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V
/ CAT II 600 V
EMC Designed to comply with IEC 61326-1:2000-11 (EMC)
whenusedwithshieldedcommunicationscables. This
meter has shown susceptibility to radiated frequencies
greaterthan1V/mfrom250to450MHz
Triggering
Characteristic Description
Samples per Trigger 1to50,000
Trigger Delay 0 s to 3600 s; in 10 s increments
External Trigger Delay <1 ms
External Trigger Jitter <500 s
Trigger Input TTL Levels
Trigger Output 5 V max (open collector)
Memory
Model Description
DMM4050/4040 10,000 measurements, internal, and up to 2 Gigabyte
capacity with USB memory module (available
separately) through front-panel USB port
Math Functions
Zero, dBm, dB, MX+B, Offset, DCV ratio and TrendPlot, Histogram, Statistics
(min/max/average/standard deviation), and Limit Test
Electrical
Characteristic Description
Input Protection 1000 V all ranges
Overrange 20% on all ranges except 1000 VDC, 1000 VAC, Diode,
and 10 A ranges
Remote Interfaces
RS-232C, DTE 9-pin, 1200 to 230400 baud (RS-232C to USB cable available
to connect the meter to a PC USB port) IEEE 488.2.
LAN and “Ethernet 10/100Base-T with DHCP (for IP address) option”.
Warranty
Three years
4www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
Electrical Specications
DC Voltage Specications
Accuracy specications are valid for digit resolution mode after at least a 1 hour
warm-up with Auto Zero enabled.
24-hour specications are relative to calibration standards and assume a controlled
electromagnetic environment per EN 61326-1:2000-11.
Characteristic Description
Maximum Input 1000 V on any range
Common Mode
Rejection
140 dB at 50 or 60 Hz ±0.1% (1 kunbalance)
Normal Mode Rejection 60 dB for NPLC of 1 or greater with analog lter off and
power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog lter on and
power line frequency ±0.1%
Measurement Method Multiramp A/D
A/D Linearity 0.0002% of measurement + 0.0001% of range
Input Bias Current <30 pA at 25 °C
Auto-zero Off Operation Following instrument warm-up at calibration temperature
±1 °C and less than 10 minutes, add error: 0.0002%
range additional error +5 V
Analog Filter When using the analog lter, specications are relative
towithinonehourofusingtheZEROfunctionforthat
range and NPLC setting
DC Ratio Accuracy is ±(Input accuracy + Reference accuracy),
where Input accuracy = DC Voltage accuracy for the HI
to LO Input (in ppm of the Input voltage), and Reference
accuracy = DC Voltage accuracy for the HI to LO
(Sense) Reference (in ppm of the Reference voltage)
Settling Considerations Measurement settling times are affected by source
impedance, cable dielectric characteristics, and input
signal changes
Input Characteristics
ResolutionRange Resolution
Digits Digits Digits
Input Impedance
100 mV 100.0000 mV 10 V1V 100 nV 10 Mor >10 G*2
1 V 1.000000 V 100 V10V1V10 Mor >10 G*2
10 V 10.00000 V 1 mV 100 V10V10 Mor >10 G*2
100 V 100.0000 V 10 mV 1 mV 100 V10 M±1%
1000 V 1,000.000 V 100 mV 10 mV 1 mV 10 M±1%
*2Inputs beyond ±14 V are clamped through 200 ktypical. 10 Mis default input impedance.
www.tektronix.com/dmm4050_4040 5
Data Sheet
DMM4050 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range 24 Hour
(23 °C ±1 °C)
90 Days
(2C)
1Year
(23 °C ±5 °C)
Temperature Coefcient/°C
Outside18to2C
100 mV 0.0025 + 0.003 0.0025 + 0.0035 0.0037 + 0.0035 0.0005 + 0.0005
1 V 0.0018 + 0.0006 0.0018 + 0.0007 0.0025 + 0.0007 0.0005 + 0.0001
10 V 0.0013 + 0.0004 0.0018 + 0.0005 0.0024 + 0.0005 0.0005 + 0.0001
100 V 0.0018 + 0.0006 0.0027 + 0.0006 0.0038 + 0.0006 0.0005 + 0.0001
1000 V 0.0018 + 0.0006 0.0031 + 0.001 0.0041 + 0.001 0.0005 + 0.0001
DMM4040 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range 24 Hour
(23 °C ±1 °C)
90 Days
(2C)
1Year
(23 °C ±5 °C)
Temperature Coefcient/°C
Outside18to2C
100 mV 0.003 + 0.003 0.004 + 0.0035 0.005 + 0.0035 0.0005 + 0.0005
1 V 0.002 + 0.0006 0.003 + 0.0007 0.004 + 0.0007 0.0005 + 0.0001
10 V 0.0015 + 0.0004 0.002 + 0.0005 0.0035 + 0.0005 0.0005 + 0.0001
100 V 0.002 + 0.0006 0.0035 + 0.0006 0.0045 + 0.0006 0.0005 + 0.0001
1000 V 0.002 + 0.0006 0.0035 + 0.0010 0.0045 + 0.0010 0.0005 + 0.0001
Additional Errors
Digits NPLC Additional NPLC Noise Error
100 0% of range
10 0% of range
10.001% of range
0.2 0.0025% of range ±12 V
0.02 0.017% of range ±17 V
6www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
AC Voltage Specications
AC Voltage specications are for AC sinewave signals >5% of range. For inputs from
1% to 5% of range and <50 kHz, add an additional error of 0.1% of range, and for
50 kHz to 100 kHz, add 0.13% of range.
Characteristic Description
Maximum Input 1000 VRMS or 1414 V peak or 8 × 107Volts-Hertz product
(whichever is less) for any range
Measurement Method AC-coupled true RMS. Measures the AC component of
input with up to 1000 VDC bias on any range
AC Filter Bandwidth
Slow 3Hz–300kHz
Medium 20 Hz 300 kHz
Fast 200 Hz 300 kHz
Common Mode
Rejection
70 dB at 50 Hz or 60 Hz ±0.1% (1 kunbalance)
Maximum Crest Factor 5:1 at Full Scale
Additional Crest Factor
Errors (<100 Hz)
Crest factor 1-2, 0.05% of full scale
Crestfactor2-3,0.2%offullscale
Crestfactor3-4,0.4%offullscale
Crestfactor4-5,0.5%offullscale
Only applies for non-sinusoid signals
Input Characteristics
ResolutionRange Resolution
Digits Digits Digits
Input Impedance
100 mV 100.0000 mV 10 V1V 100 nV
1 V 1.000000 V 100 V10V1V
10 V 10.00000 V 1 mV 100 V10V
100 V 100.0000 V 10 mV 1 mV 100 V
1000 V 1,000.000 V 100 mV 10 mV 1 mV
1M±2% shunted by
<100 pf
www.tektronix.com/dmm4050_4040 7
Data Sheet
DMM4050/4040 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range Frequency 24 Hour
(23 °C ±1 °C)
90 Days
(2C)
1 Year
(2C)
Temperature
Coefcient/°C
Outside18to2C
3–5Hz 1.0+0.03 1.0+0.04 1.0+0.04 0.1+0.004
5 10 Hz 0.35 + 0.03 0.35 + 0.04 0.35 + 0.04 0.035 + 0.004
10 Hz 20 kHz 0.04 + 0.03 0.05 + 0.04 0.06 + 0.04 0.005 + 0.004
20 50 kHz 0.1 + 0.05 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
100 mV
100 300 kHz*34.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.20 + 0.02
3–5Hz 1.0+0.02 1.0+0.03 1.0+0.03 0.1+0.003
5 10 Hz 0.35 + 0.02 0.35 + 0.03 0.35 + 0.03 0.035 + 0.003
10 Hz 20 kHz 0.04 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003
20 50 kHz 0.1 + 0.04 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
1V
100 300 kHz*34.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.2 + 0.02
3–5Hz 1.0+0.02 1.0+0.03 1.0+0.03 0.1+0.003
5 10 Hz 0.35 + 0.02 0.35 + 0.03 0.35 + 0.03 0.035 + 0.003
10 Hz 20 kHz 0.04 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003
20 50 kHz 0.1 + 0.04 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
10 V
100 300 kHz*34.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.2 + 0.02
3–5Hz 1.0+0.02 1.0+0.03 1.0+0.03 0.1+0.003
5 10 Hz 0.35 + 0.02 0.35 + 0.03 0.35 + 0.03 0.035 + 0.003
10 Hz 20 kHz 0.04 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003
20 50 kHz 0.1 + 0.04 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 100 kHz 0.55 + 0.08 0.6 + 0.08 0.6 + 0.08 0.06 + 0.008
100 V
100 300 kHz*34.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.2 + 0.02
3–5Hz 1.0+0.015 1.0+0.0225 1.0+0.0225 0.1+0.00225
5 10 Hz 0.35 + 0.015 0.35 + 0.0225 0.35 + 0.0225 0.035 + 0.00225
10 Hz 20 kHz 0.04 + 0.015 0.05 + 0.0225 0.06 + 0.0225 0.005 + 0.00225
20 50 kHz 0.1 + 0.03 0.11 + 0.0375 0.12 + 0.0375 0.011 + 0.00375
50 100 kHz*40.55 + 0.06 0.6 + 0.06 0.6 + 0.06 0.06 + 0.006
1000 V
100 300 kHz*3, 4 4.0 + 0.375 4.0 + 0.375 4.0 + 0.375 0.2 + 0.015
*3Typically 30% reading error at 1 MHz.
*41000 V range is limited to 8 × 107Volt-Hertz.
Additional Low Frequency Errors
Errorisstatedas%ofreading.
AC Filter
Frequency
3 Hz (Slow) 20 Hz (Medium) 200 Hz (Fast)
10 20 Hz 0 0.25
20 40 Hz 0 0.02
40 100 Hz 0 0.01 0.55
100 200 Hz 0 0 0.2
200 Hz 1 kHz 0 0 0.02
>1 kHz 0 0 0
8www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
Resistance
Specications are for 4-wire resistance function, 2 × 4-wire resistance, or 2-wire
resistance with zero. If zero is not used, add 0.2 for 2-wire resistance plus lead
resistance, and add 20 mfor 2 × 4-wire resistance function.
Characteristic Description
Measurement Method Current source referenced to LO input
Max Lead Resistance
(4-wire ohms)
10% of range per lead for 10 ,100,1kranges.
1kper lead on all other ranges
Input Protection 1000 V on all ranges
Common Mode
Rejection
140 dB at 50 or 60 Hz ±0.1% (1 kunbalance)
Normal Mode Rejection 60 dB for NPLC of 1 or greater with analog lter off and
power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog lter on and
power line frequency ±0.1%
Analog Filter When using the analog lter, specications are relative
towithinonehourofusingtheZEROfunctionforthat
range and NPLC setting
Input Characteristics
ResolutionRange Resolution
Digits Digits Digits
Source Current
10 10.00000 1m100  10  5mA/13V
100 100.0000 10 m1m100  1mA/6V
1k1.000000 k100 m10 m1m1mA/6V
10 k10.00000 k1100 m10 m100 A/6V
100 k100.0000 k10 1100 m100 A/13V
1M1.000000 M100 10 110 A/13V
10 M10.00000 M1k100 10 1A/13V
100 M100.0000 M10 k1k100 1A||10M/10V
1.0 G1.000000 G100 k10 k1k1A||10M/10V
DMM4050/4040 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range 24 Hour
(23 °C ±1 °C)
90 Days
(2C)
1Year
(23 °C ±5 °C)
Temperature Coefcient/°C
Outside18to2C
10 0.003 + 0.01 0.008 + 0.03 0.01+ 0.03 0.0006 + 0.0005
100 0.003 + 0.003 0.008 + 0.004 0.01 + 0.004 0.0006 + 0.0005
1k0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
10 k0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
100 k0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
1M0.002 + 0.001 0.008 + 0.001 0.01 + 0.001 0.001 + 0.0002
10 M0.015 + 0.001 0.02 + 0.001 0.04 + 0.001 0.003 + 0.0004
100 M0.3 + 0.01 0.8 + 0.01 0.8 + 0.01 0.15 + 0.0002
1G1.0 + 0.01 1.5 + 0.01 2.0 + 0.01 0.6 + 0.0002
Additional Ohms Errors
Digits NPLC Additional NPLC Noise Error
100 0% of range
10 0% of range
10.001% of range
0.2 0.003% of range ±7 m
0.02 0.017% of range ±15 m
www.tektronix.com/dmm4050_4040 9
Data Sheet
DC Current
Characteristic Description
Input Protection Tool-accessible 11 A / 1000 V and 440 mA / 1000 V
fuses, limits of 400 mA continuous 550 mA for 2 minutes
on, 1 minute off
Common Mode
Rejection
140 dB at 50 or 60 Hz ±0.1% (1 kunbalance)
Normal Mode Rejection 60 dB for NPLC of 1 or greater with analog lter off and
power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog lter on and
power line frequency ±0.1%
Analog Filter When using the analog lter, specications are relative
to within onehourofusingtheZEROfunctionforthat
range and NPLC setting
Input Characteristics
ResolutionRange Resolution
Digits Digits Digits
Shunt Resistance
(Ohms)
Burden Voltage
100 A 100.0000 A10nA 1nA 100pA 100 <0.015 V
1 mA 1.000000 mA 100 nA 10 nA 1 nA 100 <0.15 V
10 mA 10.00000 mA 1 A 100 nA 10 nA 1<0.025 V
100 mA 100.0000 mA 10 A1A 100 nA 1<0.25 V
400 mA*7400.000 mA 100 A10A1A1<0.50 V
1A*
61.000000 A 100 A10A1A0.01 <0.05 V
3A*
53.00000 A 1 mA 100 A10A0.01 <0.15 V
10 A 10.00000 A 1 mA 100 A10A0.01 <0.5 V
*5Part of 10 A range.
*6Available on the front-panel terminal only.
*7400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
DMM4050/4040 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range 24 Hour
(23 °C ±1 °C)
90 Days
(2C)
1Year
(23 °C ±5 °C)
Temperature Coefcient/°C
Outside18to2C
100 A 0.01 + 0.02 0.04 + 0.025 0.05 + 0.025 0.002 + 0.003
1 mA 0.007 + 0.005 0.030 + 0.005 0.05 + 0.005 0.002 + 0.0005
10 mA 0.007 + 0.02 0.03 + 0.02 0.05 + 0.02 0.002 + 0.002
100 mA 0.01 + 0.004 0.03 + 0.005 0.05 + 0.005 0.002 + 0.0005
400 mA*70.03 + 0.004 0.04 + 0.005 0.05 + 0.005 0.005 + 0.0005
1A*
60.03 + 0.02 0.04 + 0.02 0.05 + 0.02 0.005 + 0.001
3A*
5, 6 0.05 + 0.02 0.08 + 0.02 0.1 + 0.02 0.005 + 0.002
10 A*60.1 + 0.008 0.12 + 0.008 0.15 + 0.008 0.005 + 0.0008
*5Part of 10 A range.
*6Available on the front-panel terminal only.
*7400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
Additional Current Errors
Digits NPLC Additional NPLC Noise Error
for 1 mA, 100 mA, 400 mA,
3A,and10A
Additional NPLC Noise Error for
100 μA, 10 mA, 1 A
100 0% of range 0% of range
10 0% of range 0% of range
10.001% of range 0.01% of range
0.2 0.11% of range ±4 A 0.11% of range ±4 A
0.02 0.04% of range ±4 A 0.28% of range ±4 A
10 www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
AC Current
The following AC current specications are for sinusoidal signals with amplitudes
greater than 5% of range. For inputs from 1% to 5% of range, add an additional
error of 0.1% of range.
Characteristic Description
Input Protection Tool-accessible 11 A / 1000 V and 440 mA / 1000 V
fuses, limits of 400 mA continuous 550 mA for 2 minutes
on, 1 minute off
Measurement Method AC-coupled true RMS, DC-coupled to the fuse and shunt
(no blocking capacitor)
AC Filter Bandwidth
Slow 3Hzto10kHz
Medium 20 Hz to 10 kHz
Fast 200 Hz to 10 kHz
Maximum Crest Factor 5:1 at Full Scale
Additional Crest Factor
Errors (<100 Hz)
Crest factor 1-2, 0.05% of full scale
Crestfactor2-3,0.2%offullscale
Crestfactor3-4,0.4%offullscale
Crestfactor4-5,0.5%offullscale
Only applies to non-sinusoid signals
Input Characteristics
ResolutionRange Resolution
Digits Digits Digits
Shunt Resistance
(Ohms)
Burden Voltage
100 A 100.0000 A10nA 1nA 100pA 100 <0.015 V
1 mA 1.000000 mA 100 nA 10 nA 1 nA 100 <0.15 V
10 mA 10.00000 mA 1 A 100 nA 10 nA 1<0.025 V
100 mA 100.0000 mA 10 A1A 100 nA 1<0.25 V
400 mA*9400.000 mA 100 A10A1A1<0.50 V
1A*
81.000000 A 100 A10A1A0.01 <0.05 V
3A*
5, 8 3.00000 A 1 mA 100 A10A0.01 <0.05 V
10 A*810.00000 A 1 mA 100 A10A0.01 <0.5 V
*5Part of 10 A range.
*8Available at front-panel connectors only.
*9400 mA available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400 mA.
www.tektronix.com/dmm4050_4040 11
Data Sheet
DMM4050/4040 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range Frequency 24 Hour
(23 °C ±1 °C)
90 Days
(2C)
1 Year
(2C)
Temperature
Coefcient/°C
Outside18to2C
3–5Hz 1.1+0.06 1.1+0.06 1.1+0.06 0.2+0.006
5 10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.1 + 0.006
10 Hz 5 kHz 0.15 + 0.06 0.15 + 0.06 0.15 + 0.06 0.015 + 0.006
100 A
5–10kHz 0.35+0.7 0.35+0.7 0.35+0.7 0.03+0.006
3–5Hz 1.0+0.04 1.0+0.04 1.0+0.04 0.1+0.006
5 10 Hz 0.3 + 0.04 0.3 + 0.04 0.3 + 0.04 0.035 + 0.006
10 Hz 5 kHz 0.1 + 0.04 0.1 + 0.04 0.1 + 0.04 0.015 + 0.006
1mA
5 10 kHz 0.2 + 0.25 0.2 + 0.25 0.2 + 0.25 0.03 + 0.006
3–5Hz 1.1+0.06 1.1+0.06 1.1+0.06 0.2+0.006
5 10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.1 + 0.006
10 Hz 5 kHz 0.15 + 0.06 0.15 + 0.06 0.15+ 0.06 0.015 + 0.006
10 mA
5–10kHz 0.35+0.7 0.35+0.7 0.35+0.7 0.03+0.006
3–5Hz 1.0+0.04 1.0+0.04 1.0+0.04 0.1+0.006
5 10 Hz 0.3 + 0.04 0.3 + 0.04 0.3 + 0.04 0.035 + 0.006
10 Hz 5 kHz 0.1 + 0.04 0.1 + 0.04 0.1 + 0.04 0.015 + 0.006
100 mA
5 10 kHz 0.2 + 0.25 0.2 + 0.25 0.2 + 0.25 0.03 + 0.006
3–5Hz 1.0+0.1 1.0+0.1 1.0+0.1 0.1+0.006
5 10 Hz 0.3 + 0.1 0.3 + 0.1 0.3 + 0.1 0.035 + 0.006
10 Hz 5 kHz 0.1 + 0.1 0.1 + 0.1 0.1 + 0.1 0.015 + 0.006
400 mA*7
5 10 kHz 0.2 + 0.7 0.2 + 0.7 0.2 + 0.7 0.03 + 0.006
3–5Hz 1.0+0.04 1.0+0.04 1.0+0.04 0.1+0.006
5 10 Hz 0.3 + 0.04 0.3 + 0.04 0.3 + 0.04 0.035 + 0.006
10 Hz 5 kHz 0.1 + 0.04 0.1 + 0.04 0.1 + 0.04 0.015 + 0.006
1A*
6
5–10kHz 0.35+0.7 0.35+0.7 0.35+0.7 0.03+0.006
3–5Hz 1.1+0.06 1.1+0.06 1.1+0.06 0.1+0.006
5 10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.035 + 0.006
10 Hz 5 kHz 0.15 + 0.06 0.15 + 0.06 0.15 + 0.06 0.015 + 0.006
3A*
5, 6
5–10kHz 0.35+0.7 0.35+0.7 0.35+0.7 0.03+0.006
3–5Hz 1.1+0.06 1.1+0.06 1.1+0.06 0.1+0.006
5 10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.035 + 0.006
10 Hz 5 kHz 0.15 + 0.06 0.15 + 0.06 0.15 + 0.06 0.015 + 0.006
10 A*6
5–10kHz 0.35+0.7 0.35+0.7 0.35+0.7 0.03+0.006
*5Part of 10 A range.
*6Available on the front-panel terminal only.
*7400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
Additional Low-frequency Errors
Errorisstatedas%ofreading.
AC Filter
Frequency
3 Hz (Slow) 20 Hz (Medium) 200 Hz (Fast)
10 20 Hz 0 0.25
20 40 Hz 0 0.02
40 100 Hz 0 0.01 0.55
100 200 Hz 0 0 0.2
200 Hz 1 kHz 0 0 0.02
>1 kHz 0 0 0
12 www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
Frequency
Characteristic Description
Gate Time Programmable to 1 s, 100 ms, and 10 ms
Measurement
Method
Flexible counting technique. AC-coupled input using
the AC voltage measurement function
Settling
Considerations
When measuring frequency or period after a DC
offset voltage change, errors may occur. For the
most accurate measurement, wait up to 1 second for
the input blocking capacitor to settle
Measurement
Considerations
To minimize measurement errors, shield inputs
from external noise when measuring low-voltage,
low-frequency signals
DMM4050/4040 Accuracy
Accuracy is given as ± % measurement
Range Frequency 24 Hour
(23 °C ±1 °C)
90 Days
(2C)
1 Year
(2C)
Temperature
Coefcient/°C
Outside18to2C
3–5Hz 0.1 0.1 0.1 0.005
5 10 Hz 0.05 0.05 0.05 0.005
10 40 Hz 0.03 0.03 0.03 0.001
40 Hz 300 kHz 0.006 0.01 0.01 0.001
100 mV to 1000 V*10, 11
300 kHz 1 MHz 0.006 0.01 0.01 0.001
*10 Limitedto8×10
7Volt-Hertz
*11 Input >100 mV. For 10 100 mV, multiply percent measurement error by 10.
Gate Time vs. Resolution
Gate Time Resolution
0.01
0.1
1.0
Additional Low-frequency Errors
Error stated as percent of measurement for inputs >100 mV. For 10 100 mV,
multiply percent by 10.
NPLC
Frequency
3 5 Hz 0 0.12 0.12
5 10 Hz 0 0.17 0.17
10–40Hz 0 0.2 0.2
40 100 Hz 0 0.06 0.21
100 300 Hz 0 0.03 0.21
300 Hz 1 kHz 0 0.01 0.07
>1 kHz 0 0 0.02
www.tektronix.com/dmm4050_4040 13
Data Sheet
Capacitance (DMM4050 Only)
Accuracy is stated as ±(% of measurement + % of range)
Range Resolution 1 Year Accuracy*12
(23 °C ±5 °C)
Temperature Coefcient/°C
Outside18to2C
1nF 1pF 2% ±2.5% 0.05 + 0.05
10 nF 10 pF 1% ±0.5% 0.05 + 0.01
100 nF 100 pF 1% ±0.5% 0.01 + 0.01
1F1nF
1% ±0.5% 0.01 + 0.01
10 F10nF
1% ±0.5% 0.01 + 0.01
100 F100nF
1% ±0.5% 0.01 + 0.01
1mF 1F1% ±0.5% 0.01 + 0.01
10 mF 10 F1% ±0.5% 0.01 + 0.01
100 mF 100 F4% ±0.2% 0.05 + 0.05
*12 Stated accuracy is attained when Zero function is used.
Temperature (DMM4050 only)
Test Current: 1 mA
Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 lead resistance. The accuracy listed in the table below are valid
only when using the 4-wire RTD measurement function. Specications do not include probe accuracy, which must be added.
AccuracyRange Resolution
90 Days
(2C
)
1Year
(23 °C ±5 °C)
Temperature Coefcient/°C
Outside18to2C
-200 °C 0.001 °C 0.06 0.09 0.0025
-100 °C 0.001 °C 0.05 0.08 0.002
0 °C 0.001 °C 0.04 0.06 0.002
100 °C 0.001 °C 0.05 0.08 0.002
300 °C 0.001 °C 0.1 0.12 0.002
600 °C 0.001 °C 0.18 0.22 0.002
Additional Errors
Digits NPLC Additional NPLC Noise Error
100 C
10 C
10.03 °C
0.2 0.12 °C
0.02 0.6 °C
14 www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
Continuity
Characteristic Description
Continuity Threshold Selectable between 1 and 1000
Test Current 1mA
Response Time 300 S/s with audible tone
Accuracy is given as ±(% measurements + % of range)
Range 24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(2C)
Temperature
Coefcient/°C
Outside 18
to 28 °C
1000.0 0.002 + 0.01 0.008 + 0.02 0.01 + 0.02 0.001 + 0.002
Diode Test
Characteristic Description
Test Cur rent 100 Aor1mA
Response Time 300 S/s with audible tone
Accuracy is given as ±(% measurements + % of range)
Range 24 Hour
(2C)
90 Days
(23 °C ±5 °C)
1 Year
(2C)
Temperature
Coefcient/°C
Outside 18
to 28 °C
5.0000 V 0.002 + 0.002 0.008 + 0.002 0.01 + 0.002 0.001 + 0.002
10.0000 V 0.002 + 0.001 0.008 + 0.002 0.01 + 0.002 0.001 + 0.002
Measurement Rates (IEEE488*16)
Measurements/Second*13
Function Digits Setting Integration Time
60 Hz (50 Hz) DMM4040 DMM4050
100 NPLC 1.67 (2) s 0.6 (0.5) 0.6 (0.5)
10 NPLC 167 (200) ms 6 (5) 6 (5)
1 NPLC 16.7 (20) ms 60 (50) 60 (50)
0.2 NPLC 3.3 ms 270 270
DC Volts, DC Current, and
Resistance
0.02 NPLC 500 s 995 995
3Hz 0.47 0.47
20 Hz 1.64 1.64
AC Voltage and AC
Current*14
200 Hz*15 4.5 4.5
1s 1 1
100 ms 9.8 9.8
Frequency and Period
10 ms 80 80
Capacitance NA 2
*13 Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.
*14 Maximum measurement rates for 0.01% of AC step. When DC input varies, additional settling delay is required.
*15 For remote operation or external trigger using default settling delay.
*16 Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurement rates for RS232 can vary depending on the baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is
711 measurements. The LAN bus has a maximum measurement rate of 963 measurements.
www.tektronix.com/dmm4050_4040 15
Data Sheet
Ordering Information
Models
Model Description
DMM4040 6.5 Digit Multimeter
DMM4050 6.5 Digit Multimeter
DMM4050/4040 Includes: Meter, TL710 test leads, line cord, spare line fuse,
statement of cal practices, Warranty statement, Safety and Installation Guide,
CD-ROM with user manual (English, French, Italian, German, Spanish, Simplied
Chinese, Traditional Chinese, Korean, Russian, Japanese), RS-232 to USB Adapter
Cable, National Instruments LabVIEW SignalExpress™ Tektronix Edition, Limited
Edition Software.
Please specify power plug when ordering.
Instrument Options
Power Plug Options
Option Description
Opt. A0 North America
Opt. A1 Universal Euro
Opt. A2 United Kingdom
Opt. A3 Australia
Opt. A5 Switzerland
Opt. A6 Japan
Opt. A10 China
Opt. A11 India
Opt. A99 No power cord
Opt. E1 Euro and UK power cords
Service Options*17
Option Description
Opt. CA1 Provides a single calibration event or coverage for the
designated calibration interval, whichever comes rst
Opt. C3 Calibration Service 3 Years
Opt. C5 Calibration Service 5 Years
Opt. D1 Calibration Data Report
Opt. R5 Repair Service 5 Years (including warranty)
*17 Test Leads and accessories are not covered by the DMM warranty and Service Offerings. Refer to the
datasheet of each Test Lead and accessory model for its unique warranty and calibration terms.
Recommended Accessories and Software
Accessory Description
Calibration Manual 077-0362-xx
Programmer’s Manual 077-0363-xx
TP750 100 RTD Temperature Probe (DMM4050 only)
TL710 Premium Test Leads (196-3250-xx)
TL705 2×4 Wire Ohm 1000 V Precision Test Lead
TL725 2×4 Wire Ohm SMD Test Tweezers
AC4000 Soft Transit Case
HCTEK4321 Hard Carrying Case
Y8846S Single Rackmount Kit
Y8846D Dual Rackmount Kit
013-0369-xx Calibration Fixture 4-terminal short
SIGEXPTE NI LabVIEW SignalExpress Tektronix Edition Software
Full Version
Product(s) are manufactured in ISO registered facilities.
Product(s) complies with IEEE Standard 488.1-1987 and RS-232C.
16 www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
www.tektronix.com/dmm4050_4040 17
Data Sheet
18 www.tektronix.com/dmm4050_4040
Digital Multimeters Tektronix DMM4050 and DMM4040
www.tektronix.com/dmm4050_4040 19
Data Sheet Contact Tektronix:
ASEAN / Australasia (65) 6356 3900
Austria +41 52 675 3777
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 07 81 60166
Brazil +55 (11) 3759-7627
Canada 1 (800) 661-5625
Central East Europe, Ukraine, and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Denmark +4580881401
Finland +41526
75 3777
France +33(0)169868181
Germany +49 (221) 94 77 400
Hong Kong (852) 2585-6688
India (91) 80-42922600
Italy +39 (02) 25086 1
Japan 81 (3) 6714-3010
Luxembourg +44 (0) 1344 392400
Mexico, Central/South America & Caribbean 52 (55) 54247900
Middle East, Asia, and North Africa +41 52 675 3777
The Netherlands 090 02 021797
Norway 800 16098
People’s Republic of China 86 (10) 6235 1230
Poland +41 52 675 3777
Portugal 80 08 12370
Republic of Korea 82 (2) 6917-5000
Russia & CIS +7 (495) 7484900
South Africa +27 11 206 8360
Spain (+34) 901 988 054
Sweden 020 08 80371
Switzerland +41 52 675 3777
Taiwan 886 (2) 2722-9622
United Kingdom & Ireland +44 (0) 1344 392400
USA 1 (800) 426-2200
For other areas contact Tektronix, Inc at: 1 (503) 627-7111
Updated 5 August 2009
For Further Information. Tektronix maintains a comprehensive, constantly expanding
collection of application notes, technical briefs and other resources to help engineers working
on the cutting edge of technology. Please visit www.tektronix.com
Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents,
issued and pending. Information in this publication supersedes that in all previously published material.
Specication and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of
Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks
of their respective companies.
10 Sep 2009 3MW-23595-0